 
 
 
European Medicines Agency 
 
 
 
7 Westferry Circus, Canary Wharf, London, E14 4HB, UK 
Tel. (44-20) 74 18 84 00   Fax (44-20) 74 18 86 60 
E-mail: mail@emea.eu.int     http://www.emea.eu.int 
©EMEA 2009 Reproduction and/or distribution of this document is authorised for non commercial purposes only provided the EMEA is acknowledged 
1 
2 
3 
 
London, 19 November 2009 
 
Doc. Ref: EMEA/CHMP/EWP/192217/2009 
 
 
COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE 
4 
(CHMP) 
5 
6 
DRAFT 
GUIDELINE ON VALIDATION OF BIOANALYTICAL METHODS 
7 
8 
 
DRAFT AGREED BY THE EFFICACY WORKING PARTY 
September 2009 
ADOPTION BY CHMP FOR RELEASE FOR CONSULTATION 
19 November 2009 
END OF CONSULTATION (DEADLINE FOR COMMENTS) 
31 May 2010 
 
9 
Comments should be provided using this template to EWPSecretariat@emea.europa.eu  
 
10 
KEYWORDS 
CHMP, EMEA, Guideline, validation, bioanalytical method, analyses 
 
2/17 
 
GUIDELINE ON VALIDATION OF BIOANALYTICAL METHODS 
11 
12 
13 
14 
15 
16 
17 
18 
19 
20 
21 
22 
23 
24 
25 
26 
27 
28 
29 
30 
31 
32 
33 
34 
35 
36 
37 
38 
TABLE OF CONTENTS 
1. 
INTRODUCTION (BACKGROUND)........................................................................................ 3 
2. 
SCOPE............................................................................................................................................ 3 
3. 
LEGAL BASIS .............................................................................................................................. 3 
4. 
METHOD VALIDATION............................................................................................................ 4 
4.1 
COMPLETE VALIDATION OF AN ANALYTICAL METHOD........................................................... 4 
4.1.1 
Selectivity......................................................................................................................... 4 
4.1.2 
Carry-over....................................................................................................................... 5 
4.1.3 
Lower limit of quantitation.............................................................................................. 5 
4.1.4 
Calibration curve ............................................................................................................ 5 
4.1.5 
Accuracy.......................................................................................................................... 6 
4.1.6 
Precision.......................................................................................................................... 7 
4.1.7 
Dilution integrity............................................................................................................. 7 
4.1.8 
Matrix effect .................................................................................................................... 7 
4.1.9 
Stability............................................................................................................................ 8 
4.2 
PARTIAL VALIDATION.............................................................................................................. 9 
4.3 
CROSS VALIDATION................................................................................................................. 9 
4.4 
LIGAND-BINDING ASSAYS........................................................................................................ 9 
5. 
ANALYSIS OF STUDY SAMPLES.......................................................................................... 10 
5.1 
ANALYTICAL RUN.................................................................................................................. 11 
5.2 
ACCEPTANCE CRITERIA OF AN ANALYTICAL RUN................................................................. 11 
5.3 
CALIBRATION RANGE ............................................................................................................ 12 
5.4 
REANALYSIS OF STUDY SAMPLES.......................................................................................... 12 
5.5 
INTEGRATION......................................................................................................................... 13 
6. 
INCURRED SAMPLES REANALYSIS................................................................................... 13 
7. 
STUDY REPORT........................................................................................................................ 13 
DEFINITIONS .................................................................................................................................... 16 
 
3/17 
 
39 
40 
41 
42 
43 
45 
46 
47 
48 
49 
50 
51 
52 
53 
54 
56 
57 
58 
59 
60 
61 
62 
64 
65 
66 
67 
68 
69 
70 
71 
72 
73 
74 
75 
76 
77 
78 
EXECUTIVE SUMMARY 
This guideline defines key elements and provides recommendations for the validation of bioanalytical 
methods. The guideline focuses on the validation of the analytical methods used for pharmacokinetic 
sample analysis. In addition, guidance will be provided with regard to the actual analysis of study 
samples.  
1. 
INTRODUCTION (background) 
44 
Measurement of drug concentrations in biological matrices is an important aspect of medicinal product 
development for those products containing new active substances as well as for line extensions and 
generic products. Such data may be required to support new applications as well as variations to 
authorised drug products. The results of toxicokinetic, pharmacokinetic and bioequivalence studies are 
used to make critical decisions supporting the safety and efficacy of a medicinal drug substance or 
product. It is therefore paramount that the applied bioanalytical methods used are well characterised, 
fully validated and documented to a satisfactory standard in order to yield reliable results. 
Acceptance criteria wider than those defined in this guideline may need to be used in special 
situations, such as analysis of complex matrices (e.g. solid tissues), when usual acceptance criteria 
cannot be met. This should be justified and prospectively defined.   
2. 
SCOPE 
55 
This guideline provides requirements for the validation of bioanalytical methods.  
In addition, specific aspects of the bioanalytical method itself will be addressed, e.g. the actual 
analysis of samples from toxicokinetic studies and clinical trials. 
Furthermore, this guideline will describe when partial validation or cross validation may represent an 
appropriate alternative approach to the complete validation of an analytical method.  
Some special techniques such as radio-labelled analysis methods using 14C labelled drugs, are not 
covered here, but even in such cases efforts should be made to apply to the principles of this guideline. 
3. 
LEGAL BASIS 
63 
This guideline has to be read in conjunction with the introduction and general principles (4) and Part I 
and II of the Annex I to Directive 2001/83 as amended. It applies to Marketing Authorisation 
Applications for human medicinal products submitted in accordance with the Directive 2001/83/EC as 
amended, and Regulation (EC) No. 726/2004, in which the analysis of drug concentrations in a 
biological matrix is part of the application.  
The validation of bioanalytical methods and the analysis of study samples should be performed in 
accordance with the principles of Good Laboratory Practice (GLP). However, as human bioanalytical 
studies fall outside of the scope of GLP, as defined in Directive 2004/10/EC, the sites conducting the 
human studies are not required to be monitored as part of a national GLP compliance programme. In 
addition, for clinical trials in humans the principles of Good Clinical Practice (GCP) should be 
followed. 
Furthermore, reference is made to the following EMEA guidelines: 
• 
Note for guidance on good clinical practices (CPMP/ICH/135/95). 
• 
Note for guidance on validation of analytical procedures: text and methodology 
(CPMP/ICH/381/95). 
 
4/17 
 
81 
82 
83 
84 
85 
86 
87 
88 
89 
90 
91 
92 
93 
94 
95 
96 
97 
98 
99 
100 
101 
102 
103 
104 
105 
106 
107 
108 
109 
110 
111 
112 
113 
114 
115 
116 
117 
119 
120 
121 
4. 
METHOD VALIDATION 
79 
4.1 
Complete validation of an analytical method  
80 
A complete method validation should be performed for any analytical method whether new or based 
upon literature. 
The main objective of method validation is to demonstrate the reliability of a particular method for the 
determination of an analyte concentration in a specific biological matrix, such as blood, plasma, urine, 
saliva or tissue. Moreover, validation should be performed using the same anticoagulant as for the 
study samples. A full validation should be performed for each species concerned.  
In some cases, it may be problematic for validation purposes to obtain an identical matrix compared to 
the matrix of the study samples. A suitable alternative matrix may be used, e.g. synthetically prepared 
cerebrospinal fluid, if justified. 
The main characteristics of a bioanalytical method that are essential to ensure the acceptability of the 
performance and the reliability of analytical results are: selectivity, lower limit of quantitation, the 
response function (calibration curve performance), accuracy, precision, matrix effects, stability of the 
analyte(s) and any internal standard in the biological matrix and the stock and working solutions under 
the entire period of storage and processing conditions. 
Usually one analyte or drug has to be determined, but on occasions it may be appropriate to measure 
more than one analyte. This may involve two different drugs, but can also involve a parent drug with 
its metabolites, or the enantiomers or isomers of a drug. In these cases the principles of validation and 
analysis apply to all analytes of interest.  
Reference standards 
During method validation, a blank biological matrix will be spiked with the analyte of interest using 
solutions of reference standard. In addition, an internal standard (IS) is normally used in 
chromatographic methods.  
It is important that the quality of the reference standard and IS is ensured, as the quality (purity) may 
affect the outcome of the analysis, and therefore the outcome of the study data. Therefore the 
reference standards used for the analytical validation and analysis should be obtained from an 
authentic and traceable source. Suitable reference standards, include certified standards such as 
compendial standards (EPCRS, USP, WHO), commercial available standards, or fully characterised 
standards prepared in-house or by an external non-commercial organisation. Suitability of the 
reference standard should be scientifically justified. The use of certified standards is not needed for IS, 
as long as the suitability for use is demonstrated, e.g. lack of interference is shown for the substance 
itself or any impurities thereof. 
Whoever the supplier, a certificate of analysis is required to ensure quality, stability, storage 
conditions, expiration date, batch number and purity of the reference standards. 
When MS detection is used in the bioanalytical method, a stable isotope-labelled IS is recommended 
to be used whenever possible. However, it is essential that the labelled standard is of the highest 
isotope purity and that no isotope exchange reaction occurs. The presence of any unlabelled analyte 
would otherwise introduce a bias in the results. 
4.1.1 
Selectivity 
118 
The analytical method should be able to differentiate the analyte(s) of interest and IS from endogenous 
components in the matrix (i.e. blood, plasma, urine) or other components in the sample. Selectivity 
should be proven by using at least 6 sources of the appropriate blank matrix, which are individually 
 
5/17 
 
122 
123 
124 
125 
126 
127 
128 
129 
130 
131 
132 
133 
134 
135 
136 
137 
138 
139 
141 
142 
143 
144 
145 
146 
147 
148 
149 
151 
152 
153 
155 
156 
157 
158 
159 
160 
161 
162 
163 
164 
165 
166 
analysed and evaluated for interference. Absence of interfering components is accepted where the 
response is less than 20% of the lower limit of quantitation for the analyte. 
It may also be necessary to investigate the extent of any interference caused by metabolites of the 
drug(s), interference from degradation products formed during sample preparation, and interference 
from possible co-administered medications. Co-medications normally used in the subject population 
studied should be taken into account.  
The possibility of back-conversion of a metabolite into parent analyte during the successive steps of 
the analysis (including extraction procedures) should also be evaluated, when relevant (e.g. acidic 
metabolites to ester, unstable N-oxides or glucuronide metabolites, lactone-ring structures). 
Preferably, blank matrix (and/or samples spiked with analyte at a concentration not higher than  
3 times the lower limit of quantitation) should be spiked with concentrations of the metabolite of 
interest, representing the actual highest in vivo metabolite concentrations, the sample should be 
processed, and the chromatogram should be evaluated for the formation of the parent analyte. The 
extent of back-conversion should be established and the impact on the study results discussed. It is 
acknowledged that this evaluation will not be possible early during drug development of a new 
chemical entity when the metabolism is not yet evaluated. However, it is expected that this issue is 
taken into account and the analytical method is revalidated as further knowledge regarding metabolism 
of the active substance is gained during drug development. 
4.1.2 
Carry-over 
140 
Carry-over should be addressed and minimised during method development. Carry-over may not 
affect accuracy and precision (see section 4.1.5 and 4.1.6). During validation carry-over should be 
assessed by injecting blank samples after a high concentration sample or calibration standard. If it 
appears that carry-over is unavoidable, specific measures should be considered, tested during the 
validation and applied during the analysis of the study samples. This could include the injection of 
blank samples after samples with an expected high concentration, before the analysis of the next study 
sample. 
Randomisation of samples should be avoided, as this may interfere with the detection and assessment 
of carryover problems. 
4.1.3 
Lower limit of quantitation 
150 
The lower limit of quantitation (LLOQ) is the lowest amount of analyte in a sample which can be 
quantified reliably, with an acceptable accuracy and precision (see Accuracy and Precision). The 
LLOQ should be adapted to expected concentrations and to the aim of the study.  
4.1.4 
Calibration curve 
154 
The response of the instrument with regard to the analyte should be known, and should be evaluated 
over a specified concentration range. The concentrations to be analysed (calibration standards) should 
be prepared in the same matrix as the matrix of the intended study samples by spiking the blank matrix 
with known concentrations of the analyte (and IS). There should be one calibration curve for each 
analyte studied in the method validation and for each analytical run. 
Before carrying out the validation of the analytical method it should be known what concentration 
range is expected. This range should be covered by the calibration curve range, defined by the LLOQ 
being the lowest calibration standard and the upper limit of quantitation (ULOQ), being the highest 
calibration standard. The range should be justified based on scientific information.  
A minimum of six calibration concentration levels should be used, excluding the blank sample 
(processed matrix sample without analyte and without IS) and a zero sample (processed matrix with 
IS).  
 
6/17 
 
167 
168 
169 
170 
171 
172 
173 
174 
175 
176 
177 
178 
179 
180 
181 
183 
184 
185 
186 
187 
188 
189 
190 
191 
192 
193 
194 
195 
196 
197 
198 
199 
200 
201 
202 
203 
204 
205 
206 
207 
208 
A relationship which can simply and adequately describe the response of the instrument with regard to 
the analyte should be applied. The blank and zero samples should not be taken into consideration to 
calculate the calibration curve parameters.   
The calibration curve parameters should be submitted (slope and intercept in case of linear fit). In 
addition, the back calculated concentrations of the calibration standards should be presented together 
with the calculated mean accuracy values (see definition of Accuracy below). At least 3 calibration 
curves should be evaluated.  
The back calculated concentrations of the calibration standards should be within ±15% of the nominal 
value, except for the LLOQ for which it should be within ±20%. At least 75% of the calibration 
standards with a minimum of six, must fulfil this criterion. In case a calibration standard does not 
comply with these criteria, this calibration standard sample should be rejected, and the calibration 
curve without this calibration standard should be re-evaluated, including regression analysis. 
Although it may be clear from stability data that the analyte is sufficiently stable in the matrix of 
interest, it is recommended that freshly prepared calibration curves are used during validation of the 
bioanalytical method. 
4.1.5 
Accuracy 
182 
The accuracy of an analytical method describes the closeness of the determined value obtained by the 
method to the true concentration of the analyte (expressed in percentage). Accuracy should be 
assessed on samples spiked with known amounts of the analyte, the quality control samples (QC 
samples).  
During method validation accuracy should be determined by replicate analysis using a minimum of  
5 determinations at a minimum of 4 concentration levels which are covering the calibration curve 
range: the LLOQ, within three times the LLOQ (low QC), around 50% of the calibration curve range 
(medium QC), and at about 75% of the upper calibration curve range (high QC).  
The QC samples are analysed against the calibration curve, and the obtained concentrations are 
compared with the nominal value. The accuracy should be reported as percent of the nominal value. 
Accuracy should be evaluated for the values of the QC samples obtained within a single run (the 
within run accuracy) and in different runs (the between-run accuracy). The latter will support the 
accuracy over time.  
To enable evaluation of any trends over time within one run, it is recommended to demonstrate 
accuracy of QC samples over at least one of the runs with a size equivalent to a prospective analytical 
run. 
Within-run accuracy 
For the validation of the within-run accuracy, there should be a minimum of five samples per 
concentration level at LLOQ, low, medium and high QC samples in a single run. The mean accuracy 
value should be within 15% of the nominal values for the QC samples, except for the LLOQ which 
should be within 20% of the nominal value.    
Between –run accuracy 
For the validation of the between-run accuracy at least five determinations per concentration per run at 
LLOQ, low, medium and high QC samples from three runs analysed on at least two different days 
should be evaluated. The mean accuracy value should be within 15% of the nominal values for the QC 
samples, except for the LLOQ which should be within 20% of the nominal value.    
 
7/17 
 
209 
210 
211 
213 
214 
215 
216 
217 
218 
219 
220 
221 
222 
223 
225 
226 
227 
228 
229 
231 
232 
233 
234 
235 
236 
237 
238 
239 
240 
241 
242 
243 
244 
245 
246 
247 
248 
249 
Reported method validation data and the determination of accuracy and precision should include all 
outliers; however, calculations of accuracy and precision excluding values that are statistically 
determined as outliers should additionally be reported.  
4.1.6 
Precision 
212 
The precision of the analytical method describes the closeness of repeated individual measures of 
analyte. Precision is expressed as the coefficient of variation (CV). The statistical method for 
estimation of the precision should be predefined and calculated according standard practise. Precision 
should be demonstrated for the LLOQ, low, medium and high QC samples, within a single run and 
between different runs, i.e. using the results generated for demonstration of accuracy.  
Within-run precision 
The within-run CV value should not exceed 15% for the QC samples, except for the LLOQ which 
should not exceed 20%.    
Between –run precision 
The between-run CV value should not exceed 15% for the QC samples, except for the LLOQ which 
should not exceed 20%. 
4.1.7 
Dilution integrity 
224 
Dilution of samples should not interfere with the accuracy and precision. Dilution integrity should be 
demonstrated by spiking the matrix with an analyte concentration above the ULOQ and dilution of this 
sample with blank matrix (at least five determinations per dilution factor). Accuracy and precision 
should be within the set criteria, i.e. within ±15%. Dilution integrity should cover the applied dilution 
of the study samples. 
4.1.8 
Matrix effect 
230 
Matrix effects should be investigated when using mass spectrometric methods, using at least 6 lots of 
matrix including haemolysed, hyperlipidaemic and if applicable, sample matrix from special 
populations, such as renally or hepatically impaired populations. 
For each analyte and the internal standard, the matrix factor (MF) should be calculated in each lot of 
matrix, by calculating the ratio of the peak area in the presence of matrix (measured by analysing 
blank matrix spiked with analyte at a concentration of maximum 3 times the LLOQ after extraction), 
to the peak area in absence of matrix (pure solution of the analyte). The IS normalised MF should also 
be calculated by dividing the MF of the analyte by the MF of the IS.  
The CV of the IS-normalised MF calculated from the 6 batches of matrix should not be greater than  
15 %. 
If this approach cannot be used, for instance in the case of on-line sample preparation, the variability 
of the response from batch to batch should be assessed by analysing at least 6 batches of matrix in 
triplicate, spiked at a concentration of a maximum of 3 times the LLOQ. The validation report should 
include the peak areas of the analyte and of the IS and the calculated concentration for each individual 
sample. The overall CV calculated for the concentration should not be greater than 15 %. The mean 
concentration should be within 15 % of the nominal concentration. The mean concentration should 
also be reported for each individual batch of matrix; a deviation of this mean from the nominal 
concentration of more than 20 % in any individual batch of matrix should lead to additional 
investigations. 
 
8/17 
 
250 
251 
252 
253 
254 
255 
256 
258 
259 
260 
261 
262 
263 
264 
265 
266 
267 
268 
269 
270 
271 
272 
273 
274 
275 
276 
277 
278 
279 
280 
281 
282 
283 
284 
285 
286 
287 
288 
If the matrix is difficult to obtain, less than 6 different batches of matrix may be used, but this should 
be justified. However, matrix effects should still be investigated. 
If a formulation for injection to be administered to the subjects or animals contains excipients known 
to be responsible for matrix effects, for instance polyethylene glycol or polysorbate, matrix effects 
should be studied with matrix containing these excipients, in addition to blank matrix. The matrix used 
for this evaluation should be obtained from subjects or animals administered the excipient, unless it 
has been demonstrated that the excipient is not metabolised or transformed in-vivo. 
4.1.9 
Stability 
257 
Evaluation of the stability should be carried out to ensure that every step taken during sample 
preparation and sample analysis, as well as the storage conditions used do not affect the concentration 
of the analyte. Any deviation from the initial concentration that does occur must be within acceptable 
limits.  
Stability should be ensured for every step in the analytical method, meaning that the conditions 
applied to the stability tests, such as sample matrix, materials storage and analytical conditions should 
be similar to those used for the actual study samples. Stability cannot be proven by literature data. 
Stability of the analyte and IS in the studied matrix is evaluated using at least triplicates samples of the 
low and high QC samples which are analysed immediately after preparation and after the applied 
storage conditions that are to be evaluated. The QC samples are analysed against a calibration curve, 
obtained from freshly prepared calibration standards, and the obtained concentrations are compared to 
the nominal concentrations. The deviation should be within ±15%.   
Stability of the stock and working solutions should be tested with an appropriate dilution, taking into 
consideration the linearity and measuring range of the detector. 
Stability studies should investigate different storage conditions over time periods that equal or exceed 
those applied to the actual study samples. 
Normally, as an example, the following stability tests should be evaluated: 
• 
stock solution and working solution stability, 
• 
freeze and thaw stability of the analyte in the matrix from freezer storage conditions to room 
temperature, 
• 
stability of the analyte in matrix stored in the refrigerator, if applicable, 
• 
bench top stability of the analyte in matrix at room temperature, 
• 
long term stability of the analyte in matrix stored in the freezer at the same storage 
temperature as the study samples, 
• 
bench top stability of the processed sample at room temperature or under the storage 
conditions to be used during the study (dry extract or in the injection phase), if applicable, 
• 
on-instrument/ autosampler stability of the processed sample at injector or autosampler 
temperature.  
Regarding the freeze and thaw stability: The QC samples are stored and frozen in the freezer at the 
intended temperature and thereafter thawed at room temperature. After thawing, samples are refrozen 
again applying the same conditions. At each cycle, samples should be frozen for at least 12 hours 
 
9/17 
 
289 
290 
291 
292 
293 
294 
295 
296 
297 
298 
299 
301 
302 
303 
304 
305 
306 
307 
309 
310 
311 
312 
313 
314 
315 
316 
318 
319 
320 
321 
322 
323 
324 
325 
326 
327 
328 
329 
330 
before they are thawed. The number of cycles in the freeze-thaw stability should equal or exceed that 
of the freeze/thaw cycles of study samples.  
Regarding long term stability of the analyte in matrix stored in the freezer: The QC samples should be 
stored in the freezer under the same storage conditions and at least for the same duration as the study 
samples. For the evaluation of the long term stability it is not acceptable to use study samples, as the 
nominal concentration is unknown, and can therefore not be used as reference. It is recommended that 
evaluation of long term stability is carried out before the start of the actual study. 
Sufficient attention should be paid to the stability of the analyte in the sampled matrix directly after 
blood sampling of subjects and further preparation before storage, to ensure that the obtained 
concentrations by the analytical method reflect the concentrations of the analyte in the subject at the 
moment of sampling.. 
4.2 
Partial validation 
300 
In situation where minor changes are made to an analytical method that has already been validated, a 
full validation may not be necessary, depending onto the nature of the applied changes. Changes for 
which a partial validation may be needed include transfer of the bioanalytical method to another 
laboratory, change in equipment, calibration concentration range, storage conditions etc. All 
modifications should be reported and the scope of revalidation or partial validation justified. 
In most cases, provision of additional accuracy and precision data or relevant additional stability data 
on the modified issue may be sufficient.  
4.3 
Cross validation 
308 
Where data are obtained from different study sites, comparison of those data is needed, and a cross 
validation of the applied analytical methods should be carried out. Differences in sample preparation 
or the use of another analytical method may result in different outcomes between the study sites. Cross 
validation should be performed in advance of study samples being analysed if possible. For the cross 
validation, the same set of QC samples should be analysed by both analytical methods. The outcome 
of the cross validation is critical in determining whether the obtained data are reliable and whether 
they can be compared and used. The difference between the two measurements should not exceed 
15%. 
4.4 
Ligand-binding assays 
317 
Ligand-binding assays or immunoassays are nowadays especially used for macromolecules. In these 
assays, quantification is based on macromolecular interactions between the macromolecule and 
antibody.  
The validation principles and the issues with regard to analysis of study samples as indicated before, 
should also be applied in general for ligand-binding assays. However, the following issues need 
special attention.  
One of the issues is the fact that macromolecules tend to be heterogeneous (small differences in for 
instance glycosylation, or phosphorylation). Therefore validation must cover issues such as batch-to-
batch differences in glycosylation and metabolic differences in phosphorylation. In addition, 
macromolecules may be structurally comparable to endogeneous compounds. Therefore, specificity of 
the antibody and selectivity of the assay are critical parameters for ligand-binding assays.  
Specificity of an antibody refers to its ability to mainly bind the antigen of interest. Ideally the 
antibody should be specific such that no cross-reactivity with structural related compounds occurs.  
 
10/17 
 
331 
332 
333 
334 
335 
336 
337 
338 
339 
340 
341 
342 
343 
344 
345 
346 
347 
348 
349 
350 
351 
352 
353 
354 
355 
356 
357 
358 
359 
360 
361 
362 
363 
364 
365 
366 
367 
368 
369 
371 
372 
373 
374 
Specificity is validated by using at least 10 sources of sample matrix, spiked at or near the LLOQ. The 
presence of endogenous antibodies to the analyte may interfere with the analysis and should be taken 
into account. As interference may be concentration dependent, it would be helpful to estimate the 
concentration below which interference occurs. Due to interference of endogenous compounds, the 
blank response may exceed 20% of the LLOQ, however this may be acceptable, as long as it does not 
affect accuracy. If a surrogate matrix is used for validation, the comparability with regard to study 
sample matrix should be demonstrated, because of the disease state of a subject, the study matrix may 
contain different components that may interfere. It is important that the sample matrix and the 
standard matrix are equivalent. If not, it must be shown that the dose-response relation (e.g. the slope 
and asymptotes of a four-parameter curve) is unaffected by the matrix. 
Reference standards should be selected in such a manner that specificity is ensured, and binding 
characteristics are durable and stable regarding antibody/antigen complex formation. A change in 
reference standard during analysis may affect this, and suitability of the new reference standard should 
thus be demonstrated.   
With regard to the calibration curve, the response is measured indirectly and as a result the response 
function is in most cases nonlinear. Moreover, generally the calibration curve range is limited (<2 
orders of magnitude). In case anchor calibration standards are used, the additive value should be 
clearly documented. Accuracy of the back calculated concentrations should be within ±20%, except 
for LLOQ and ULOQ which should not exceed 25%. 
At least three QC samples should be included (low, medium and high) in each analytical run. The 
within-run and the between-run accuracy and precision should be within ±20%. Furthermore, the total 
error (sum of the absolute value of mean accuracy and precision should be less than 30% (40% at the 
LLOQ and ULOQ). The results of all runs should be included for validation, except in cases where 
errors are obvious and documented. 
It is recommended that the size of the run during validation is comparable to that of a run during 
analysis of study samples (or visa versa).  
It should be demonstrated that characteristics of the analyte are not affected by the methods of sample 
preparation, additives (e.g. anticoagulants), and stability during the whole process. Evaluation should 
not only address chemical and physical properties, but also biological integrity (i.e. maintenance of 
antibody binding affinity). In addition, this also accounts for changes after validation, e.g. a change in 
used diluents or reagents during analysis of study samples.  
If for analysis of study samples microtiter plate-based assays are used, a batch may comprise several 
individual plates, but each plate should contain an individual set of calibration standards and QC 
samples, to compensate for difference in plate performance.  
Commercial kits 
Commercial kits may have been developed for other use or purposes than the intended use, e.g. 
analysis of samples obtained in bioequivalence studies. Therefore, commercial kits need to be re-
validated, meaning that the LOQ and the QC samples in the actual concentration range perform 
accurately and precisely. The principles of validation listed before apply. 
5. 
ANALYSIS OF STUDY SAMPLES 
370 
After complete validation of the analytical method, analysis of study or subject samples may be 
carried out. Depending on the time period between validation and the analysis of the study samples it 
may be necessary to verify the performance of the method before start of the analysis of study 
samples. 
 
11/17 
 
375 
376 
377 
379 
380 
381 
382 
383 
384 
385 
386 
387 
388 
389 
390 
391 
392 
393 
394 
395 
397 
398 
399 
400 
401 
402 
403 
404 
405 
406 
407 
408 
409 
410 
411 
412 
413 
414 
415 
416 
417 
418 
419 
The study samples should be processed in accordance with the validated analytical method, and 
together with the QC samples and the calibration curve to ensure the acceptability of the analytical 
run.  
5.1 
Analytical run 
378 
An analytical run consists of the blank sample (processed matrix sample without analyte and without 
IS) and a zero sample (processed matrix with IS), a set of calibration standards at a minimum of 6 
concentration levels, at least 3 levels of QC samples (low, medium and high) in duplicate (or at least 5 
% of the number of study samples, whichever is higher), and study samples to be analysed. All 
samples (calibration standards, QC, and study samples) should be processed and extracted as one 
single batch of samples. Analysing as a single analytical run samples prepared separately as several 
batches should be avoided. If such an approach cannot be avoided, for instance due to bench-top 
stability limitations, each batch of samples should preferably include low, medium and high QC 
samples. Acceptance criteria should be pre-established in a Standard Operating Procedure (SOP) or in 
the study plan and should be defined both globally for the full analytical run, and for each individual 
batch of samples. 
It is advised to analyse all samples of one subject together in one analytical run to reduce the 
variability in outcome. Moreover, it is considered acceptable to analyse study samples of more than 
one subject in one analytical run. The QC samples should be divided over the run i.e. at the beginning, 
middle and at the end of the run. It is also acceptable to place one of each low, medium and high QC 
samples at the beginning of the analytical run followed by the study samples and at the end of the run 
again one of each low, medium and high QC samples. 
5.2 
Acceptance criteria of an analytical run 
396 
Criteria for acceptance or rejection of an analytical run should be defined in the analytical study 
protocol, in the study plan or in a SOP. The following acceptance criteria should apply: 
Accuracy:  
The back calculated concentrations of the calibration standards should be within ±15% of the nominal 
value, except for the LLOQ for which it should be within ±20%. At least 75% of the calibration 
standards with a minimum of six, must fulfil this criterion. If one of the calibration standards does not 
meet these criteria, this calibration standard sample should be rejected and the calibration curve 
without this calibration standard should be re-evaluated, and regression analysis performed. Criteria 
for decision to exclude calibration standards or not should be pre-defined in a SOP, and should be 
independent from the results of the QC samples. 
If the rejected calibration standard is the LLOQ, it should be realised that the LLOQ for this analytical 
run is the next acceptable higher calibration standard concentration of the calibration curve.  If the 
highest calibration standard is rejected, the ULOQ, for this analytical run is the concentration of the 
next acceptable lower calibration standard of the calibration curve. The revised calibration range must 
cover all QC samples (low, medium and high). 
The accuracy values of the QC samples should be within ±15% of the nominal values. At least 4 out 
of 6 (67%) QC samples and at least 50% at each concentration level should comply with this criterion. 
In case these criteria are not fulfilled the analytical run should be rejected, and the study samples 
reanalysed.  
The between-run (mean) accuracy of the QC samples should be within 15% of the nominal value. 
In the case of the simultaneous determination of several analytes, if an analytical run is acceptable for 
one analyte but has to be rejected for another analyte, the data for the accepted analyte can be used,  
but the samples should be re-analysed for determination of the rejected analyte. 
 
12/17 
 
420 
421 
422 
423 
424 
425 
426 
427 
428 
429 
431 
432 
433 
434 
435 
436 
437 
439 
440 
441 
442 
443 
444 
445 
446 
447 
448 
449 
450 
451 
452 
453 
454 
455 
456 
457 
458 
459 
Precision:  
The between-run precision should not exceed 15%. 
Ligand-binding assays: 
The back calculated concentrations of the calibration standards should be within ±20% of the nominal 
value, except for the LLOQ and the ULOQ for which it should be within ±25%. At least 75% of the 
calibration standards with a minimum of six, must fulfil this criterion. This requirement does not apply 
to ‘anchor’ calibrators. 
The accuracy values of the QC samples should be within ±20% of the nominal values. At least 4 out 
of 6 (67%) QC samples and at least 50% at each concentration level should apply to this criterion. 
Exceptions to this criterion should be justified.   
5.3 
Calibration range 
430 
If a large number of the analyte concentrations of the study samples appear to be above the ULOQ, the 
calibration curve range should be extended, if possible. The same applies if it appears that the 
calibration curve range is too wide, such that most analyte concentrations fall in the lower part of the 
calibration curve range. The calibration curve range should then be narrowed, or an additional QC 
sample level should be included so that at least 2 QC sample levels fall within the range of 
concentrations measured in study samples. If the calibration curve range is extended, the analytical 
method should be revalidated to ensure accuracy and precision. 
5.4 
Reanalysis of study samples 
438 
Reanalysis of study samples should be predefined in the study protocol or SOP, before the actual start 
of the analysis of the samples. The number of samples (and percentage of total number of samples) 
that have been reanalysed should be discussed in the study report. The reanalysed samples should be 
identified and the initial value, the reason for reanalysis, the values obtained in the reanalyses, the 
finally accepted value and a justification for the acceptance should be provided.  
The following reasons can be identified to reanalyse study samples: 
• 
rejection of an analytical run because the run did not fulfil the acceptance criteria with regard 
to accuracy and precision of the calibration standards and/or QC samples, 
• 
internal standard response significantly different from the response for the calibration standard 
and QC samples, if such criteria have been pre-defined in a SOP, 
• 
improper sample injection or malfunction of equipment, 
• 
the obtained concentration is above the ULOQ or below the run’s LLOQ, in runs where the 
lowest standard sample has been rejected from a calibration curve, resulting in a higher LLOQ 
compared with other run’s,  
• 
identification of sample analyte in pre-dose samples or placebo sample, 
• 
poor chromatography. 
Normally reanalysis of study samples because of a pharmacokinetic reason is not acceptable. This is 
especially important for bioequivalence studies, as this may affect and bias the outcome of such a 
study. However reanalysis might be considered as part of laboratory investigations, to identify 
possible reasons for results considered as abnormal and to prevent the recurrence of similar problems 
in the future.  
 
13/17 
 
460 
461 
462 
463 
465 
466 
468 
469 
470 
471 
472 
473 
474 
475 
476 
477 
478 
479 
480 
481 
482 
483 
484 
485 
486 
488 
489 
490 
491 
492 
493 
494 
495 
496 
497 
498 
499 
500 
Re-injection of samples can be made in case of instrument failure if reinjection reproducibility and on-
injector stability have been demonstrated during validation. Re-injection of a full analytical run or of 
individual calibration standard samples or QC samples, simply because the calibration or QCs failed, 
without any identified analytical cause, is not acceptable.  
5.5 
Integration 
464 
Chromatogram integration should be described in a SOP. Any deviation from this SOP should be 
discussed in the analytical report. 
6. 
INCURRED SAMPLES REANALYSIS 
467 
The use of calibration standards and QC samples during validation may not mimic the actual study 
samples. Differences for instance in protein binding, back-conversion of known and unknown 
metabolites, sample inhomogeneity or concomitant medications, may affect the accuracy and precision 
of the analyte in such samples during processing and storage. It is therefore recommended to evaluate 
accuracy of incurred samples by reanalysis of study samples over a certain time period. The extent of 
testing depends on the analyte and the study samples, and should be based upon in-depth 
understanding of the analytical method and analyte. It should at least provide sufficient confidence 
that the concentration being reported is accurate. In case incurred sample analysis showed deviating 
results, this should be investigated, and adequate steps should be taken to minimize inaccuracy (and 
imprecision).  
If pharmacokinetic parameters represent the primary endpoints of a study, incurred sample analysis is 
recommended. In toxico-kinetic studies it is sufficient to address this issue once per species. For 
(human) study samples evaluation of incurred samples should be carried out for every subject or 
patient population, unless otherwise justified. For bioequivalence studies analysis of incurred samples 
should always be carried out. Incurred sample analysis should be evaluated as early as possible. It is 
recommended that study samples are obtained from several subjects close to the expected maximal 
concentration and in the elimination phase. Samples should not be pooled, as pooling may limit 
anomalous findings. The difference between the two values obtained should be within 20% of the 
mean (30% for ligand-binding assays) for at least 67% of the repeats.  
7. 
STUDY REPORT 
487 
The study director should sign and date the final report to indicate acceptance of responsibility for the 
validity of the data and to indicate the extent to which the study is complies with the principles of 
good laboratory practice.  
The validation report should include complete documentation of the protocol, conduct and evaluation 
of the analysis, in accordance with the principles of GLP-rules and in compliance with EU and ICH 
guidelines. Information regarding conducted audits/inspection should be included in the report.  
SOP’s for relevant analysis specific procedures should be appended to the study report. 
All individual data should be available in electronic format to be provided upon request. 
Any deviation from the analytical protocol should also be discussed in the analytical report. 
The validation report should include at least the following information: 
• 
summary of the validation report in a table format, 
• 
summary of the applied analytical method and where appropriate, the source of the analytical 
method (references from literature and/or modifications in the procedure), 
 
14/17 
 
501 
502 
503 
504 
505 
506 
507 
508 
509 
510 
511 
512 
513 
514 
515 
516 
517 
518 
519 
520 
521 
522 
523 
524 
525 
526 
527 
528 
529 
530 
531 
532 
533 
534 
535 
536 
537 
538 
539 
540 
541 
542 
• 
summary of the assay procedure (analyte, IS, details of the sample pre-treatment, extraction 
and analysis), 
• 
reference standards (origin, batch, certificate of analysis, stability), 
• 
calibration standards and QC samples (matrix including anti-coagulant, preparation, 
preparation dates, and storage conditions), 
• 
run acceptance criteria,  
• 
sample tracking (conditions and duration of storage), 
• 
analysis: 
� 
table of all analytical runs with analysis dates, 
� 
table of calibration results of all analytical runs, including calibration range, response 
function, back-calculated concentrations, within- and between-run precision and 
accuracy,  
� 
table of QC results of all analytical runs (within- and between-run precision and 
accuracy), 
� 
stability data of stock solution, working solution, QC, covering the applied storage 
conditions,  
� 
data on selectivity, LLOQ, carry-over, matrix effect, dilution integrity;  
• 
deviating results obtained during validation with full justification of the action taken, 
• 
deviations from method and/or SOPs (description of deviations, impact on study, supportive 
data). 
All measurements with the individual calculated concentrations have to be presented in the report. 
Furthermore a specific detailed description of the analysis of the study samples should be written as a 
separate report and should include at least: 
• 
reference standards (origin, batch, certificate of analysis, stability, storage conditions) 
• 
calibration standards and QC samples (storage conditions) 
• 
run acceptance criteria (short description, method reference) 
• 
assay procedure (short description) 
• 
sample tracking (dates of receipt and contents, sample conditions on receipt, storage location 
and conditions, if applicable) 
• 
analysis: 
� 
set up of the analytical run, 
� 
table of all analytical runs with analysis dates and results, indicating which samples 
have been analysed in which analytical run, 
� 
table of calibration results of all (passed) analytical runs; values outside acceptance 
criteria should be clearly marked, 
� 
table of QC results of all (passed) analytical runs; values outside acceptance 
criteria should be clearly marked; 
• failed analytical runs (identity, assay date, reason for failure), 
• deviations from method and/or SOPs (description of deviations, impact on study, supportive 
data), 
• re-assay (table of sample identification, reason for re-assay, original and re-assay values), 
• incurred sample analysis, 
 
15/17 
 
543 
544 
• all chromatograms of 5 – 20% of the subjects, including the corresponding QC samples and 
calibration standards; for bioequivalence studies at least of 20% of the subjects.  
 
16/17 
 
545 
547 
548 
550 
551 
553 
554 
556 
557 
558 
559 
561 
562 
564 
566 
567 
569 
570 
572 
573 
575 
576 
578 
579 
581 
582 
584 
585 
DEFINITIONS 
1. 
Anchor calibrators 
546 
Anchor calibrators are standards points outside of the range of quantitation, used to assist in fitting the 
non linear regression of the standard curve in ligand-binding assays. 
2. 
Accuracy 
549 
The accuracy of an analytical procedure expresses the closeness of the determined value to the value 
which is accepted either as a conventional true value or an accepted reference value. 
3. 
Analytical Procedure 
552 
The analytical procedure refers to the way of performing the analysis. It should describe in detail the 
steps necessary to perform each analysis.  
4. 
Calibration range 
555 
The range of an analytical procedure is the interval between the upper and lower concentration 
(amounts) of analyte in the sample (including these concentrations) for which it has been 
demonstrated that the analytical procedure meets requirements for precision, accuracy and response 
function. 
5. 
Carry over 
560 
Carry-over is the appearance of an analyte signal in blank sample peaks after the analysis of samples 
with a high analyte concentration. 
6. 
Incurred samples 
563 
Study samples from dosed subjects or animals. 
7. 
Incurred sample reanalysis 
565 
The analysis of a portion of the incurred samples to determine whether the original analytical results 
are reproducible. 
8. 
Lower limit of quantitation (LLOQ) 
568 
The lower quantitation limit of an individual analytical procedure is the lowest amount of analyte in a 
sample which can be quantitatively determined with pre-defined precision and accuracy. 
9. 
Matrix effect 
571 
The direct or indirect alteration or interference in response due to the presence of unintended analytes 
(for analysis) or other interfering substances in the sample. 
10. Precision 
574 
The precision of an analytical procedure expresses the closeness of agreement (degree of scatter) 
between a series of measurements obtained under the prescribed conditions. 
11. Quality control (QC) sample 
577 
A spiked sample used to monitor the performance of a bioanalytical method and to assess the integrity 
and validity of the results of the unknown samples analyzed in an individual batch. 
12. Response function 
580 
The response function of an analytical procedure is its ability (within a given range) to obtain test 
results which are directly proportional to the concentration (amount) of analyte in the sample. 
13. Selectivity 
583 
Selectivity is the ability of the bioanalytical method to measure and differentiate the analyte in the 
presence of components which may be expected to be present. 
 
17/17 
 
587 
588 
590 
591 
14. Specificity 
586 
For ligand binding assays, specificity is the ability to measure the analyte unequivocally in the 
presence of other compounds, either exogenous or endogenous, in the matrix. 
15. Upper limit of quantitation (ULOQ) 
589 
The upper quantitation limit of an individual analytical procedure is the highest amount of analyte in a 
sample which can be quantitatively determined with pre-defined precision and accuracy.  
