Official address  Domenico Scarlattilaan 6  ●  1083 HS Amsterdam  ●  The Netherlands
An agency of the European Union
Address for visits and deliveries  Refer to www.ema.europa.eu/how-to-find-us 
Send us a question Go to www.ema.europa.eu/contact
Telephone +31 (0)88 781 6000
 
© European Medicines Agency, 2021. Reproduction is authorised provided the source is acknowledged.
31 March 2022
EMA/CHMP/ICH/195040/2022
Committee for Medicinal Products for Human Use 
ICH guideline Q14 on analytical procedure development
Step 2b 
Transmission to CHMP 
8 March 2022
Adoption by CHMP 
24 March 2022
Release for public consultation 
31 March 2022
Deadline for comments 
31 July 2022
Comments should be provided using this template. The completed comments form should be 
sent to ich@ema.europa.eu 
 
 
 
 
 
INTERNATIONAL COUNCIL FOR HARMONISATION OF TECHNICAL 
REQUIREMENTS FOR PHARMACEUTICALS FOR HUMAN USE 
 
 
ICH HARMONISED GUIDELINE 
 
 
ANALYTICAL PROCEDURE DEVELOPMENT 
Q14 
 
 
Draft version  
Endorsed on 24 March 2022 
Currently under public consultation 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
At Step 2 of the ICH Process, a consensus draft text or guideline, agreed by the 
appropriate ICH Expert Working Group, is transmitted by the ICH Assembly to the 
regulatory authorities of the ICH regions for internal and external consultation, 
according to national or regional procedures. 
 
 
 
 
 
Q14 
Document History 
 
Code 
History 
Date 
Q14 
Endorsement by the Members of the ICH Assembly 
under Step 2 and release for public consultation. 
24 March 2022 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Legal notice: This document is protected by copyright and may, with the exception of the ICH 
logo, be used, reproduced, incorporated into other works, adapted, modified, translated or 
distributed under a public license provided that ICH's copyright in the document is acknowledged 
at all times.  In case of any adaption, modification or translation of the document, reasonable 
steps must be taken to clearly label, demarcate or otherwise identify that changes were made to 
or based on the original document.  Any impression that the adaption, modification or translation 
of the original document is endorsed or sponsored by the ICH must be avoided. 
The document is provided "as is" without warranty of any kind.  In no event shall the ICH or the 
authors of the original document be liable for any claim, damages or other liability arising from 
the use of the document. 
The above-mentioned permissions do not apply to content supplied by third parties.  Therefore, 
for documents where the copyright vests in a third party, permission for reproduction must be 
obtained from this copyright holder. 
 
i 
ICH HARMONISED GUIDELINE  
ANALYTICAL PROCEDURE DEVELOPMENT 
Q14  
ICH Consensus Guideline 
 
TABLE OF CONTENTS 
 
1. 
INTRODUCTION .................................................................................................................. 2 
1.1 
Objective of the Guideline ....................................................................................................... 2 
2. 
SCOPE .................................................................................................................................... 2 
2.1 
General Considerations for Analytical Procedure Development and Lifecycle Management . 2 
2.2 
Minimal versus Enhanced Approaches to Analytical Procedure Development ....................... 3 
2.3 
The Analytical Procedure Lifecycle ......................................................................................... 4 
3. 
ANALYTICAL TARGET PROFILE (ATP) ........................................................................ 4 
4. 
KNOWLEDGE AND RISK MANAGEMENT IN ANALYTICAL PROCEDURE 
DEVELOPMENT AND CONTINUAL IMPROVEMENT........................................................... 5 
4.1 
Knowledge Management ......................................................................................................... 5 
4.2 
Risk Management .................................................................................................................... 5 
5. 
EVALUATION OF ROBUSTNESS AND PARAMETER RANGES OF ANALYTICAL 
PROCEDURES ................................................................................................................................. 6 
5.1 
Robustness ............................................................................................................................... 6 
5.2 
Analytical Procedure Parameter Ranges .................................................................................. 6 
6. 
ANALYTICAL PROCEDURE CONTROL STRATEGY ................................................. 7 
6.1 
Established Conditions for Analytical Procedures ................................................................... 8 
7. 
LIFECYCLE MANAGEMENT AND POST-APPROVAL CHANGES OF ANALYTICAL 
PROCEDURES ................................................................................................................................. 8 
8. 
DEVELOPMENT OF MULTIVARIATE ANALYTICAL PROCEDURES ................... 12 
9. 
DEVELOPMENT OF ANALYTICAL PROCEDURES FOR REAL TIME RELEASE 
TESTING: SPECIAL CONSIDERATONS .................................................................................. 16 
10. 
SUBMISSION OF ANALYTICAL PROCEDURE RELATED INFORMATION ........ 16 
10.1 General Regulatory Considerations and Documentation ....................................................... 16 
10.2 Documentation for the Enhanced Approach .......................................................................... 17 
10.3 Documentation for Multivariate Analytical Procedures and RTRT ....................................... 17 
11. 
GLOSSARY .......................................................................................................................... 19 
12. 
REFERENCES ..................................................................................................................... 26 
13. 
ANNEX ................................................................................................................................. 26 
ICH Q14 Guideline 
 
1 
 
13.1 Annex A – Analytical Procedure Lifecycle ............................................................................ 26 
13.1.1 Measurement of Stereoisomers as Specific Process Related Impurities in a Small 
Molecule Drug Substance (DS).............................................................................................. 28 
13.1.2 Measurement of Potency for an anti-TNF-alpha Monoclonal Antibody .................... 37 
13.2 Annex B: Validation Strategies for MODRs .......................................................................... 59 
13.3 Annex C: Example of Multivariate Model Lifecycle Components ....................................... 61 
 
 
 
ICH Q14 Guideline 
 
2 
 
1. INTRODUCTION 
1 
1.1 Objective of the Guideline 
2 
This guideline describes science and risk-based approaches for developing and maintaining analytical 
3 
procedures suitable for the assessment of the quality of drug substances and drug products. The 
4 
systematic approach suggested in ICH Q8 Pharmaceutical Development together with principles of 
5 
ICH Q9 Quality Risk Management can also be applied to the development and lifecycle management 
6 
of analytical procedures. When developing an analytical procedure, a minimal (also known as 
7 
traditional) approach or elements of an enhanced approach can be applied.  
8 
Furthermore, the guideline describes considerations for the development of multivariate analytical 
9 
procedures and for real time release testing (RTRT). 
10 
This guideline is intended to complement ICH Q2 Validation of Analytical Procedures. Submitting 
11 
knowledge and information related to development of analytical procedures to regulatory agencies 
12 
may provide additional evidence to demonstrate that the analytical procedure is appropriate for its 
13 
intended purpose. 
14 
Using the tools described in ICH Q12 Technical and Regulatory Considerations for Pharmaceutical 
15 
Product Lifecycle Management, the guideline describes principles to support change management of 
16 
analytical procedures based on risk management, comprehensive understanding of the analytical 
17 
procedure and adherence to predefined criteria for performance characteristics. Knowledge gained 
18 
from application of an enhanced approach to analytical procedure development can provide better 
19 
assurance of the performance of the procedure, can serve as a basis for the analytical procedure 
20 
control strategy and can provide an opportunity for more efficient regulatory approaches to related 
21 
post approval changes. 
22 
The guideline also describes submission of analytical procedure development and related lifecycle 
23 
information in the Common Technical Document (CTD) format (ICH M4Q, The Common Technical 
24 
Document for the Registration of Pharmaceuticals for Human Use: Quality – M4Q). 
25 
 
26 
2. SCOPE 
27 
This guideline applies to new or revised analytical procedures used for release and stability testing of 
28 
commercial drug substances and products (chemical and biological/biotechnological). The guideline 
29 
can also be applied to other analytical procedures used as part of the control strategy (ICH Q10, 
30 
Pharmaceutical Quality System) following a risk-based approach. The scientific principles described 
31 
in this guideline can be applied in a phase-appropriate manner during clinical development. This 
32 
guideline may also be applicable to other types of products, with appropriate regulatory authority 
33 
consultation as needed. Development of pharmacopoeial analytical procedures is out of scope. 
34 
 
35 
2.1 General Considerations for Analytical Procedure Development and Lifecycle Management 
36 
The goal of development is to obtain an analytical procedure fit for its intended purpose: to measure 
37 
an attribute or attributes of the analysed material with the needed specificity/selectivity, accuracy 
38 
and/or precision over the reportable range. 
39 
In this section the minimal and enhanced approaches to analytical procedure development are 
40 
described. While the minimal approach remains acceptable, some or all elements of the enhanced 
41 
approach might be used to support development and lifecycle management of analytical procedures. 
42 
ICH Q14 Guideline 
 
3 
 
In certain cases, an established analytical procedure can be applied to multiple products with little or 
43 
no modification of measurement conditions. For a new application of such platform analytical 
44 
procedures, the subsequent development can be abbreviated, and certain validation tests can be 
45 
omitted based on a science- and risk-based justification. Details of the performance characteristics 
46 
considered for analytical procedure validation are described in ICH Q2. 
47 
In general, data gained during the development studies (e.g., robustness data from a design of 
48 
experiments (DoE study)) can be used as validation data for the related analytical procedure 
49 
performance characteristics and does not necessarily need to be repeated. 
50 
 
51 
2.2 Minimal versus Enhanced Approaches to Analytical Procedure Development  
52 
 
53 
Minimal Approach  
54 
Analytical procedure development should include the following elements as appropriate: 
55 
• Identifying which attributes of the drug substance or drug product need to be tested by the 
56 
analytical procedure. 
57 
• Selecting an appropriate analytical procedure technology and related instruments or suitable 
58 
apparatus. 
59 
• Conducting appropriate development studies to evaluate analytical procedure performance 
60 
characteristics such as specificity, accuracy and precision over the reportable range (including 
61 
the calibration model, limits at lower and/or higher range ends) and robustness. 
62 
• Defining an appropriate analytical procedure description including the analytical procedure 
63 
control strategy (e.g., parameter settings and system suitability). 
64 
 
65 
Enhanced Approach 
66 
The enhanced approach offers a systematic way of developing and refining knowledge of an 
67 
analytical procedure. An enhanced approach should include one or more of the following elements in 
68 
addition to those already described for the minimal approach:  
69 
• An evaluation of the sample properties and the expected variability of the sample based on 
70 
manufacturing process understanding. 
71 
• Defining the analytical target profile (ATP). 
72 
• Conducting risk assessment and evaluating prior knowledge to identify the analytical 
73 
procedure parameters that can impact performance of the procedure. 
74 
• Conducting uni- or multi-variate experiments to explore ranges and interactions between 
75 
identified analytical procedure parameters. 
76 
• Defining an analytical procedure control strategy based on enhanced procedure understanding 
77 
including appropriate set-points and/or ranges for relevant analytical procedure parameters 
78 
ensuring adherence to performance criteria. 
79 
• Defining a lifecycle change management plan with clear definitions and reporting categories 
80 
of established conditions (ECs), proven acceptable ranges (PARs) or method operational 
81 
design regions (MODRs) as appropriate. 
82 
 
83 
Applying elements of the enhanced approach to development can lead to more robust analytical 
84 
procedures, better understanding of the impact of analytical procedure parameters and more flexibility 
85 
for lifecycle management such as wider operating ranges, a more appropriate set of ECs and 
86 
associated reporting categories for changes. 
87 
 
88 
ICH Q14 Guideline 
 
4 
 
The enhanced approach potentially offers several advantages, including: 
89 
• 
Understanding of which analytical procedure attributes are essential to procedure 
90 
performance (i.e., ECs). 
91 
• Employing predefined performance characteristics (e.g., in the ATP) linked to critical quality 
92 
attributes (CQAs) and their acceptance criteria to provide purpose driven protocols for 
93 
validation of analytical procedures and for future comparisons between current and new 
94 
analytical procedures/technologies. 
95 
• 
Improving analytical procedure control resulting in more reliable operation.  
96 
• 
Enabling preventative measures and facilitating continual improvement by using more 
97 
analytical procedure knowledge. 
98 
• 
Reducing the amount of effort across the analytical procedure lifecycle. 
99 
 
100 
2.3 The Analytical Procedure Lifecycle 
101 
Figure 1 depicts elements of the analytical procedure lifecycle. Analytical procedure development 
102 
and change management approaches are described in this guideline whereas analytical procedure 
103 
validation is described in ICH Q2. Depending on the intended use of the analytical procedure and the 
104 
development approach taken, the order and extent of each element could vary, and several elements 
105 
could occur simultaneously. 
106 
Figure 1: The Analytical Procedure Lifecycle 
107 
 
108 
Control strategy –
Attributes 
Requiring Testing
Analytical Target 
Profile (ATP)
Technology 
Selection
Risk 
Assessment
Identification of 
Parameter Set-Points 
and/or Ranges
Definition of 
Analytical Procedure 
Control Strategy
Validation
Routine Use and On-
going Monitoring
Continual 
Improvement
Changes to 
Product and/or 
Control Strategy –
Attributes 
Requiring Testing
Change
Product
and Process 
Understanding
Prior Knowledge
Analytical Procedure Lifecycle
Change Management
Analytical Procedure Development
Analytical 
Procedure 
Validation
 
109 
 
110 
3. ANALYTICAL TARGET PROFILE (ATP) 
111 
Product and process understanding (ICH Q8 and ICH Q11 Development and Manufacture of Drug 
112 
Substances) leads to the identification of quality attributes requiring analytical measurement for 
113 
control which are described (for example) in a quality target product profile (QTPP). Measurement 
114 
needs can be captured in an ATP which forms the basis for development of the analytical procedure. 
115 
ICH Q14 Guideline 
 
5 
 
An ATP consists of a description of the intended purpose, appropriate details on the product attributes 
116 
to be measured and relevant performance characteristics with associated performance criteria. The 
117 
ATP includes the performance requirements for a single attribute or a set of quality attributes. The 
118 
ATP drives the choice of analytical technology. Multiple available analytical techniques may meet 
119 
the performance requirements. Consideration of the operating environment (e.g., at-line, in-line or 
120 
off-line) should be included in the technology selection. Once a technology has been selected, the 
121 
ATP serves as a foundation to derive the appropriate analytical procedure attributes and acceptance 
122 
criteria for analytical procedure validation (ICH Q2). Formal documentation and submission of an 
123 
ATP is optional but can facilitate regulatory communication irrespective of the chosen development 
124 
approach. 
125 
The ATP facilitates the selection of the technology, the procedure design and development as well as 
126 
the subsequent performance monitoring and continual improvement of the analytical procedure. The 
127 
ATP is maintained over the lifecycle and can also be used as a basis for lifecycle management to 
128 
ensure that the analytical procedure remains suitable for the intended use. 
129 
Illustrative examples of ATPs are provided in Annex A. 
130 
 
131 
4. KNOWLEDGE AND RISK MANAGEMENT IN ANALYTICAL PROCEDURE 
132 
DEVELOPMENT AND CONTINUAL IMPROVEMENT 
133 
4.1 Knowledge Management  
134 
As with product and manufacturing process development (ICH Q10), knowledge management plays 
135 
a critical role in analytical procedure development and during the subsequent lifecycle of the 
136 
analytical procedure. 
137 
Prior knowledge is explicitly or implicitly used for informing decisions during analytical procedure 
138 
development and lifecycle management. Prior knowledge can be internal knowledge from a 
139 
company’s proprietary development and analytical experience, external knowledge such as reference 
140 
to scientific and technical publications or established scientific principles. 
141 
Prior product knowledge plays an important role in identifying the appropriate analytical technique. 
142 
Knowledge of best practices and current state-of-the-art technologies as well as current regulatory 
143 
expectations contributes to the selection of the most suitable technology for a given purpose. Existing 
144 
platform analytical procedures (e.g., protein content determination by UV spectroscopy for a protein 
145 
drug) can be leveraged to evaluate the attributes of a specific product without conducting additional 
146 
procedure development. 
147 
As additional information is obtained, knowledge related to analytical procedures should be actively 
148 
managed throughout the product lifecycle. 
149 
 
150 
4.2 Risk Management 
151 
The use of quality risk management is encouraged to aid in the development of a robust analytical 
152 
procedure to reduce risk of poor performance and reporting incorrect results. Risk assessment is 
153 
typically performed early in analytical procedure development and is repeated as more information 
154 
becomes available. Risk assessment can be formal or informal and can be supported by prior 
155 
knowledge. 
156 
Risk assessment tools as described in ICH Q9 Annex 1 can be used to  
157 
ICH Q14 Guideline 
 
6 
 
• identify analytical procedure parameters (factors and operational steps) with potential impact 
158 
on its performance, e.g., Annex A Figures 1 and 2 (Ishikawa diagrams).  
159 
• assess the potential impact of analytical procedure parameters on the analytical procedure 
160 
performance.  
161 
• identify and prioritise analytical parameters to be investigated experimentally. 
162 
 
163 
Risk control principles can be used to establish an analytical procedure control strategy. To maintain 
164 
a state of control for analytical procedure performance, ongoing monitoring is recommended as part 
165 
of risk review.  
166 
Risk communication should be used to support continual improvement of the analytical procedure 
167 
performance throughout its lifecycle. The outcome of quality risk management should be documented 
168 
within the applicant’s pharmaceutical quality system (PQS).   
169 
 
170 
5. EVALUATION OF ROBUSTNESS AND PARAMETER RANGES OF ANALYTICAL 
171 
PROCEDURES 
172 
5.1 Robustness 
173 
The robustness of an analytical procedure is a measure of its capacity to meet the expected 
174 
performance requirements during normal use. Robustness is tested by deliberate variations of 
175 
analytical procedure parameters. Prior knowledge and risk assessment can inform the selection of 
176 
parameters to investigate during the robustness study. Those parameters likely to influence procedure 
177 
performance over the intended period of use should be studied.  
178 
 
179 
For most procedures, robustness evaluation is conducted during development. If the evaluation of 
180 
robustness was already conducted during development, it does not need to be repeated during 
181 
validation as discussed in ICH Q2. Data from validation studies (e.g., intermediate precision) can be 
182 
used to complement robustness evaluation. For some analytical procedures with inherent high 
183 
parameter variability (e.g., those requiring biological reagents) wider ranges may need to be 
184 
investigated during robustness studies. Robustness of multivariate procedures may require additional 
185 
considerations (see chapter 8). The outcome of the evaluation of robustness should be reflected in the 
186 
analytical procedure control strategy. 
187 
 
188 
5.2 Analytical Procedure Parameter Ranges 
189 
Experiments to investigate parameter ranges can provide additional knowledge about the analytical 
190 
procedure performance. The respective analytical procedure attributes and associated criteria could 
191 
be derived from the ATP. Univariate examination of a single parameter can establish proven 
192 
acceptable ranges (PAR) for the analytical procedure. 
193 
In an enhanced approach, the ranges for the relevant parameters and their interactions can be 
194 
investigated in multi-variate experiments (DoE). Risk assessment and prior knowledge should be 
195 
used to identify parameters, attributes and appropriate associated ranges to be investigated 
196 
experimentally. Categorical variables (e.g., different instruments) can also be considered as part of 
197 
the experimental design.  
198 
The outcome of development studies including DoE can provide an understanding of the relationships 
199 
between analytical procedure variables (inputs) and the responses of the analytical procedure 
200 
(outputs). Based on the results, fixed set-points may be defined for some parameters. For others, 
201 
PARs could be defined while still others could be included into an MODR. An MODR consists of 
202 
ICH Q14 Guideline 
 
7 
 
combined ranges for two or more variables within which the analytical procedure is shown to be fit 
203 
for the intended use. 
204 
Parameter ranges (e.g., PAR or MODR) can be proposed by the applicant based on development data 
205 
and are subject to regulatory approval. Moving within an established parameter range does not require 
206 
regulatory notification. 
207 
For practical reasons and following a risk-based approach, it may not be necessary or possible to 
208 
validate the entirety of a MODR. The part of a PAR or a MODR intended for routine use in the 
209 
analytical procedure must be covered by validation data. Validation approaches for MODRs are 
210 
described in Annex B including an example table to present the performance characteristics combined 
211 
with the analytical procedure attribute acceptance criteria, parameter ranges, analytical procedure 
212 
control strategy and validation strategy. Analytical procedure validation is required only for those 
213 
performance characteristics not covered by data from analytical procedure development. An 
214 
analytical procedure validation strategy, e.g., as part of the analytical procedure validation protocol, 
215 
can define the necessary extent of additional validation. 
216 
 
217 
6. ANALYTICAL PROCEDURE CONTROL STRATEGY 
218 
An analytical procedure control strategy should ensure that the analytical procedure performs as 
219 
expected during routine use throughout its lifecycle and consists of a set of controls, derived from 
220 
current understanding of the analytical procedure including development data, risk assessment and 
221 
robustness. Prior knowledge could also be used to develop the analytical procedure control strategy. 
222 
The analytical procedure control strategy should be defined before validation (ICH Q2) and should 
223 
be confirmed after validation has been finalized. 
224 
The analytical procedure control strategy includes analytical procedure parameters needing control 
225 
and the system suitability test (SST) which is part of the analytical procedure description. The 
226 
analytical procedure description should include the steps necessary to perform each analytical test. 
227 
This can include (but is not limited to) the sample, the reference materials and the reagents, sample 
228 
and control preparations, use of the apparatus, generation of the calibration curve, use of the formulae 
229 
for the calculation of the reportable results and other necessary steps. The level of detail should 
230 
enable a skilled analyst to perform the analysis and interpret the results (such as the level of detail in 
231 
a regional pharmacopoeia for a similar substance). 
232 
The SST depends on the type and intent of the analytical procedure and is typically conducted with 
233 
one or more predefined materials (including use of positive or negative controls). The SST is designed 
234 
to verify selected analytical procedure attributes. The acceptance criteria should be based on 
235 
analytical procedure performance criteria. The components of the SST should be selected using risk 
236 
assessment as well as knowledge and understanding from development data. The test is used to verify 
237 
that the measurement system and the analytical operations associated with the analytical procedure 
238 
are adequate during the intended time period of analysis and enable the detection of potential failures. 
239 
Validity of the results of the analytical procedure depends on the outcome of the SST. In the enhanced 
240 
approach, a well-designed set of SST parameters and criteria to ensure method performance could 
241 
represent an important aspect of risk mitigation. For analytical procedures relying on multivariate 
242 
models, data quality should be verified using appropriate software tools.  
243 
In addition to SST, sample suitability assessment may be required to ensure acceptable sample 
244 
response. A sample and/or sample preparation is considered suitable if the measurement response of 
245 
the sample satisfies pre-defined acceptance criteria for the analytical procedure attributes that have 
246 
been developed for the validated analytical procedure (often used for biologics). In these cases, 
247 
sample suitability is a prerequisite for the validity of the result along with a satisfactory outcome of 
248 
ICH Q14 Guideline 
 
8 
 
the SST. For analytical procedures relying on multivariate models, sample suitability assessment can 
249 
be verified using appropriate software tools which check if the sample fits within the model space. 
250 
This is commonly called data quality check. 
251 
 
252 
Ongoing monitoring of selected analytical procedure outputs is recommended to look for any trends, 
253 
in line with PQS expectations. Review of analytical procedure outputs facilitates the procedure 
254 
lifecycle management and enables proactive intervention to avoid failures.  
255 
 
256 
6.1 Established Conditions for Analytical Procedures 
257 
In line with ICH Q12, applicants may define established conditions (ECs) for an analytical procedure. 
258 
ECs are proposed and justified by the applicant and approved by the regulatory authorities. ECs can 
259 
be identified using tools highlighted in Chapter 4 including risk assessment, prior knowledge, and 
260 
learnings from uni- and/or multi-variate experimentation. The nature and extent of ECs will depend 
261 
on the development approach, the complexity of the analytical procedure and a demonstrated 
262 
understanding of how parameters and other factors impact its performance.  
263 
With a minimal approach to development, the number of ECs may be extensive with fixed analytical 
264 
procedure parameters and set points.  
265 
With an enhanced approach to development, there should be an increased understanding of the 
266 
relationship between analytical procedure parameters and performance to facilitate identification of 
267 
which factors require control and thus enable a more appropriate set of ECs. These can focus on 
268 
performance characteristics (e.g., specificity, accuracy, precision). 
269 
 
270 
ECs could consist of performance criteria (e.g., in the ATP or as part of SST), the analytical procedure 
271 
principle (i.e., the physicochemical basis or specific technology), and set points and/or ranges for one 
272 
or more parameters. Analytical procedure parameters which need to be controlled to ensure the 
273 
performance of the procedure as well as those where the need for control cannot be reasonably 
274 
excluded should be identified as ECs. If a parameter is controlled through performance characteristics 
275 
and criteria, that parameter may not necessarily need to be defined as an EC or may be assigned a 
276 
lower reporting category. 
277 
Use of the enhanced approach should not lead to providing a less detailed description of analytical 
278 
procedures in a regulatory submission. A suitably detailed description of the analytical procedures in 
279 
Module 3 of the CTD is expected to provide a clear understanding regardless of the approach used to 
280 
identify ECs for analytical procedures. Description of analytical procedures can include supportive 
281 
information as well as identified ECs. 
282 
Identification of reporting categories for ECs and the utilization of ECs in change management are 
283 
described in the next chapter. 
284 
 
285 
7. LIFECYCLE MANAGEMENT AND POST-APPROVAL CHANGES OF ANALYTICAL 
286 
PROCEDURES 
287 
Changes to analytical procedures can occur throughout the product lifecycle and could involve 
288 
modification of existing procedures or a complete replacement including introduction of a new 
289 
technology. Major changes in the performance characteristics or additional information on attributes 
290 
could, in certain instances, lead to reevaluation of the ATP itself and/or a new procedure. Typically, 
291 
process knowledge, analytical procedure knowledge and continual improvement are drivers for 
292 
ICH Q14 Guideline 
 
9 
 
change. If possible, changes should lead to improved analytical procedures in line with best practices 
293 
and instrumentation. The tools and enablers discussed in ICH Q12 are applicable to analytical 
294 
procedures, irrespective of the development approach and consist of: 
295 
• Existing risk-based categorisation of changes to analytical procedures (in applicable regional 
296 
regulatory framework)  
297 
• ECs  
298 
• Post-Approval Change Management Protocols (PACMPs) which provide a detailed 
299 
explanation of how future changes will be managed and provide the marketing authorization 
300 
holder (MAH) with certainty about the acceptability of future changes and an associated 
301 
reduced reporting category. 
302 
• The Product Lifecycle Change Management (PLCM) document which can facilitate 
303 
regulatory communication about likely post-approval changes. 
304 
• The PQS (documentation of all changes including those not requiring regulatory submission, 
305 
e.g., within a MODR or for parameters deemed not to have an impact on the method 
306 
performance) 
307 
• The structured approach to frequent CMC changes (ICH Q12 Chapter 8). 
308 
 
309 
If a minimal approach to development is taken, then any changes should be reported according to 
310 
existing regional reporting requirements. The use of different elements of the enhanced approach can 
311 
facilitate management and regulatory communication of post-approval changes. 
312 
If appropriately justified and validated (see Chapter 5.2), a PAR or MODR allows flexibility within 
313 
the approved range(s) to be managed within a company’s PQS. Changes outside of the approved 
314 
ranges or expansion of said ranges require regulatory reporting. 
315 
In cases where ECs are proposed, the risk associated with prospective changes should be assessed up 
316 
front to define the appropriate reporting category. Factors to consider include the importance of the 
317 
quality attribute being measured, the complexity of the technology and the extent of the change. 
318 
Relevant risk reduction measures should be identified based on product and process knowledge as 
319 
well as analytical procedure understanding and the proposed analytical procedure control strategy. 
320 
Finally, the level of risk (high, medium or low) should be assigned. 
321 
In general, an understanding of the analytical procedure robustness and/or prior knowledge can be 
322 
used to support risk mitigation associated with future changes. Submitting the outcomes of the risk 
323 
assessments to regulatory agencies when ECs are identified can help justify reporting categories for 
324 
future changes to analytical procedures. 
325 
Figure 2 summarizes how risk assessment and risk reduction measures can help identify appropriate 
326 
reporting categories for ECs. Fixing performance criteria for performance characteristics identified 
327 
as ECs, for example, in an ATP, can help mitigate risk associated with changes. This ensures that the 
328 
analytical procedure remains fit for purpose subsequent to changes and thus forms the basis of a 
329 
bridging strategy. Changes to parameters that are not ECs should be documented in the PQS but do 
330 
not require regulatory reporting. 
331 
The ATP could also form the basis of a PACMP which would allow changes (e.g., a change between 
332 
technologies) to be reported at a lower reporting category provided that the pre-defined requirements 
333 
for a change are met.  
334 
 
 
335 
ICH Q14 Guideline 
 
10 
 
 
336 
Figure 2: Risk-based approach to identification of ECs and reporting categories for associated 
337 
changes in the enhanced approach  
338 
 
339 
Is the factor proposed as an EC?
Yes
Considering product and procedure 
knowledge and understanding*, what is the 
risk associated with the prospective 
change(s) to the analytical procedure?
Are criteria of  relevant 
performance characteristics 
defined as ECs which ensure 
the post-change quality of the 
measured result after the 
change?
Reporting category as 
notification moderate***
Reporting category as 
notification low
Reporting category 
as prior approval
Manage changes within PQS
Low
High
Medium
Are criteria of  relevant 
performance characteristics 
defined as ECs which ensure 
the post-change quality of the 
measured result after the 
change?
Yes**
Yes**
No
No
No
Perform Risk Assessment/Development 
Studies to inform the selection of ECs
 
340 
 
*     Including analytical procedure control strategy 
341 
 
**   Sufficient information or prior knowledge should be available to design appropriate  
342 
                   future bridging studies  
343 
 
*** In some cases, moderate risk changes proposed by the company may require prior 
344 
                   approval based on health authority feedback 
345 
 
346 
In Annex A examples are given on how appropriate reporting categories can be proposed.   
347 
When implementing changes to analytical procedures, QRM can be used to evaluate the impact of 
348 
the changes and re-confirm that the originally agreed reporting category is still appropriate. The 
349 
outcome of this risk assessment informs the design and extent of the studies needed to support the 
350 
change including an appropriate bridging strategy to demonstrate that the revised or new procedure 
351 
is fit for purpose. The implementation of an already validated analytical procedure at a different 
352 
location, including the concepts of the analytical procedure transfer, should follow the same 
353 
verification and bridging strategies (Tables 1 and 2). 
354 
 
 
355 
ICH Q14 Guideline 
 
11 
 
Table 1: Relationship between knowledge, risk and extent of studies for changes to analytical 
356 
procedures  
357 
Knowledge 
 Low                                          High  
Risk associated with the change 
Low                                                                                          High                                          
Confirmatory study according 
to previously defined protocol 
or prior knowledge 
 
In depth study according to 
previously defined protocol  
 
 
 
Confirmatory study including 
study design 
 
In depth evaluation including 
study design  
 
358 
For product and process changes, a re-assessment and potential adaptation of the ATP, if used, and a 
359 
re-assessment of the suitability of the analytical procedure may be necessary.  
360 
If an applicant proposes a new analytical procedure, a thorough risk assessment and evaluation should 
361 
be conducted to determine any impact on the performance. The analytical procedure control strategy 
362 
for the new procedure should be established. ECs associated with the new procedure should be 
363 
justified when reporting the change. 
364 
Table 2 provides examples of data recommended to support a change dependent on the extent of the 
365 
change and the identified risk category.  
366 
 
 
367 
ICH Q14 Guideline 
 
12 
 
Table 2: Examples of Analytical Procedure Change Evaluation 
368 
Risk Factor: Extent of change 
Bridging strategy 
Evidence of the suitability of 
a new procedure 
Change of analytical procedure 
principle 
(physicochemical/biochemical 
basis) 
Full validation of new procedure 
And 
Comparative analysis of 
representative samples and standards.  
And/or 
Demonstration that the analytical 
procedure’s ability to discriminate 
between acceptable and non-
acceptable results remains 
comparable 
Analytical procedure performance 
characteristics are evaluated and 
criteria are met after the change 
And 
Results are comparable after 
change or differences are 
acceptable and potential impact 
on specification evaluated 
Change within same analytical 
procedure principle, for example: 
 
1. Modification of procedures 
2. Transfer of procedures to 
different 
locations/environments 
 
Partial or full re-validation of the 
analytical procedure performance 
characteristics affected by the change 
And/or 
Comparative analysis of 
representative samples and standards 
Analytical procedure attributes 
are evaluated and criteria are met 
after change 
And/or 
Results are comparable after 
change or differences are 
acceptable and potential impact 
on specification evaluated 
 
369 
To support the use of the tools described in this guideline, the company´s PQS change management 
370 
process should be effective and in line with recommendations described in ICH Q12. During the 
371 
lifecycle the MAH should evaluate performance, perform trend analysis, assess knowledge gained 
372 
and re-evaluate if the analytical procedure remains fit for purpose.  
373 
 
374 
8. DEVELOPMENT OF MULTIVARIATE ANALYTICAL PROCEDURES 
375 
Multivariate analytical procedures are those where a result is determined through a multivariate 
376 
calibration model utilizing more than one input variable. The considerations provided here are for 
377 
models using latent variables that are mathematically related to directly measured variables. Other 
378 
approaches, in machine learning, such as neural networks, or optimization techniques could use 
379 
similar principles although the specific approach may vary and will not be discussed in detail. 
380 
Development of a robust multivariate analytical procedure includes scientifically justified sample 
381 
selection and distribution over the range, sample size, model variable selection and data pre-
382 
processing. 
383 
Sample and sample population 
384 
Multivariate models link measured model variables with values obtained from a validated reference 
385 
procedure or from reference samples. Therefore, samples in multivariate analysis consist of input 
386 
measurements and their corresponding reference values, which are numeric values for quantitative 
387 
measurements (e.g., assay) and classification categories for qualitative methods (e.g., identity). In 
388 
some cases, one set of input measurements could be used for multiple models provided that more 
389 
than one reference value exists. The reference values are determined using reference analytical 
390 
procedure(s) or prepared reference samples with known values. Care should be taken to ensure that 
391 
ICH Q14 Guideline 
 
13 
 
uncertainty in the reference analytical procedure is sufficiently low in relation to the intended 
392 
performance of the multivariate analytical procedure and that prepared reference samples are 
393 
homogeneous. The approach to the reference procedure(s) or prepared reference samples should be 
394 
explained and justified. 
395 
 
396 
The ranges of multivariate models are typically constructed by data from samples. Therefore, a 
397 
careful strategy for sample selection is essential for obtaining the relevant information from the 
398 
analytical data and contributes to the robustness of the resulting model. Based on the method and 
399 
measurement principle, the sample population should encompass the sources of variability likely to 
400 
occur during manufacture and analysis, such as raw material quality, manufacturing process 
401 
variability, storage conditions, sample preparation and testing. Use of risk assessment tools can help 
402 
to identify sources of variability with the potential to influence the measurements and resulting model 
403 
outputs. 
404 
Obtaining samples with appropriate variability at commercial scale can be challenging. Therefore, 
405 
development laboratory and pilot scale samples are often utilized to provide enough variability to 
406 
improve accuracy and robustness of the model. Inclusion of commercial scale samples is 
407 
recommended to capture variability related to specific equipment and/or processing conditions. 
408 
Careful consideration should also be given to sample distribution in the calibration and validation 
409 
sets, as this will influence the model predictive capability. 
410 
The number of samples used to create a calibration model for quantitative analysis will depend on the 
411 
complexity of the sample matrix and/or interference by the matrix in the analyte signal of interest 
412 
(i.e., for more complex sample matrices, generally more samples are needed). 
413 
Sufficient samples should be available to allow for creation of independent calibration and validation 
414 
sets of appropriate size and variability, i.e., samples in the validation set are not incorporated in 
415 
calibration or internal testing sets. A validation sample set generated with samples from independent 
416 
batches can be used to demonstrate model robustness. 
417 
Variable selection  
418 
Variable selection is performed during model development.  For example, wavelength range selection 
419 
is frequently applied in spectroscopic applications to select a region of a spectrum that gives the best 
420 
estimation of the selected chemical or physical property to be evaluated (modeled). Variable selection 
421 
depends on the measurement principle, application and other factors, and should be justified. 
422 
Data transformation  
423 
The selection of the data transformation method(s) can be driven by the type of data, instrument or 
424 
sample, the intended use of the model and/or prior knowledge. Caution should be exercised when 
425 
performing any transformation because artefacts can be introduced, or essential information lost. Any 
426 
transformation of data should be documented and justified. 
427 
Robustness 
428 
Model development should minimize the prediction error and provide a robust model that consistently 
429 
assures the long-term performance of multivariate models. The robustness should be built into the 
430 
model by including relevant sources of variability related to materials, process, environment, 
431 
instrumentation or other factors. Sources of variability can be identified from prior knowledge and 
432 
risk assessments and evaluated using statistical tools. Robustness depends on multiple factors, e.g., 
433 
composition of the calibration set, data transformation method, variable selection and the number of 
434 
latent variables. 
435 
Optimization of the multivariate model is an important step in development and often requires a trade-
436 
off between accuracy and robustness. A critical factor is the number of latent variables to be used in 
437 
the calibration model which ensures the model is optimized for its intended purpose. Selection of the 
438 
ICH Q14 Guideline 
 
14 
 
number of latent variables occurs during model development and is confirmed during internal testing. 
439 
Too many latent variables can result in model overfitting, potentially resulting in decreased 
440 
robustness and a need for more frequent model updates. Justification for the final number of latent 
441 
variables used should be provided. Diagnostic plots provided by software packages can be useful to 
442 
support the justification.  
443 
Re-calibration and model maintenance 
444 
Tracking the calibration model performance is an important part of ongoing monitoring for a 
445 
multivariate analytical procedure. Various statistical tools can be employed as diagnostics to ensure 
446 
that the model assumptions are upheld. For latent variable models, these diagnostic tools can include: 
447 
• examination of residuals to determine unmodeled features of the data (e.g., x-residuals or F-
448 
probability) 
449 
• outlier diagnostics to determine if the data is within the bounds of the model construction 
450 
(e.g., Hotelling’s T-squared or Mahalanobis distance) 
451 
Software packages allow for the application of diagnostic tools for every model prediction.  
452 
Additionally, continued performance of the calibration model should be confirmed on a periodic and 
453 
event-driven basis by comparison of the model predictions with the reference samples or reference 
454 
method results. This confirmatory testing helps to ensure that the calibration model continues to 
455 
perform as expected. Examples of events that could trigger confirmatory testing include new known 
456 
process variability, unexpected process events or scheduled instrument maintenance.  
457 
Monitoring of the model can be used to trigger model rebuilding (recalibration) as a part of continual 
458 
improvement. In general, the same considerations hold as for the original model building and internal 
459 
testing. Based on the cause of the model update (e.g., a process shift), new data may need to be 
460 
included and old non-relevant data may be taken out. 
461 
Once the new calibration model is established, the updated analytical procedure can be validated 
462 
against the same performance criteria as the one included in the original model. Aspects that are not 
463 
expected to change from the model update may not need to be evaluated (e.g., specificity).  
464 
 
 
465 
ICH Q14 Guideline 
 
15 
 
Figure 3: Multivariate Model Lifecycle 
466 
Multivariate Analytical 
Procedure Implementation  
& Performance Monitoring 
Multivariate Model 
Validation
Ongoing Multivariate 
Model Confirmatory 
Testing
(as needed)
Model 
Development
Multivariate Model 
Selection
Model assessment 
(within PQS)
Model Assumptions & 
Ranges / Initial Risk 
Assessment
Model Establishment
Routine Production
Model Maintenance
Establish (or update) 
Multivariate Model 
Maintenance Plan
 
467 
The multivariate model lifecycle is iterative and can be broken down into 3 major components: (1) 
468 
model establishment, (2) routine production and (3) model maintenance. 
469 
The choice of a multivariate model is based on the analytical procedure requirements and the 
470 
measurement technology selected. Prior to model development, the performance factors for the model 
471 
are defined, including the underlying model assumptions and desired ranges for model applicability. 
472 
An initial risk assessment can be valuable to understand potential sources of variability in the 
473 
materials and process that could affect the model performance and therefore should be considered 
474 
during the model calibration. Model development, including calibration and internal testing, follows 
475 
the considerations outlined in this chapter. Once the model is developed, it is validated using 
476 
independent data not previously used in the calibration set. The last step in model establishment is 
477 
development of a multivariate model maintenance plan, which includes the procedures and limits for 
478 
outlier diagnostics, and defines the frequency and circumstances for confirmatory testing, if needed.  
479 
Routine analysis of the multivariate analytical procedure typically includes monitoring the 
480 
appropriateness of every measurement using outlier diagnostics. Confirmatory testing against a 
481 
reference procedure is recommended on a pre-defined periodic or event driven basis (e.g., equipment 
482 
maintenance, new raw materials or process changes). Model assessment can be triggered by failure 
483 
of confirmatory testing or outlier diagnostics to meet the predefined criteria, or from data trending 
484 
indicating potential issues with the model, the process or the materials being measured (examples of 
485 
multivariate model lifecycle components are provided in Annex C).  
486 
Model assessment is performed within the PQS and utilizes knowledge management and risk 
487 
assessment. If an issue is identified, model development and revalidation may be needed, for example, 
488 
to add samples into the calibration set and remove those that are no longer relevant. In some cases, 
489 
the model may be performing appropriately, but additional experience may identify the need to 
490 
modify the limits of the model maintenance plan. In other cases, the issue identified could be related 
491 
ICH Q14 Guideline 
 
16 
 
to the measurement system (e.g., a misaligned sample interface) and no model update would be 
492 
needed. The dashed arrows in the figure illustrates reintroduction into the lifecycle flow based on the 
493 
potential outcomes of the model assessment.  
494 
 
495 
9. DEVELOPMENT OF ANALYTICAL PROCEDURES FOR REAL TIME RELEASE 
496 
TESTING: SPECIAL CONSIDERATONS 
497 
Real Time Release Testing (RTRT) is the ability to evaluate and ensure the quality of in-process and/or 
498 
final product based on process data, which typically include a valid combination of measured material 
499 
attributes and process controls (ICH Q8). RTRT measurements work in conjunction with all elements 
500 
of the control strategy (e.g., process monitoring or in-process controls) to ensure product quality. 
501 
RTRT can be applied to active substances, intermediates and finished products. 
502 
RTRT can be based on an appropriate combination of one or more process measurements and/or 
503 
material attributes to provide a prediction of one or more product CQAs and needs to be specific for 
504 
that CQA. The relationship between the RTRT approach and the product CQAs, as well as acceptance 
505 
criteria, should be fully justified. As appropriate, an RTRT procedure should be validated as 
506 
recommended in ICH Q2 and it should be demonstrated that the process measurements have 
507 
appropriate specificity for the targeted product quality attribute. 
508 
Sampling and the sample interface are important considerations when designing any on-line or in-
509 
line test method, including those used for RTRT. The measurement point(s) should be chosen to be 
510 
representative of the entire material being processed with the sample duration or amount appropriately 
511 
chosen (e.g., relative to a unit dose). Additionally, the sample interface should remain consistent over 
512 
the duration of manufacturing and should be robust to expected processing and environmental 
513 
variations. 
514 
The RTRT approach should be included in the product specification along with a reference to the 
515 
RTRT analytical procedure(s) and the related acceptance criteria, which are discussed in ICH Q6A 
516 
and Q6B. Quantitative RTRT results should be expressed in the same units as those for traditional 
517 
testing. The product specification will typically also include the analytical procedures to be used for 
518 
off-line testing. If the dossier includes a registered alternate control strategy to RTRT (e.g., traditional 
519 
end-product testing for when process analytics are unavailable), the related analytical procedures and 
520 
when they would be applied should also be included in the submitted product specifications. 
521 
 
522 
10. SUBMISSION OF ANALYTICAL PROCEDURE RELATED INFORMATION  
523 
10.1 
General Regulatory Considerations and Documentation 
524 
The analytical procedure description(s) should be included in the ICH M4Q CTD section 3.2.S.4.2 
525 
for drug substance or section 3.2.P.5.2 for drug product. Validation data and any supportive 
526 
information needed to justify the analytical procedure control strategy should be included in the CTD 
527 
section 3.2.S.4.3 for drug substance or section 3.2.P.5.3 for drug product. Other analytical procedures 
528 
used as part of the control strategy can be included in relevant CTD sections (e.g., 3.2.S.2, 3.2.P.3 
529 
and 3.2.P.4). The analytical procedure should describe the steps in sufficient detail for a skilled 
530 
analyst to perform the analysis as elaborated in Chapter 6. Submission of validation data should 
531 
follow the recommendations in ICH Q2. The criteria used in the validation study should be included 
532 
in the submission. In some cases, depending on the intended use (e.g., dissolution testing) and/or the 
533 
selected technique it may be appropriate to submit development data as justification.  
534 
ICH Q14 Guideline 
 
17 
 
Where ECs are proposed for analytical procedures as elaborated in Chapter 6, the ECs should be 
535 
clearly differentiated from supportive information. Additional development and validation 
536 
information can be included in sections 3.2.S.4.3 and 3.2.P.5.3 to justify ECs and their reporting 
537 
categories. When other lifecycle management elements as described in ICH Q12 are included in the 
538 
submission, the applicant should follow the principles described in ICH Q12 and Chapter 7 of this 
539 
document. 
540 
10.2 
Documentation for the Enhanced Approach  
541 
If an enhanced approach to development leads to the incorporation of enhanced elements into the 
542 
analytical procedure control strategy, then these should be justified.  
543 
Performance characteristics and acceptance criteria (e.g., described in an ATP) and other elements of 
544 
the enhanced approach (e.g., MODRs or PARs), should be described in the dossier sections for 
545 
analytical procedure description (e.g., 3.2.S.4.2 and 3.2.P.5.2). If ECs are proposed, then these should 
546 
also be included in the analytical procedure description, accompanied by supportive information. Use 
547 
of the enhanced approach should not lead to providing a less detailed description of analytical 
548 
procedures in a regulatory submission. 
549 
If ECs are proposed, risk-based categorization of changes and corresponding reporting categories 
550 
should be included in the submission. Appropriate justification should be given for parameters that 
551 
are ECs and those that are not ECs (see Chapter 6). For parameters that are not ECs and are typically 
552 
not included in a minimal procedure description a justification is not expected. 
553 
Appropriate information from analytical procedure risk assessment and development studies to 
554 
support the proposed lifecycle management strategy should be summarized and submitted in the 
555 
regulatory submission sections for analytical procedure validation (e.g., 3.2.S.4.3 and 3.2.P.5.3). 
556 
10.3 
Documentation for Multivariate Analytical Procedures and RTRT 
557 
Development information related to multivariate analytical procedures should be provided 
558 
commensurate with the level of impact of the model (Guide for ICH Q8/Q9/Q10 Implementation). 
559 
The process development section of the dossier (e.g., 3.2.S.2.6 or 3.2.P.2) should include the model 
560 
development information for multivariate models used as part of manufacturing development studies 
561 
or for in-process controls or tests. Supportive development information for RTRT multivariate models 
562 
can be included in either the appropriate analytical procedure validation or process development 
563 
section. 
564 
Validation information for multivariate analytical procedures used for release of drug product or drug 
565 
substance, including RTRT, should be included in the validation information section of the dossier 
566 
(e.g., 3.2.S.4.3 or 3.2.P.5.3). Additionally, these sections should include validation information on 
567 
analytical procedures used as reference methods. The model development, calibration and validation 
568 
information can be included directly in the CTD section or be in an appended document. 
569 
For multivariate models used as part of drug substance or drug product specifications, including 
570 
RTRT approaches, the description of the validation approach and results should include:  
571 
• Description of the independent validation sample set 
572 
• The performance criteria to be met during validation of the multivariate model 
573 
• Evaluation of the model validation results against the performance criteria 
574 
• Discussion of the relationship between the model performance criteria and the attribute 
575 
specification limits 
576 
• High level overview of the PQS elements for model monitoring and maintenance, such as 
577 
diagnostic tools for determining the appropriateness of the sample data for the model and the 
578 
approach taken when outliers are identified.  
579 
 
580 
ICH Q14 Guideline 
 
18 
 
The description of the multivariate analytical procedure used for RTRT should be provided in the 
581 
CTD section 3.2.S.4.2 for drug substance or section 3.2.P.5.2 for drug product and typically includes: 
582 
• The property or attribute of interest to be determined by the multivariate analytical 
583 
procedure and the desired quantitative ranges or limits  
584 
• A description of the measurement principle and pertinent instrument operating parameters 
585 
(e.g., sample presentation, sample interrogation time and measurement frequency) 
586 
• An overview of how the multivariate model calibration data are obtained (e.g., sample 
587 
preparation approach, reference method) 
588 
• The type of multivariate model (e.g., principal component analysis) 
589 
• A description of reference analytical procedure or high-level description of prepared 
590 
reference samples preparation 
591 
• Any calculations needed to adjust the model output into the reported value 
592 
 
593 
Additionally, section 3.2.S.4.2 for drug substance or section 3.2.P.5.2 for drug product should include 
594 
description of any analytical procedures that are part of a registered alternate control strategy to 
595 
RTRT. 
596 
 
 
597 
ICH Q14 Guideline 
 
19 
 
11. GLOSSARY 
598 
ACCURACY  
599 
The accuracy of an analytical procedure expresses the closeness of agreement between the value 
600 
which is accepted either as a conventional true value or as an accepted reference value and the value 
601 
measured. (ICH Q2) 
602 
ANALYTICAL PROCEDURE 
603 
The analytical procedure refers to the way of performing the analysis. The analytical procedure 
604 
description should include in detail the steps necessary to perform each analytical test. (ICH Q2) 
605 
ANALYTICAL PROCEDURE ATTRIBUTE 
606 
A technology specific property that should be within an appropriate limit, range, or distribution to 
607 
ensure the desired quality of the measured result. For example, attributes for chromatography 
608 
measurements may include peak symmetry factor and resolution. (ICH Q14) 
609 
ANALYTICAL PROCEDURE CONTROL STRATEGY 
610 
A planned set of controls derived from current analytical procedure understanding that ensures the 
611 
analytical procedure performance and the quality of the measured result. (ICH Q14) 
612 
ANALYTICAL PROCEDURE PARAMETER 
613 
Any factor (including reagent quality) or analytical procedure operational step that can be varied 
614 
continuously (e.g., flow rate) or specified at controllable, unique levels. (ICH Q14) 
615 
ANALYTICAL PROCEDURE VALIDATION STRATEGY 
616 
An analytical procedure validation strategy describes how to select the analytical procedure 
617 
performance characteristics for validation. In the strategy, data gathered during development studies 
618 
(e.g., using MODR or PAR) and system suitability tests (SSTs) can be applied to validation and an 
619 
experimental scheme for future movements of parameters within an MODR/PAR can be predefined. 
620 
(ICH Q14) 
621 
ANALYTICAL TARGET PROFILE (ATP) 
622 
A prospective summary of the performance characteristics describing the intended purpose and the 
623 
anticipated performance criteria of an analytical measurement. (ICH Q14) 
624 
CALIBRATION MODEL 
625 
A model based on analytical measurements of known samples that relates the input data to a value 
626 
for the property of interest (i.e., the model output). (ICH Q2) 
627 
 
628 
 
629 
 
630 
ICH Q14 Guideline 
 
20 
 
CONTROL STRATEGY 
631 
A planned set of controls, derived from current product and process understanding, that assures 
632 
process performance and product quality. The controls can include parameters and attributes related 
633 
to drug substance and drug product materials and components, facility and equipment operating 
634 
conditions, in-process controls, finished product specifications, and the associated methods and 
635 
frequency of monitoring and control. (ICH Q10) 
636 
CO-VALIDATION 
637 
Demonstration that the analytical procedure meets its predefined performance criteria when used at 
638 
different laboratories for the same intended purpose. Co-validation can involve all (full revalidation) 
639 
or a subset (partial revalidation) of performance characteristics potentially impacted by the change in 
640 
laboratories. (ICH Q2) 
641 
CRITICAL QUALITY ATTRIBUTE (CQA) 
642 
A physical, chemical, biological or microbiological property or characteristic that should be within 
643 
an appropriate limit, range or distribution to ensure the desired product quality. (ICH Q8) 
644 
CROSS-VALIDATION 
645 
Demonstration that two or more analytical procedures meet the same predefined performance criteria 
646 
and can therefore be used for the same intended purpose. (ICH Q2) 
647 
DETECTION LIMIT 
648 
The detection limit is the lowest amount of an analyte in a sample which can be detected but not 
649 
necessarily quantitated as an exact value. (ICH Q2) 
650 
DETERMINATION 
651 
The reported value(s) from single or replicate measurements of a single sample preparation as per the 
652 
validation protocol. (ICH Q2) 
653 
ESTABLISHED CONDITIONS (ECs) 
654 
ECs are legally binding information considered necessary to assure product quality. As a 
655 
consequence, any change to ECs necessitates a submission to the regulatory authority. (ICH Q12) 
656 
INTERMEDIATE PRECISION 
657 
Intermediate precision expresses within-laboratories variations. Factors to be considered should 
658 
include potential sources of variability, for example, different days, different environmental 
659 
conditions, different analysts and different equipment. (ICH Q2)  
660 
KNOWLEDGE MANAGEMENT 
661 
A systematic approach to acquiring, analysing, storing and disseminating information related to 
662 
products, manufacturing processes and components. (ICH Q10) 
663 
ICH Q14 Guideline 
 
21 
 
METHOD OPERABLE DESIGN REGION (MODR) 
664 
A combination of analytical procedure parameter ranges within which the analytical procedure 
665 
performance criteria are fulfilled and the quality of the measured result is assured. (ICH Q14) 
666 
ONGOING MONITORING 
667 
The collection and evaluation of analytical procedure performance data to ensure the quality of 
668 
measured results throughout the analytical procedure lifecycle. (ICH Q14) 
669 
PERFORMANCE CHARACTERISTIC 
670 
A technology independent description of a characteristic to ensure the quality of the measured result. 
671 
Typically, accuracy, precision, specificity/selectivity and range may be considered. The term was 
672 
previously called VALIDATION CHARACTERISTIC. (ICH Q2) 
673 
PERFORMANCE CRITERION 
674 
An acceptance criterion describing a numerical range, limit or desired state to ensure the quality of 
675 
the measured result. (ICH Q14) 
676 
PLATFORM ANALYTICAL PROCEDURE 
677 
A platform analytical procedure can be defined as a multi-product method suitable to test quality 
678 
attributes of different products without significant change to its operational conditions, system 
679 
suitability and reporting structure.  This type of method would apply to molecules that are sufficiently 
680 
alike with respect to the attributes that the platform method is intended to measure. (ICH Q2) 
681 
PRECISION  
682 
The precision of an analytical procedure expresses the closeness of agreement (degree of scatter) 
683 
between a series of measurements obtained from multiple samplings of the same homogeneous 
684 
sample under the prescribed conditions. Precision can be considered at three levels: repeatability, 
685 
intermediate precision and reproducibility. 
686 
The precision of an analytical procedure is usually expressed as the variance, standard deviation or 
687 
coefficient of variation of a series of measurements. (ICH Q2) 
688 
PROVEN ACCEPTABLE RANGE FOR ANALYTICAL PROCEDURES (PAR) 
689 
A characterised range of an analytical procedure parameter for which operation within this range, 
690 
while keeping other parameters constant, will result in an analytical measurement meeting relevant 
691 
performance criteria. (ICH Q14) 
692 
QUALITY RISK MANAGEMENT 
693 
A systematic process for the assessment, control, communication and review of risks to the quality 
694 
of the drug (medicinal) product across the product lifecycle. (ICH Q9)  
695 
 
696 
ICH Q14 Guideline 
 
22 
 
QUANTITATION LIMIT 
697 
The quantitation limit is the lowest amount of analyte in a sample which can be quantitatively 
698 
determined with suitable precision and accuracy. The quantitation limit for an analytical procedure 
699 
should not be more than the reporting threshold. The quantitation limit is a parameter used for 
700 
quantitative assays for low levels of compounds in sample matrices, and, particularly, is used for the 
701 
determination of impurities and/or degradation products. (ICH Q2) 
702 
RANGE 
703 
The range of an analytical procedure is the interval between the lowest and the highest reportable 
704 
results in which the analytical procedure has a suitable level of precision, accuracy and response. 
705 
(ICH Q2) 
706 
REPORTABLE RANGE 
707 
The reportable range of an analytical procedure includes all values from the lowest to the highest 
708 
reportable result for which there is a suitable level of precision and accuracy. Typically, the 
709 
reportable range is given in the same unit as the specification. (ICH Q2) 
710 
WORKING RANGE 
711 
The working range of an analytical procedure is the lowest and the highest concentration that 
712 
the analytical procedure provides meaningful results. Working ranges may be different before 
713 
sample preparation (sample working range) and when presented to the analytical instrument 
714 
(instrument working range). (ICH Q2) 
715 
REAL TIME RELEASE TESTING (RTRT) 
716 
The ability to evaluate and ensure the quality of the in-process and/or final product based on process 
717 
data, which typically include a valid combination of measured material attributes and process 
718 
controls. (ICH Q8) 
719 
REPEATABILITY 
720 
Repeatability expresses the precision under the same operating conditions over a short interval of 
721 
time. Repeatability is also termed intra-assay precision. (ICH Q2) 
722 
REPORTABLE RESULT 
723 
The result as generated by the analytical procedure after calculation or processing and applying the 
724 
described sample replication. (ICH Q2) 
725 
REPRODUCIBILITY 
726 
Reproducibility expresses the precision between laboratories (e.g., inter-laboratory studies, usually 
727 
applied to standardization of methodology). (ICH Q2) 
728 
 
729 
 
730 
ICH Q14 Guideline 
 
23 
 
RESPONSE 
731 
The response of an analytical procedure is its ability (within a given range) to obtain a signal which 
732 
is effectively related to the concentration (amount) of analyte in the sample by some known 
733 
mathematical function. (ICH Q2) 
734 
REVALIDATION 
735 
Demonstration that an analytical procedure is still fit for its intended purpose after a change to the 
736 
product, process or the analytical procedure itself. Revalidation can involve all (full revalidation) or 
737 
a subset (partial revalidation) of performance characteristics. (ICH Q2) 
738 
ROBUSTNESS 
739 
The robustness of an analytical procedure is a measure of its capacity to meet the expected 
740 
performance requirements during normal use. Robustness is tested by deliberate variations of 
741 
analytical procedure parameters. (ICH Q14) 
742 
SAMPLE SUITABILITY ASSESSMENT 
743 
A sample or sample preparation is considered suitable if the measurement response on the sample 
744 
satisfies pre-defined acceptance criteria for the analytical procedure attributes that have been 
745 
developed for the validated analytical procedure. Sample suitability is a pre-requisite for the validity 
746 
of the result along with a satisfactory outcome of the system suitability test. Sample suitability 
747 
assessment generally consists of the assessment of the similarity of the response between a standard 
748 
and the test sample and may include a requirement of no interfering signals arising from the sample 
749 
matrix. (ICH Q14) 
750 
SPECIFICITY/SELECTIVTY 
751 
Specificity and selectivity are both terms to describe the extent to which other substances interfere 
752 
with the determination of a substance according to a given analytical procedure. Such other 
753 
substances might include impurities, degradation products, related substances, matrix or other 
754 
components present in the operating environment. Specificity is typically used to describe the 
755 
ultimate state, measuring unequivocally a desired analyte.  Selectivity is a relative term to describe to 
756 
which extent particular analytes in mixtures or matrices can be measured without interferences from 
757 
other components with similar behaviour. (ICH Q2) 
758 
SYSTEM SUITABILITY TEST (SST) 
759 
These tests are developed and used to verify that the measurement system and the analytical 
760 
operations associated with the analytical procedure are adequate for the intended analysis and increase 
761 
the detectability of potential failures (ICH Q14) 
762 
 
763 
 
764 
 
765 
ICH Q14 Guideline 
 
24 
 
TOTAL ANALYTICAL ERROR 
766 
Total analytical error (TAE) represents the overall error in a test result that is attributed to imprecision 
767 
and inaccuracy. TAE is the combination of both systematic error of the procedure and random 
768 
measurement error. (ICH Q14) 
769 
VALIDATION STUDY 
770 
An evaluation of prior knowledge, data or deliberate experiments to determine the suitability of an 
771 
analytical procedure for its intended purpose. (ICH Q2) 
772 
VALIDATION TEST 
773 
Validation tests are deliberate experiments designed to determine the suitability of an analytical 
774 
procedure for its intended purpose. (ICH Q2) 
775 
 
776 
MULTIVARIATE GLOSSARY  
777 
CALIBRATION DATA SET 
778 
A set of data with matched known characteristics and measured analytical results, that spans the 
779 
desired operational range. (ICH Q2) 
780 
DATA TRANSFORMATION 
781 
Mathematical operation on model input data to assume better correlation with the output data and 
782 
simplify the model structure. (ICH Q14) 
783 
INDEPENDENT SAMPLE 
784 
Independent samples are samples not included in the calibration set of a multivariate model. 
785 
Independent samples can come from the same batch from which calibration samples are selected. 
786 
(ICH Q2) 
787 
INTERNAL TESTING 
788 
Internal testing is a process of checking if unique samples processed by the model yield the correct 
789 
predictions (qualitative or quantitative). 
790 
 Internal testing serves as means to establish the optimal number of latent variables, estimate the 
791 
standard error and detect potential outliers. Internal testing is preferably done by using samples not 
792 
included in the calibration set. Alternatively, internal testing can be done using a subset of calibration 
793 
samples, while temporarily excluding them from the model calculation. (ICH Q2) 
794 
INTERNAL TEST SET 
795 
A set of data obtained from samples that have physical and chemical characteristics that span a range 
796 
of variabilities similar to the samples used to construct the calibration set. (ICH Q14) 
797 
ICH Q14 Guideline 
 
25 
 
LATENT VARIABLES 
798 
Mathematically derived variables that are directly related to measured variables and are used in 
799 
further processing. (ICH Q2) 
800 
MODEL VALIDATION 
801 
The process of determining the suitability of a model by challenging it with independent test data and 
802 
comparing the results against prespecified criteria. For quantitative models, validation involves 
803 
confirming the calibration model’s performance with an independent dataset. For identification 
804 
libraries, validation involves analysing samples (a.k.a., challenge samples) not represented in the 
805 
library to demonstrate the discriminative ability of the library model. (ICH Q2) 
806 
MODEL MAINTENANCE 
807 
Safeguards over the lifecycle of a multivariate model to ensure continued model performance, often 
808 
including outlier diagnostics and resulting actions for model redevelopment or change in the 
809 
maintenance plans. (ICH Q14) 
810 
MULTIVARIATE ANALYTICAL PROCEDURE 
811 
An analytical procedure where a result is determined through a multivariate calibration model 
812 
utilizing more than one input variable. (ICH Q2) 
813 
OUTLIER DIAGNOSTIC 
814 
Tests that can identify unusual or atypical data in a multivariate analytical procedure. (ICH Q14) 
815 
REFERENCE PROCEDURE 
816 
A separate analytical procedure used to obtain the reference values of the calibration and validation 
817 
samples for a multivariate analytical procedure. (ICH Q2) 
818 
REFERENCE SAMPLE 
819 
A sample representative of the test sample with a known value for the property of interest, used for 
820 
calibration. (ICH Q14) 
821 
VALIDATION SET 
822 
A set of data used to give an independent assessment of the performance of the calibration model, 
823 
ideally over a similar operating range. (ICH Q14) 
824 
 
 
825 
ICH Q14 Guideline 
 
26 
 
12. References 
826 
ICH Q2 Validation of Analytical Procedures 
827 
ICH Q8 Pharmaceutical Development 
828 
ICH Q9 Quality Risk Management 
829 
ICH Q10 Pharmaceutical Quality System 
830 
ICH Q12 Technical and Regulatory Considerations for Pharmaceutical Product Lifecycle 
831 
Management 
832 
ICH M4Q The Common Technical Document for the Registration of Pharmaceuticals for Human 
833 
Use: Quality – M4Q 
834 
 
835 
13. ANNEX 
836 
13.1 
Annex A – Analytical Procedure Lifecycle 
837 
The examples provided in this Annex are mock examples for illustrative purposes. They suggest how 
838 
the concepts described in ICH Q14 could be applied and should not be used as a template or the sole 
839 
basis for a regulatory submission. 
840 
The examples have been created to illustrate 
841 
• how analytical procedure performance characteristics derived from the product context and 
842 
knowledge could be summarized in an ATP 
843 
• how performance characteristics described in the ATP could be applied to select a suitable 
844 
analytical technology, guide the development of an analytical procedure and help define the 
845 
analytical procedure control strategy  
846 
• how performance characteristics described in the ATP could aid the design of the validation 
847 
study for the analytical procedure 
848 
• how to identify ECs for analytical procedures developed using the enhanced approach 
849 
• how QRM and the adherence to associated criteria for relevant performance characteristics 
850 
and/or the subsequent execution of a bridging study can ensure the post-change quality of 
851 
the measured result and help to justify the respective reporting categories for ECs and the 
852 
post approval change management of analytical procedures 
853 
 
854 
As described in chapter 4, QRM can be used to evaluate the impact of proposed changes for analytical 
855 
procedures. The paragraph below describes examples of risk factors and risk reduction measures to 
856 
identify the risk associated with the changes to an analytical procedure. The outcome of the risk 
857 
assessment (risk level: high, medium or low) feeds into the design and extent of the studies needed to 
858 
support the change  
859 
Selected Risk (risk factors) 
860 
• 
Relevance of the test  
861 
• 
Potential clinical impact of the measured attribute (efficacy, safety, 
862 
pharmacokinetics and immunogenicity), e.g., controlling CQA vs non CQA  
863 
• 
Extent of knowledge of the attribute 
864 
• 
Attribute covered by other elements of the control system (testing or process control) 
865 
• 
Complexity of the technology 
866 
ICH Q14 Guideline 
 
27 
 
• 
Simple vs. complex technology 
867 
• 
Platform technologies 
868 
• 
Novel vs. established technology (e.g., in Pharmacopoeias) 
869 
• 
Several attributes reported as a sum (e.g., charge variants for large molecules) 
870 
• 
Biological assays, cell-based assays, immunochemical assays 
871 
• 
Multiattribute assays 
872 
• 
Multivariate assays 
873 
• 
Extent of the change  
874 
• 
Change of one or several parameters within MODR/PAR 
875 
• 
Change of one or several parameters outside the already proven ranges 
876 
• 
Change of the analytical procedure within existing analytical procedure 
877 
performance characteristics 
878 
• 
Change to analytical procedure performance characteristics (e.g., due to tightening 
879 
a specification limit or a change to the intended purpose of the procedure to 
880 
measure additional attributes) 
881 
 
882 
Risk reduction 
883 
Risk reduction is defined in ICH Q9 as actions taken to lessen the probability of occurrence of harm 
884 
and the severity of that harm. 
885 
Different kinds of knowledge can lead to reduction of risk, for example: 
886 
• 
Product and Process knowledge 
887 
- 
Knowledge about CQAs of the product/active substance and their acceptable ranges 
888 
- 
Well justified AP performance criteria cover/link to CQAs and their acceptable 
889 
range 
890 
- 
Knowledge about CPPs of the manufacturing process including risk assessment of 
891 
the process control capability over the CQA 
892 
- 
Evidence to control the CQAs through the process parameter settings 
893 
- 
Knowledge of the degradation pathways demonstrated by the analysis of relevant 
894 
stressed samples 
895 
- 
Other product knowledge (e.g., impurity profile, particle size and distribution) 
896 
• Analytical Procedure understanding and analytical procedure control strategy 
897 
- 
Knowledge about analytical procedure parameters and their impact on measurement 
898 
performance  
899 
- 
Proven analytical procedure robustness, e.g., harmonized procedures (compendial 
900 
tests) 
901 
- 
Enhanced method understanding (e.g., DoE studies) supporting justification of 
902 
acceptable ranges (e.g., PAR, MODR) to ensure quality of the result 
903 
- 
Other knowledge from development of analytical procedure 
904 
- 
System Suitability Test covers relevant analytical procedure attributes 
905 
- 
Ongoing monitoring of method output 
906 
- 
Clear link between signal and CQA to be measured (e.g., peak characterization 
907 
available, specificity) 
908 
• Subsequent Bridging strategy for the actual change 
909 
- 
Availability of well characterized reference material, relevant historical and or 
910 
stressed samples to support method output assessment against performance 
911 
requirements (demonstrated ability to control the CQA) 
912 
- 
Comparison to output of previous method (understanding and acceptance of risk for 
913 
potential differences) 
914 
ICH Q14 Guideline 
 
28 
 
- 
Demonstrated understanding of risks associated with parameter changes and 
915 
potential interactions with other parameters (if applicable) 
916 
- 
Prior experience or literature with similar changes, analyte or technology 
917 
- 
Reference to previous filings or to platform analytical procedures (if appropriate). 
918 
 
919 
13.1.1 Measurement of Stereoisomers as Specific Process Related Impurities in a Small Molecule Drug 
920 
Substance (DS) 
921 
 
922 
Introduction and Background 
923 
“Sakuratinib Maleate” is a small molecule DS with multiple chiral centers. The chirality of the 
924 
molecule, its degradation pathway and the impurities are well characterized.  From this knowledge 
925 
and the established manufacturing process controls the 5 Stereoisomers (Impurity A-E) were found 
926 
to be potentially present in the final product. Based on toxicological considerations, Impurity A-E 
927 
was specified at NMT 0.1%. One Stereoisomer F was found to be a process-related impurity but not 
928 
a degradation product. The stereoisomer was specified for release and re-test at NMT 0.5 % based on 
929 
toxicological data. Impurities G-J were other process-related impurities, of which process impurity J 
930 
was found to be also a degradation product of the DS. All specified impurities are isolated and 
931 
available as well characterized substances for procedure development and validation. 
932 
Table 1: Analytical Target Profile: 
933 
Intended Purpose 
Quantification of the stereoisomers A-F in Sakuratinib Maleate API for release testing. 
 
Link to CQA (Chiral Purity)  
The analytical procedures should allow for the individual quantification and determination of the total sum of 
the stereoisomers A-F to verify the CQA Chiral Purity ≥99.0% 
Characteristics of the Reportable Results  
Characteristic 
Acceptance Criteria 
Rationale 
Performance Characteristics  
Accuracy 
80-120% average recovery of spiked DS with Impurity 
A-E 
90-110% average recovery of spiked DS with Impurity F 
 
The values were derived 
from considerations of the 
significance of rounded 
values. At a specification 
level of 0.1%, 20% bias 
would lead to a variation of 
the analytical result of 
0.02%, which was found 
acceptable for a release 
decision. 
In a similar fashion, values 
for precision were derived. 
The recovery criteria for 
accuracy were set with 
respect to the reported result 
and taking into 
consideration any correction 
or response factors. 
Precision 
For impurity A-E 
Intermediate Precision RSD (n≥6): 
Impurity A-E ≤15% 
Impurity F ≤10%  
Specificity 
Analytical procedure should demonstrate to quantify 
with an acceptable bias of not more than 0.01% 
impurities A-F in presence of other likely process related 
substances or DS degradation products, which could be 
induced during chemical synthesis (Impurities G-J), and 
the salt forming agent. 
Potential interference with 
quantification of specified 
impurities by other regular 
components in the sample 
Reportable Range 
Impurity A-E: at least 0.05-0.12% 
Impurity F: at least 0.05-0.6%  
Reporting threshold to 120% 
of specification limit 
ICH Q14 Guideline 
 
29 
 
Initial Technology Selection 
934 
Multiple analytical technologies for chiral separations were available: Chromatographic methods 
935 
such as gas chromatography (GC), liquid chromatography (HPLC), supercritical fluid 
936 
chromatography (SFC) and thin-layer chromatography (TLC) were established technologies using 
937 
different chiral separation principles. More recently, capillary zone electrophoresis (CZE) and 
938 
capillary electrochromatography (CEC) had been shown to be alternatives to chromatographic 
939 
methods. Besides meeting the desired performance characteristics, further practical criteria were 
940 
considered in the technology selection for development, based on general technical knowledge, 
941 
operational needs, availability of equipment and capabilities in the company at the time: 
942 
• 
Complexity and robustness of technology  
943 
• 
Time and costs of analysis 
944 
• 
Standardization of technology and availability of multiple instrument suppliers  
945 
• 
Existing expertise in the company  
946 
 
947 
It was finally concluded to start method development with two technologies: Chiral HPLC and CZE. 
948 
As detection mode, UV detection was selected as it was known that the molecule had sufficient UV 
949 
absorption properties and standard for both separation techniques at the time.  
950 
 
951 
Analytical Procedure Development 
952 
 
953 
At initial development, a first screening was performed between HPLC and CZE technology. With 
954 
the technology and columns available at the time, only CZE could meet the expected performance for 
955 
specificity as described in the ATP, which served as primary endpoint for procedure development. 
956 
Therefore, the HPLC procedure development was discontinued at initial development.  
957 
 
958 
A risk analysis for the developed CZE procedure was performed. Parameters, where impact on the 
959 
performance of the procedure could not reasonably excluded were identified. See Ishikawa diagram 
960 
below: 
961 
 
962 
Figure 1: Ishikawa-Diagram 
963 
 
964 
 
965 
Analytical procedure parameters were investigated and their impact on the performance was 
966 
evaluated. The robustness of the CZE procedure was optimized and verified versus the performance 
967 
characteristics. Ultimately, the analytical procedure was optimized in the areas of sensitivity at QL, 
968 
repeatability of migration times and corrected peak areas, peak tailing of the API and stereoisomers, 
969 
and separation buffer depletion. Based on the development results, detailed instructions were given 
970 
in the analytical procedure description “Determination of the stereoisomers A-F in Sakuratinib 
971 
ICH Q14 Guideline 
 
30 
 
Maleate” and an SST was established on relative migration times resolution, LOQ, repeatability of 
972 
injection and the asymmetry of the DS peak as part of the analytical procedure control strategy.  
973 
 
974 
Table 2: Analytical Procedure Description  
975 
 
976 
Capillary: 
Uncoated fused silica, 50 µm diameter, at least 70 cm length 
Separation Buffer: 
13.2 g/l solution of ammonium phosphate adjusted to pH 6.0 with 
phosphoric acid filtered and 100 mM β-cyclodextrin, both ends of 
capillary 
Rinsing steps: 
1M sodium hydroxide, water, 0.1M sodium hydroxide  
Rinsing time at 1 psi at least 2 minutes each step  
 
 Column temperature: 
30°C 
Injection: 
Injection test solution (a) and the reference solution; injection for at least 
3 s then CZE buffer injection for 2 s at about 0.5 psi 
Separation field strength 
and polarity 
217 V/cm, normal mode 
Detection 
UV 214 nm 
 
977 
Method validation 
978 
After the analytical procedure description was finalized, a technology specific validation study was 
979 
planned according to the recommendations in ICH Q2. In alignment with the performance 
980 
characteristics, a technology and procedure specific set of attributes and criteria were derived from 
981 
the performance characteristics: 
982 
• 
The accuracy was measured by spiking three levels, 0.05, 0.1 and 0.12% for impurity 
983 
A-E, 0.05, 0.5 and 0.6% for impurity F to the DS salt form at 100% level and the 
984 
average recovery was calculated. The acceptance criteria for the average recovery of 
985 
80-120%  and 90-110% respectively were met 
986 
• 
For precision (repeatability), 6 separate preparations of the 6 stereoisomers were 
987 
made at specification limit. The RSD of 15% (Impurities A-E) respectively 10% 
988 
(Impurity F) criteria for precision of the migration time corrected peak areas were 
989 
met. Similarly, intermediate precision between operators, days and instruments were 
990 
performed and evaluated in an ANOVA experiment. 
991 
• 
Specificity was demonstrated by spiking all 6 stereoisomers to the API salt form and 
992 
impurities G-J, demonstrating sufficient baseline resolution (no detectable bias 
993 
between peaks) between the individual analytes of interest and no interference with 
994 
process related impurities. Additionally, blank injections of buffer and water were 
995 
compared with a sample to demonstrate no interference with the analyte detection. 
996 
• 
To verify the reportable range, a linearity, QL and DL experiment was performed and 
997 
compared to the technology specific acceptance criteria: 
998 
• 
DL was confirmed to be above a signal to noise ratio of 3:1 for all stereoisomers 
999 
• 
QL was confirmed by demonstrating the RSD of the corrected peak areas for the 
1000 
stereoisomers at the reporting threshold was NMT 10% 
1001 
• 
Linearity was found acceptable by demonstrating the correlation coefficient R 
1002 
was greater than 0.998 at 6 levels of stereoisomer concentrations ranging from 
1003 
0.05-2.0% for all impurities and the drug substance. A wider range was chosen 
1004 
to allow the application of the procedure for a potential wider range and allow a 
1005 
more precise determination of relative UV response factors 
1006 
• 
Linearity slopes of the stereoisomers were compared to the linearity of drug 
1007 
substance to demonstrate a UV response factor of about 1.0 for each 
1008 
stereoisomer versus the drug substance 
1009 
ICH Q14 Guideline 
 
31 
 
 
1010 
After the performance of the validation study, the results were summarized in a validation report, 
1011 
which concluded that the analytical procedure would meet the acceptance criteria for the analytical 
1012 
procedure attributes. The related performance characteristics were met. The analytical procedure was 
1013 
concluded to be fit for the intended purpose. 
1014 
 
1015 
Description of Established Conditions (ECs), Reporting Categories, and Justifications 
1016 
Based on product and process understanding and considering the procedure development data and 
1017 
risk assessment (see introduction to this annex), the applicant proposed established conditions and 
1018 
reporting categories as part of the initial submission. Justification of reporting categories for changes 
1019 
included adherence to predefined acceptance criteria described in the ATP and additional performance 
1020 
controls (e.g., system suitability testing and control samples).  
1021 
 
1022 
Note: The number of ECs and the associated reporting category listed in this table may depend on 
1023 
the extent of knowledge gained and information provided and is generated for this specific example 
1024 
only. The information provided in this example is not the entirety of the knowledge that is available 
1025 
and will be submitted to regulatory agencies and should not serve as general guidance. The extent of 
1026 
ECs, actual reporting categories, and data requirements may differ by region. Other parameters and 
1027 
conditions that are not identified as ECs in the table below may be required as EC for some cases 
1028 
depending on the region. The changes to other technologies may constitute different risks and may 
1029 
lead to different reporting categories. A PACMP may be required for some cases (e.g., a change 
1030 
between technologies) depending on region.  
1031 
 
1032 
Table 3: Proposed established conditions and reporting categories applying principles of ICH 
1033 
Q12 in the enhanced approach 
1034 
Established Condition 
Overall 
Risk 
Category 
ICH Q12 
Reporting 
Category 
Justification/ rationale 
Analytical Target Profile (ATP) 
 
High 
PA 
 
If widening the ATP is necessary, it will be 
reported as PA. 
Technology: Capillary Zone 
Electrophoresis with UV detection  
 
Suitable chiral separation technique to 
meet performance characteristics 
defined in ATP 
 
Low 
NL 
 
Adherence to ATP ensured by control strategy and 
defined bridging strategy (see below) to assess 
impact of changes. Changes to the method 
principle will be reported as NL. There is a strong 
understanding between product knowledge, 
intended purpose, and the analytical procedure 
performance established. In addition, well 
characterized analytical materials as well as a 
robust method development data set is available 
to allow a well-controlled bridging between 
technologies of similar separation capabilities 
(such as CZE to chiral HPLC). 
 
Technology Specific Analytical 
Procedure Attributes  
Low 
NL 
Accuracy and Precision (see ATP) 
 
Specificity: Baseline Separation with R NLT 2.0 
for Impurities A-F, DS, Salt forming agent and 
grouped impurities G-J. Impurities G-J do not 
need to be baseline separated amongst each other 
 
Linearity: R NLT 0.990 with at least 5 points in 
the range between 0.05%-2.0% for 
 
DL Impurities A-F: S/N NLT 3:1 below level 
0.05%  
ICH Q14 Guideline 
 
32 
 
Established Condition 
Overall 
Risk 
Category 
ICH Q12 
Reporting 
Category 
Justification/ rationale 
 
QL Impurities A-F: S/N NLT 10:1 at level 0.05%  
System Suitability Test and parameter-
control relationship as part of the 
overall Analytical Procedure Control 
Strategy: 
 
SST 1: Verification of relative 
migration times of analytes as listed in 
the analytical procedure. Asymmetry 
factor of the DS ≤ 1.5, Controlled 
factors:  
• 
Electric Field Strength 
• 
Rinsing agents & times 
• 
Separation buffer concentration and 
pH  
• 
Effective Capillary Length 
• 
Capillary material 
• 
Chiral buffer additive type and 
concentration 
 
SST2: Resolution between critical 
peak pair: API Main Peak and 
Impurity D ≥ 2.0, Controlled factors: 
• 
Chiral buffer additive type and 
concentration 
• 
Buffer composition 
• 
Buffer pH 
• 
Injection time/pressure (=volume) 
• 
Reference/Test solution concentration 
 
• 
SST3: S/N at LOQ API at 0.05% 
>10 :1, Controlled factors: Detection 
• 
Injection time and pressure 
• 
Sample and reference standard 
concentrations 
 
SST 4: Repeatability of injection of 
API at 0.5% level ≤ 5%, Controlled 
factors: 
• 
Injection parameters 
• 
buffer filtration 
 
Low 
NL 
SST was developed for the CZE procedure based 
on a risk analysis in alignment with the 
performance characteristics described in the ATP. 
The SST criteria are focused on critical 
performance characteristics during the regular 
application of the analytical procedure. Control 
relationships were established through prior 
knowledge (general principles of technique) or 
during method development. See further details 
with the parameters described below. 
 
A change in the SST should ensure similar or 
improved control of the associated factors listed 
in the left column. 
Separation Principle: 
Capillary: Material: uncoated fused 
silica capillary (diameter Ø = 50 µm) 
and β-cyclodextrin 
 
suitable instrumental and injection and 
buffer conditions to meet SST 
 
 
 
 
 
 
 
 
Low 
NL 
The capillary material, diameter and the chiral 
agent are the main parameters, defining the 
separation mechanism and component migration 
order. Changing these parameters would likely 
result in the adaptation of the SST, and therefore 
the same reporting category in alignment with the 
SST is proposed. It was demonstrated that SST 1 
and 2 provide controls for the parameters, 
therefore detectability is high, and the overall risk 
associated with changing these parameters was 
categorized as low. 
ICH Q14 Guideline 
 
33 
 
Established Condition 
Overall 
Risk 
Category 
ICH Q12 
Reporting 
Category 
Justification/ rationale 
 
The following conditions are not ECs in this example: 
 
Buffer Conditions 
Chemicals (Pharmacopeial quality) 
Separation buffer (CZE):  
13.2 g/l solution of ammonium 
phosphate adjusted to pH 6.0 with 
phosphoric acid filtered and 100 mM 
β-cyclodextrin 
 
Low 
- 
During robustness studies, the variations of buffer 
pH +/- 0.5, ammonium phosphate concentration, 
and cyclodextrin concentration +/-10% were 
shown not having an impact on the performance 
of the analytical procedure. The relationship 
between the parameters and SST 1 and SST 2 was 
demonstrated during development. The data is 
provided as part of the Analytical Procedure 
Validation Report.  
Instrumental conditions: 
Detection: 214 nm (UV) 
Electric Field Strength: 217 V/cm 
Temperature: 30 °C 
Separation: Separation buffer at both 
ends of the capillary  
Capillary effective length = at least 70 
cm 
Low 
-  
During robustness studies, typical variations in 
capillary temperature, and buffer concentrations 
and detection wavelength around +/-10% were 
shown not having an impact on the performance 
of the analytical procedure. The data is provided 
as part of the Analytical Procedure Validation 
Report. The relationship of electric field strength, 
voltage and capillary length is following scientific 
relationships as prior knowledge1  
During method development, SST 1-3 were 
demonstrated to be indicative for correct 
separation conditions. The data is provided as part 
of the Analytical Procedure Validation Report. 
 
 
 
 
Capillary rinsing conditions: 1M 
sodium hydroxide, water, 0.1M 
sodium hydroxide Instrument 
parameters, Rinsing time at least 2 
minutes each step at pressures greater 
than 1 PSI 
 
Low 
- 
During method development, rinsing times were 
chosen to allow the capillary surface to be 
equilibrated with no impact on migration times 
within a wide range of rinsing (i.e., +/-0.5 
minutes). Clear scientific relationships between 
pressure, capillary length and rinsing volume 
exist, allowing adjustments between various 
equipment1Erreur ! Signet non défini.. During method 
development, SST1 was demonstrated to be 
indicative for correct rinsing conditions. The data 
is provided as part of the Analytical Procedure 
Validation Report. 
Sample Analysis Injection test solution 
(a) and the reference solution; 
injection for at least 3 s then CZE 
buffer injection for 2 s, about 0.5 psi 
pressure. 
 
Low 
-  
Clear scientific relationships between pressure, 
capillary length and injection volume exist, 
allowing adjustments between various 
equipment1. During method development, SST1-3 
were demonstrated to be indicative for correct 
injection conditions. The data is provided as part 
of the Analytical Procedure Validation Report. 
API Reference Standard: 
Concentration of test solutions and 
reference standards: 1 mg/ml API in 
water 
Low 
-  
The performance over the reportable working 
range has been demonstrated though the linearity 
experiments at validation. The lower 
concentration range control was established 
through SST3 based on clear scientific principles 
(Beer-Lambert law). The upper concentration 
limit is influenced by the ionic strength of the 
sample and a clear scientific relationship between 
ionic strength, field strength, Joule heating and 
resulting band broadening exists2. A control 
relationship was established with SST 1 and 
SST2. 
1 Harmonized pharmacopoeial chapters of Capillary Electrophoresis such as Phar. Eur. 2.2.47, USP <727>, 
1035 
ICH Q14 Guideline 
 
34 
 
Japanese Pharmacopoeia (general information capillary electrophoresis) 
1036 
2 M. I. Jimidar, Capillar Electrophoresis Methods for Pharmaceutical Analysis, Volume 9, 2008,  
1037 
9-42 ISSN: 0149-6395 
1038 
 
1039 
Change assessment and bridging strategy 
1040 
 
1041 
The assumption is that the information in the table above (ECs and reporting categories) has been 
1042 
agreed upon up front with the regulatory agency. 
1043 
For every change, the MAH will perform a structured risk assessment to evaluate potential impact on 
1044 
the performance characteristics and the link to CQA (purity) as defined in the respective ATP. As a 
1045 
potential outcome of the risk assessment, experimental bridging studies to demonstrate adherence to 
1046 
the performance characteristics and associated criteria will be performed. These can include, if 
1047 
necessary, partial or full (re-)validation of the analytical procedure performance characteristics 
1048 
affected by the change and/or comparative analysis of representative samples and standards. 
1049 
 
1050 
The MAH commits to not implement the modified analytical procedure using the predefined reporting 
1051 
category if adherence to the performance characteristics and associated criteria defined in the ATP 
1052 
cannot be demonstrated during the bridging studies. If the precondition of adherence to the ATP 
1053 
cannot be met, a higher reporting category may apply. 
1054 
 
1055 
Change description and management 
1056 
The following scenario illustrate examples of post- approval changes and illustrate the steps a MAH 
1057 
would follow when actually implementing the change. 
1058 
 
1059 
Change #1: Change of buffer pH 
1060 
Background: 
1061 
 
1062 
The company has monitored and trended the migration times of the stereoisomers during routine use 
1063 
and found that the migration times could be reproduced in a more stable manner by shifting the buffer 
1064 
pH from 6.0 to 6.5.  
1065 
 
1066 
Application of Enhanced Understanding 
1067 
 
1068 
Elements of the enhanced approach (understanding the relationship between SST1 and procedure 
1069 
performance, procedure control strategy) were used to define a control relationship between buffer 
1070 
pH and SST1 and SST 2, as communicated in the submission.  
1071 
 
1072 
Risk assessment: 
1073 
 
1074 
The intended change was a change of the analytical procedure parameter, and this was agreed to be 
1075 
managed within the company’s quality system following the adherence to commitments made (i.e., 
1076 
the parameter was not an EC). 
1077 
 
1078 
a) Risk of change to the patient, product, and manufacturing process (Relevance of the test): 
1079 
The product is well established and characterized safe and efficacious. The current control strategy 
1080 
of the product is considered as sufficient and will not be impacted by the change. As a result, the 
1081 
specifications for the chiral impurities remain unchanged.  
1082 
 
1083 
b) Complexity of the technology: 
1084 
CZE is a well-established technology and the relationship of buffer pH and ionic strength on the zeta 
1085 
potential of the analytes and the capillary surface can be predicted through mathematical equations. 
1086 
ICH Q14 Guideline 
 
35 
 
 
1087 
c) Risk of change to the performance of the analytical procedure (Extent of the change) 
1088 
The extent of the change is low as it is a minor adjustment of the buffer pH 
1089 
 
1090 
Decision Tree Question #1: Considering product and procedure knowledge and understanding, what 
1091 
is the risk associated with the proposed changes to the reported result?  
1092 
Answer: Low  
1093 
 
1094 
Decision Tree Question #2: Are criteria of relevant performance characteristics defined in the 
1095 
dossier which ensure the quality of the measured result after the change?  
1096 
Answer: Yes  
1097 
 
1098 
Demonstration of analytical procedure performance after the change   
1099 
 
1100 
As there is a clear control relationship established between buffer pH and SST1 and SST2, 
1101 
demonstration of meeting the SST criteria is considered as appropriate along with meeting the 
1102 
relevant performance characteristics and associated criteria in the ATP. 
1103 
 
1104 
Conclusions 
1105 
Based on the initial risk assessment and the additional controls of SST 1 and SST 2 in place, the risk 
1106 
of changing the buffer pH is considered to be very low.  
1107 
 
1108 
Proposed Regulatory Reporting 
1109 
The original agreement with the regulator that this parameter is not an EC was confirmed as a result 
1110 
of the steps that were performed to implement the actual change. Thus, no regulatory reporting is 
1111 
needed.  The company will document this change within the PQS. 
1112 
 
1113 
Change #2: from chiral CZE to chiral HPLC 
1114 
Background 
1115 
As chiral column technology had advanced, the company could finally identify a suitable HPLC 
1116 
column and conditions for the intended purpose. The company intends to implement the analytical 
1117 
procedure for the control of stereoisomers of API for release of the final drug in an additional 
1118 
manufacturing site. The company strategy is to use the current (CZE) and future (HPLC) analytical 
1119 
procedures as alternative procedures. A well-established technology, chiral HPLC, is targeted in the 
1120 
alternative development to allow the use of a more standardized technology platform for small 
1121 
molecule drug substances. The intended change is not related to any quality issues of the product, or 
1122 
the established CZE procedure and the company does not intend to modify the specifications for the 
1123 
chiral impurities. 
1124 
 
1125 
Application of Enhanced Understanding 
1126 
The anticipated change will neither impact the already established product understanding nor the 
1127 
expected analytical procedure performance, as described in the ATP. Additionally, the fundamentals 
1128 
of the analytical techniques are well understood as general methodology and described in 
1129 
pharmacopoeias. Technology and analyte behaviour are predictable. The product, analytes, and 
1130 
sample preparation are well characterized and understood. Elements of the enhanced approach, such 
1131 
as the clear connectivity between SST and the analytical procedure performance as described in the 
1132 
ATP and risk assessment were applied to make use of the control strategy. Similar enhanced 
1133 
methodology used in the development of the CZE procedure will also be applied for the development 
1134 
of the HPLC procedure. 
1135 
 
1136 
ICH Q14 Guideline 
 
36 
 
Risk assessment: 
1137 
The intended change is a change in technology, and this was agreed as an EC with NL following the 
1138 
adherence to commitments made. 
1139 
 
1140 
a) Risk of change to the patient, product, and manufacturing process (Relevance of the test): 
1141 
The product is well established and characterized safe and efficacious. The current analytical control 
1142 
strategy of the product is considered as sufficient and will not be impacted by the change. As a result, 
1143 
the specifications for the chiral impurities remain unchanged.  
1144 
 
1145 
b) Complexity of the technology: 
1146 
Only well-established separation technologies (HPLC and CZE) are in scope. 
1147 
 
1148 
c) Risk of change to the performance of the analytical procedure (Extent of the change) 
1149 
The performance of the analytical procedure for its intended purpose is described through accuracy, 
1150 
precision, specificity, and result range. The intended change may have an impact on the analytical 
1151 
procedure performance. Therefore, the company has used an analytical target profile as upfront 
1152 
control element to minimize the risk of change. 
1153 
 
1154 
Decision Tree Question #1: Considering product and procedure knowledge and understanding, what 
1155 
is the risk associated with the proposed changes to the reported result?  
1156 
Answer: Medium  
1157 
 
1158 
Decision Tree Question #2: Are criteria of relevant performance characteristics defined in the 
1159 
dossier which ensure the quality of the measured result after the change?  
1160 
Answer: Yes  
1161 
 
1162 
Demonstration of Analytical Procedure performance after the change   
1163 
 
1164 
The procedure will be validated by establishing a technology specific validation protocol and 
1165 
acceptance criteria. The analytical procedure will be validated in alignment with ICH Q2(R2) Annex 
1166 
2, example separation technique. The acceptance criteria for validation will be derived from the ATP 
1167 
and will result in matching or stricter technology specific tests and criteria. The company has a quality 
1168 
system in place which ensures: 
1169 
• Appropriate analytical change control and risk evaluation 
1170 
• The ATP is translated into suitable validation tests and criteria once the technology is 
1171 
selected 
1172 
• That only analytical procedures will be used and implemented, which fulfill the performance 
1173 
criteria described in the ATP 
1174 
• Therefore, at any time, the appropriate analytical procedure performance will be guaranteed 
1175 
before its implementation for regular use. 
1176 
 
1177 
Conclusions 
1178 
Based on the initial risk assessment and the additional controls in place, the risk of using an HPLC 
1179 
method as alternative method to the CZE method is considered low. The original proposed reporting 
1180 
category of NL was confirmed as a result of the additional assessment and development/validation 
1181 
data. 
1182 
 
1183 
Proposed Regulatory Reporting 
1184 
The original EC with associated reporting category as agreed upon with the regulator per Table 3 was 
1185 
confirmed as a result of the steps that were performed to implement the actual change, thus the change 
1186 
will be submitted as notification low. 
1187 
 
 
1188 
ICH Q14 Guideline 
 
37 
 
13.1.2 Measurement of Potency for an anti-TNF-alpha Monoclonal Antibody  
1189 
 
1190 
Introduction and Background  
1191 
The example presented refers to the measurement of the relative potency of the drug, in this case an 
1192 
anti-TNF-alpha monoclonal antibody, in drug substance and in drug product at release and for 
1193 
stability testing.  
1194 
 
1195 
In addition to performing measurements of product CQAs, testing of potency is a unique feature of 
1196 
the release specification testing panel for biologics. Biological activity, measured by the potency, 
1197 
describes the specific ability or capacity of a product to achieve a defined biological effect1. Often, 
1198 
for complex molecules, the physicochemical information may be extensive but unable to confirm the 
1199 
higher-order structure which, however, can be inferred from the biological activity1. 
1200 
 
1201 
For the purpose of this example, it is assumed that the mode of action of the drug is the neutralisation 
1202 
of the biological activity of soluble TNF-alpha by preventing TNF-alpha from binding to the TNF-
1203 
alpha receptor. Fc-effector functions are out of scope of the measurement described in the example. 
1204 
For the purpose of this example, it is assumed that the specification limits for the relative potency are 
1205 
80% to 125% of the activity of the reference standard representative for the product. 
1206 
 
1207 
During development, forced degradation studies highlighted some modifications in the structure of 
1208 
the molecule as confirmed by physicochemical assays. The potency assay to be developed should be 
1209 
able to detect a change and/or a shift in potency upon forced degradation. 
1210 
 
1211 
The performance characteristics of the procedure used to generate the reportable result are accuracy, 
1212 
precision, specificity and reportable range. The evaluation of the precision involves variation of the 
1213 
key sources of variability of the analytical procedure such as analyst, days, key reagents (including 
1214 
cell culture parameters, if appropriate), key equipment.  
1215 
 
1216 
 
 
1217 
                                                   
1 ICHQ6B – specifications: test procedures and acceptance criteria for biotechnological/biological 
products.  
ICH Q14 Guideline 
 
38 
 
Table 4: Analytical target profile 
1218 
Intended Purpose 
Measurement of the relative potency of an anti-TNF-alpha monoclonal antibody in Drug Substance and in Drug 
Product at release and for stability testing.  
 
Link to CQA (biological activity) 
The mode of action of the drug is the neutralisation of the biological activity of soluble TNF-alpha by preventing 
TNF-alpha from binding to the TNF-alpha receptor. The assay should be able to measure the potency of the drug 
and detect if there are significant changes in biological activity upon forced degradation conditions. 
 
Characteristics of the reportable result 
Characteristic 
Acceptance criteria 
Rationale 
Performance characteristics 
Accuracy 
Relative accuracy1 is assessed via a linearity experiment 
that covers the reportable range. No trend in relative 
bias is observed over the tested relative potency range. 
 
The 95% Confidence Interval of the slope of the fitted 
regression line between theoretical and measured 
potency falls within a range of 0.8 to 1.25. 
 
The upper and lower 90% confidence interval for the 
relative bias calculated at each potency level is not more 
than 20%2, considering the intended purpose of the 
measurement. 
 
Parameters assessed based 
on compendial guidance 
e.g., USP<1033>3 
 
Selected performance 
characteristic ensures that 
the intended method 
delivers the quality 
reportable result.  
 
 
 
Precision 
Upper 95% Confidence Interval for the average 
intermediate precision across levels across the 
reportable range (95% CI % geometric coefficient of 
variation4) is not more than 20%4, considering the 
intended purpose of the measurement.  
Total Analytical 
Error (TAE)3 
(alternative approach 
to individual 
assessment of 
accuracy and 
precision) 
Different statistical measures can be used for evaluation 
of the capability of the method such as comparison of 
the TAE (combined accuracy and precision of the 
measurement) with the specification limit.5 
 
During development the 
specification limit may be 
target limits while for 
commercial they will be the 
proposed specifications.  
Specificity 
Method is specific for the intended mechanism of action 
of the active ingredient.  
Critical characteristic of a 
bioassay to ensure 
specificity towards the 
targeted biological activity. 
No interference from relevant process related impurities 
or matrix components. 
For example, process 
related and matrix 
components do not 
significantly affect the 
characteristics of the dose 
response curve. 
Assay is stability indicating i.e., method capable of 
detecting a change in potency and/or a change in the 
shape of the dose response curve, confirmed using 
forced degraded samples (for example samples 
subjected to meaningful thermal, photostability, and 
oxidative stress).  
To ensure that the product 
remains within specification 
over its shelf-life (e.g., 
retains the required safety 
and efficacy).5 
Reportable range 
The relative potency range is the range that meets 
accuracy and precision. It should include the 
specification range as a minimum (e.g., 80% to 120% of 
the specification range in this case corresponding to 
64% to 150% for a specification of 80% to 125% 
relative potency) 
Stated range for which the 
required accuracy and 
precision characteristics are 
demonstrated. 
ICH Q14 Guideline 
 
39 
 
1 The relative accuracy of a relative potency assay is the relationship between measured relative potency and known relative 
1219 
potency. Definition from USP<1033> Biological Assay Validation, May 2017. 
1220 
2 Individual values are just an example and can be different from product to product. 
1221 
3 USP <1220> Analytical Procedure Life Cycle. USP-NF 2022 ISSUE 1; USP<1210> statistical tools for procedure validation 
1222 
and references therein; P. Jackson et al., Anal. Chem. 2019, 91, 4, 2577–2585 
1223 
4 USP <1033> Biological Assay Validation, May 2017 
1224 
5 The suitability of this approach will depend on the phase of development and/or prior knowledge on the process performance.  
1225 
 
1226 
Technology selection:  
1227 
General considerations 
1228 
Based on the ATP above, there are several current technologies that may be a suitable choice for the 
1229 
measurement of the relative potency of an anti-TNF-alpha recombinant protein as illustrated in this 
1230 
example. 
1231 
It is common for the analytical technology for the measurement of potency to evolve during the 
1232 
product lifecycle for biologics, with ELISA-based technologies often being initially utilized prior to 
1233 
the subsequent development of a more technically challenging specific cell-based assay. The two 
1234 
methods rely on the binding of the active substance to the soluble TNF-alpha. While the signal of the 
1235 
ELISA is directly measuring the binding, the cell-based assay may target a later stage event, i.e., a 
1236 
downstream event in the signalling cascade. 
1237 
Cell-based bioassays can follow several assay methodologies. In the case of anti-TNF-alpha drugs, 
1238 
this includes neutralisation assays, where the assay measures the extent of soluble TNF-alpha- 
1239 
induced cytotoxicity and apoptosis in the presence of the drug. In addition, other formats such as 
1240 
reporter gene assay can be used.   
1241 
The ATP as described above can also be used in a risk assessment if the technology platform is 
1242 
changed. 
1243 
Cell proliferation assay as a specific example 
1244 
In this example, the format of the cell-based assay chosen to measure the relative potency of the anti-
1245 
TNF-alpha recombinant protein is a neutralisation - cell proliferation assay. It is presumed in this 
1246 
example that the Fc-effector functions are not involved. 
1247 
The potency will be determined by comparison of dilutions of the sample to be tested with dilutions 
1248 
of the like for like reference standard using a suitable cell-based assay based on the inhibitory action 
1249 
of the drug on the biological activity of soluble TNF-alpha with a suitable readout for assessing the 
1250 
inhibitory effect. The cell proliferation assay was chosen. This assay has the capability to monitor the 
1251 
inhibition induced by the TNF-alpha on the proliferation of a responsive cell line (e.g., murine 
1252 
fibrosarcoma WEHI-164). The assay compares the dose response of a test sample with a designated 
1253 
standard to provide a quantitative measurement of relative potency. The cells are incubated with 
1254 
varying dilutions of test sample and reference standard in presence of TNF-alpha. The cell growth is 
1255 
assessed by a staining method using a tetrazolium salt which is converted by cellular dehydrogenases 
1256 
to a colored formazan product. The amount of released formazan is measured using a 
1257 
spectrophotometer at 450 nm and 650 nm. The spectrophotometric response is directly proportional 
1258 
to the number of living cells. 
1259 
The throughput of the cell proliferation technology was limited to a small number of samples per day. 
1260 
The test is performed on several 96-well plates and on multiple days. The number of plates run to 
1261 
generate a valid reportable result will be established during the development of the analytical 
1262 
procedure. The equipment required to run this method are commonly used in bioassay laboratories. 
1263 
There are no specific operational nor safety concerns in applying them for bioassay trained analysts. 
1264 
ICH Q14 Guideline 
 
40 
 
Analytical Procedure Development  
1265 
The development of the analytical procedure described has been performed based on extensive 
1266 
knowledge of the molecule and relative potency assays.  
1267 
The following points are considered in the establishment of the potency assay:  
1268 
- 
Purpose and context of the assay defined in the ATP:  
1269 
o 
The applicant has extensive knowledge about relevant factors that could impact the CQA 
1270 
(relative potency of the drug) based on CQA assessment and process characterization and 
1271 
has established the link between the mode of action (MOA) and the clinical performance. 
1272 
Based on these data, the appropriate cell line and antigen binding conditions for the 
1273 
potency assay have been selected.  
1274 
o 
The molecule is characterized with other functional and/or physicochemical assays that 
1275 
contribute to understanding of the molecule and binding properties (e.g., Fc effector 
1276 
function). The other characterization assays are also continuously used in the lifecycle of 
1277 
the drug.  
1278 
o 
Performance characteristics for the analytical procedure are defined (e.g., via the TAE) to 
1279 
support the specification acceptance criteria.  
1280 
o 
Relative potency will be calculated for samples as compared to signal from a well-
1281 
characterized material (e.g., a reference standard) generated in the same analysis. 
1282 
- 
Extensive Knowledge was gained from development studies and prior knowledge on: 
1283 
o The cell line and its performance (viability, cultivation conditions, cell density, cell 
1284 
line stability (e.g., minimum and maximum number of passages) are well understood. 
1285 
Robustness of the cell cultivation conditions ensuring suitable cell metabolism was 
1286 
confirmed during the development of the analytical procedure.  
1287 
o Criteria for confluence and cell viability have been defined during development to 
1288 
ensure the required cell metabolism and leading to an appropriate signal amplitude 
1289 
and dose response curve. 
1290 
o Extensive studies have been done to identify the appropriate TNF alpha solution 
1291 
(antigen) leading to a spectrophotometrically measurable sigmoidal dose response 
1292 
curve in the presence of the reference samples or test samples, with lower and upper 
1293 
asymptotes corresponding to negative and positive controls, respectively. 
1294 
o The assay conditions have been studied and the parameters which influence the assay 
1295 
performance have been identified  
1296 
o Serial dilution levels were developed to optimize the dose-response curve, e.g., to 
1297 
ensure minimally three points in the linear segment of the dose-response curve and 
1298 
two in each asymptote.  
1299 
o The relative potency of the reference standard used in the procedure was qualified, 
1300 
and criteria around its performance were established to ensure run-to-run variability 
1301 
remains within suitable limits. 
1302 
 
1303 
QRM principles were used to guide the design of development studies. Features considered during 
1304 
risk assessment are shown in Figure 2. 
1305 
ICH Q14 Guideline 
 
41 
 
Figure 2: Ishikawa diagram 
1306 
Product specific 
reference
solution preparation
Cell Preparation
Results/
Performance Characteristics/
Analytical Procedure Attributes
Sample 
preparations
Assay Execution
Dose response
curve construction
TNF Alpha reference
solution preparation
Calculation of
Reportable result
medium preparation
preparation of cell suspension
(e.g. temperature, actinomycin)
conc. TNF alpha
solution
pre-incubation duration
pre- incubation temperature
amount of reference
solution per well
amount of cells
amount of sample
solution per well
incubation temperature
incubation duration
CO2 conc.
preparation of positive control well
addition of TNF alpha
amount of tetrazolium
salt per well
incubation duration
incubation temperature
CO2 conc.
wavelenght
calculation formula
interpolation from
response curve
replicates
PQS requirements (e.g equipment qualification, operator training), human factors, material variability,
environmental controls are considered during assessment of the individual steps, as appropriate
 
1307 
 
1308 
 
 
1309 
ICH Q14 Guideline 
 
42 
 
Table 5: Summary of development data and risk assessment 
1310 
 
1311 
Unit Operation 
Procedure 
Parameter 
Defined Target 
or Range 
Investigated 
Range 
Rationale  
Risk* 
Cell preparation 
Cell Density 
(cells/mL) 
1x106 cells/mL 
50 to 150 % of 
target value 
To ensure appropriate sensitivity of the 
assay 
medium 
Actinomycin D 
(µg/mL) 
2 µg/mL 
1-3 µg/mL 
Actinomycin D is used in the assay to 
enhance cell susceptibility to TNF and will 
ensure proper sensitivity of the assay. 
medium 
Cell viability 
Minimum 80% 
70-100% 
To ensure appropriate sensitivity of the 
assay 
medium 
TNF Alpha 
reference 
standard solution 
preparation 
Concentration of the 
TNF Alpha reference 
solution  
Targeted 
working 
concentration 
50 to 150% of 
targeted 
working 
concentration   
To ensure appropriate potency 
determination of the anti-TNF drug 
low 
Reference 
Standard/Control 
Sample 
Dilution factor 
Target 
Target 
To ensure appropriate potency 
determination of the anti-TNF drug 
low 
Assay execution 
Amount of cells 
added (µL) 
50 µL 
25 µL to 75 µL 
Volume of cell suspension needed to ensure 
appropriate response of the test 
low 
Pre-Incubation 
duration (h)  
1 h 
0.5 to 1.5 h 
Combination of incubation conditions to 
allow generation of an appropriate dose 
response curve 
low 
Pre-Incubation 
temperature (°C) 
37°C 
35-38°C 
Combination of incubation conditions to 
allow generation of an appropriate dose 
response curve 
low 
CO2 concentration 
(%) 
5% 
3-7% 
Combination of incubation conditions to 
allow generation of an appropriate dose 
response curve 
low 
Incubation duration 
(h) 
20 to 24 h 
16 to 30 h 
Combination of incubation conditions to 
allow generation of an appropriate dose 
response curve. For manipulation 
convenience, between 20 and 24 h has been 
low 
ICH Q14 Guideline 
 
43 
 
selected as target 
Incubation 
temperature 
37°C 
35-38°C 
Combination of incubation conditions to 
allow generation of an appropriate dose 
response curve 
low 
CO2 concentration 
(%) 
5% 
3-7% 
Combination of incubation conditions to 
allow generation of an appropriate dose 
response curve 
low 
Dose response 
curve 
Amount of 
tetrazolium salt 
added (µL of 
reconstituted 
solution) 
10 µL 
5 µl-15 µL 
Salt needed to perform the colorimetric 
reaction and the formation of formazan  
low 
Incubation duration  
3 to 4 h 
2 to 5 h 
Duration of the incubation to ensure 
optimum formation of formazan. 
Combination of duration and temperature of 
incubation 
low 
Incubation 
temperature  
20°C 
15-25°C 
Temperature of the incubation to ensure 
optimum formation of formazan. 
Combination of duration and temperature of 
incubation 
low 
 
1312 
* Risk refers to the impact on the reportable results (considering established controls (e.g., SST are fulfilled) 
1313 
ICH Q14 Guideline 
 
44 
 
Analytical procedure description2  
1314 
Equipment:  
1315 
- 
96-well plates 
1316 
- 
Tissue culture flasks 
1317 
- 
CO2 incubator 
1318 
- 
Biosafety cabinet 
1319 
- 
Plate reader 
1320 
 
1321 
Solutions & reagents:  
1322 
- 
WEHI-164 cells (ATCC) 
1323 
- 
TNF-alpha solution:  
1324 
o Dissolve the contents of a vial of TNF-alpha according to the supplier’s 
1325 
instructions. Further dilute with assay medium to obtain a suitable working 
1326 
concentration.  The cellular response to TNF-alpha varies and a suitable TNF-
1327 
alpha concentration (e.g., ED80) is determined using a TNF-alpha dose response 
1328 
curve. 
1329 
- 
Assay medium composed of RPMI 1640, L-glutamine, heat-inactivated fetal bovine 
1330 
serum (10% v/v) and a penicillin/streptomycin solution (1% v/v) 
1331 
- 
Actinomycin D 
1332 
- 
Tetrazolium salt WST-8 (5-(2,4-disulfophenyl)-3-(2-methoxy-4-nitrophenyl)-2-(4-
1333 
nitrophenyl)-2H-tetrazol-3-ium sodium)  
1334 
- 
Reference standard 
1335 
 
1336 
Procedure: 
1337 
The number of assay plates and days for each sample will depend on the control strategy 
1338 
defined for the method.  
1339 
- 
Reference solution and test solution:  
1340 
o Dilute with assay medium to the appropriate concentration. Analyse in duplicate. 
1341 
- 
Plate preparation:  
1342 
o Add 150 µL of assay medium to the wells designated for ‘cell only control’ and 
1343 
for blanks on a 96-well microplate. 
1344 
o Add 100 µL of assay medium and 50 µL of TNF-alpha working solution to the 
1345 
wells designated for ‘cell + TNF-alpha control’.  
1346 
o Add 100 µL of assay medium to the sample wells and 200 µL of the test or 
1347 
reference solutions. 
1348 
o Further prepare a series of 2-fold dilutions. 
1349 
o Then add 50 µL of TNF-alpha working solution. 
1350 
o Incubate at 36.0-38.0°C for 1h in an incubator using 5±2% CO2.  
1351 
- 
Cell preparation 
1352 
o Prepare a suspension of WEHI-164 cells containing 1x106 cells per milliliter, 
1353 
using assay medium containing 2 µg/mL of actinomycin D.  
1354 
 
1355 
2 Contains binding information (ECs) and non-binding information 
1356 
 
1357 
ICH Q14 Guideline 
 
45 
 
- 
Plating cells 
1358 
o Add 50 µL of the cell suspension to each well maintaining the cells in a uniform 
1359 
suspension during addition. 
1360 
o Incubate at 36.0-38.0°C for 20-24 h in an incubator using 5±2% CO2. 
1361 
- 
Addition of tetrazolium salt and absorbance measurement 
1362 
o Remove 100 µL of medium from each well. 
1363 
o Add 10 µL of reconstituted WST-8 mixture to each well and reincubate for 3-4 h. 
1364 
o Measure the absorbance using a microplate reader at 450 nm and 650 nm. 
1365 
o Estimate the quantity of formazan produced by subtracting the reading at 650 nm 
1366 
from the reading at 450 nm. 
1367 
 
1368 
Calculations:  
1369 
- 
Calculate the potency of the preparation to be examined using the four-parameter 
1370 
logistic curve model. 
1371 
- 
The reportable result is calculated in accordance with the defined number of replicates 
1372 
which is determined during development. Replication strategy may include averaging of 
1373 
the results of multiple plates, typically 3. Individual results within the range of the assay 
1374 
and having passed the sample suitability assessment are used for the calculation of the 
1375 
reportable result.  
1376 
 
1377 
Analytical procedure control strategy 
1378 
The analytical procedure control strategy for relative potency determination using the cell 
1379 
proliferation assay (performed as described in the example above) can include the following 
1380 
elements: 
1381 
System Suitability Test 
1382 
- 
The dose-response curve obtained for the reference standard curve corresponds to a 
1383 
sigmoid curve with upper and lower plateaus corresponding to ‘cell only control’ and 
1384 
‘cell + TNF-alpha control’, respectively. 
1385 
- 
The dose-response curve obtained for the test sample corresponds to a sigmoid curve 
1386 
with upper and lower plateaus corresponding to ‘cell only control’ and ‘cell treated with 
1387 
TNF-alpha control’, respectively. 
1388 
- 
The coefficient of determination calculated for each standard curve (r2) is not less than 
1389 
e.g., 0.97. 
1390 
- 
Maximum value (cell only) to minimum value (TNF-alpha control) ratio: minimum e.g., 
1391 
3.0. 
1392 
Sample suitability assessment: 
1393 
E.g., Assessment of similarity/ parallelism: 
1394 
- 
The upper asymptote ratio (A std/A test): e.g., 0.8-1.2 
1395 
- 
The lower asymptote ratio (D std/D test): e.g., 0.8-1.2 
1396 
- 
The Hill slope ratio (B std/B test): e.g., 0.8-1.2 
1397 
- 
The upper to lower asymptote ratio ((D-A) std/(D-A) test): e.g., 0.8-1.2 
1398 
 
1399 
ICH Q14 Guideline 
 
46 
 
Analytical procedure validation according to ICH Q2:  
1400 
- 
Validation protocol including predefined acceptance criteria for cell-based assay 
1401 
o Performance characteristics as defined in the ATP:   
1402 
▪ Accuracy 
1403 
Established by using various starting dilutions to generate different dose 
1404 
response curves  
1405 
• Acceptance criteria: 
1406 
o Relative accuracy is assessed via a linearity experiment that 
1407 
covers the reportable range.  No trend in relative bias is 
1408 
observed over the tested relative potency range. 
1409 
o The 95% Confidence Interval of the slope of the fitted 
1410 
regression line between theoretical and measured potency falls 
1411 
within a range of 0.8 to 1.25. 
1412 
o The upper and lower 90% confidence interval for the relative 
1413 
bias calculated at each potency level is not more than 20%, 
1414 
considering the intended purpose of the measurement. 
1415 
▪ Precision 
1416 
• Acceptance criterion: 
1417 
Upper 95% confidence interval for the average intermediate precision 
1418 
across the reportable range (95% CI % geometric coefficient of 
1419 
variation) is not more than 20% considering the intended purpose of 
1420 
the measurement. 
1421 
▪ Specificity 
1422 
• Acceptance criteria:  
1423 
o The method is specific for the intended mechanism of action 
1424 
of the active ingredient, i.e., no dose response curve is 
1425 
obtained (failure of one or more of the assay acceptance 
1426 
criteria) when other biological products are tested using the 
1427 
same method parameters. 
1428 
o No interference from relevant process related impurities or 
1429 
matrix components, i.e., process related impurities and matrix 
1430 
components do not significantly affect the characteristics of 
1431 
the dose-response curve. 
1432 
o The assay is stability indicating, i.e., the method is capable of 
1433 
detecting a change in potency and/or a change in the shape of 
1434 
the dose-response curve, confirmed using forced degraded 
1435 
samples (for example samples subjected to meaningful 
1436 
thermal, photostability, or oxidative stress). 
1437 
▪ Reportable range 
1438 
• Acceptance criterion:  
1439 
The relative potency range is the range that meets accuracy and 
1440 
precision. The reportable range should include the specification range 
1441 
as a minimum (e.g., 80% to 120% of the specification range). In this 
1442 
case, the reportable range corresponds to 64% to 150% relative 
1443 
potency. 
1444 
 
1445 
 
1446 
ICH Q14 Guideline 
 
47 
 
o Technology-dependent analytical procedure attributes:  
1447 
▪ Linearity of the results 
1448 
The relative accuracy is the relationship between measured relative potency 
1449 
and known relative potency.  
1450 
• Acceptance criteria:  
1451 
o The upper and lower 90% confidence relative accuracy is 
1452 
assessed via a linearity experiment that covers the reportable 
1453 
range. No trend in relative bias is observed over the tested 
1454 
relative potency range.  
1455 
o The 95% confidence interval of the slope of the fitted 
1456 
regression line between theoretical and measured potency falls 
1457 
within a range of 0.8 to 1.25. 
1458 
▪ Working range of the analytical procedure, i.e., upper to lower levels for 
1459 
which a suitable response curve is achieved.   
1460 
Individual potency results are used to generate the reportable result according 
1461 
to the replication strategy defined in the development. 
1462 
• acceptance criteria:  
1463 
o The final reportable result is within the specifications. The 
1464 
individual results agree to a defined RSD, 20%, and are 
1465 
covered by the validation range.  
1466 
o The validated range of the method is wide enough to 
1467 
encompass the individual result.  
1468 
 
1469 
- 
Execution of the validation  
1470 
The results were summarized in a validation report, which concluded that the analytical procedure 
1471 
would meet the acceptance criteria for the analytical procedure attributes. Implicitly, the 
1472 
performance characteristics were met and, in summary, the analytical procedure was suitable for 
1473 
the intended purpose. 
1474 
 
1475 
Description of Established Conditions, Reporting Categories, and Justifications 
1476 
Based on product and process understanding, and considering the procedure development data, the 
1477 
Applicant proposed Established Conditions and reporting categories, as part of the initial submission. 
1478 
Justification of reporting categories for changes includes adherence to predefined acceptance criteria 
1479 
described in the Analytical Target Profile and additional performance controls (e.g., system suitability 
1480 
testing and control samples).  
1481 
Figure 3 illustrates which analytical procedure steps are relevant for the performance controls defined 
1482 
as established conditions together with the additional continuous performance monitoring enablers.  
1483 
Table 6 describes the ECs, their reporting categories and justification.  
1484 
Note: The number of ECs, associated reporting category listed in this table may depend on the extent 
1485 
of knowledge gained and information provided. The information provided in this example is not the 
1486 
entirety of the knowledge that is available and will be submitted to regulatory agencies. The extent of 
1487 
ECs, actual reporting categories, and data requirements may differ by region. Other parameters and 
1488 
conditions that are not identified as ECs in the table below may be required as EC for some cases 
1489 
depending on the region. The changes to other method principles may constitute different risks and 
1490 
may lead to different reporting categories. PACMP may be required for some cases (e.g., a change 
1491 
between technologies) depending on region. 
1492 
ICH Q14 Guideline 
 
48 
 
Figure 3: – Illustration of the performance control strategy of the analytical procedure 
1493 
Cell preparation
TNF Alpha reference standard 
solution preparation
Dose- response curve construction
Product- specific reference standard 
solution preparation
ECs NM
Sigmoid curve for reference and test 
sample with upper and lower 
plateaus corresponding to ‘cell only 
control’ and ‘cell + TNF-alpha 
control’, respectively.
The coefficient of determination 
calculated for the standard curve (r2) 
is not less than a pre-defined 
threshold (0.97*).
Maximum value (cell only) to 
minimum value (TNF-alpha control) 
ratio is above a pre-defined target ( 
minimum 3.0*).
Upper asymptote ratio (A std/A test): 
0.8-1.2*
Lower asymptote ratio (D std/D test): 
0.8-1.2*
Hill slope ratio (B std/B test): 0.8-1.2*
Upper to lower asymptote ratio ((D-
A) std/(D-A) test): 0.8-1.2*
ECs NL
Number of passages, confluency, 
cell counting, cell viability 
Continuous 
performance 
monitoring 
e.g., trending 
of control 
samples, long 
term assay 
performance, 
capability of 
the PQS
Analytical Procedure Steps
Established Conditions  (technology-
specific performance controls)
Additional 
support 
systems and 
other enablers
Calculation of the reportable result
Sample preparation
PQS,
computer 
system 
validation
Assay Execution
 
1494 
    
* Individual values are just an example and can be different from product to product 
1495 
 
1496 
ICH Q14 Guideline 
 
49 
 
Table 6: Proposed established conditions and reporting categories applying principles of ICH Q12 in the enhanced approach 
1497 
 
1498 
Established conditions 
ICH Q12 Reporting Category 
Justification/rationale 
Performance characteristics as reported in the 
ATP 
PA 
Relevant performance characteristics to control the CQA 
Technology (principle) 
Cell Based Assay 
PA or NM1 
Adherence to ATP ensured by control strategy and defined bridging strategy (see below) 
to assess impact of changes 
Analytical procedure parameter 
Related to the control strategy elements (SST, sample suitability assessment) 
The dose-response curve obtained for the 
reference standard curve corresponds to a 
sigmoid curve with upper and lower plateaus 
corresponding to ‘cell only control’ and ‘cell + 
TNF-alpha control’, respectively 
NM 
The long-term performance of the analytical procedure is ensured by the adherence to 
ATP and by successful execution of the bridging strategy and PQS. 
The dose-response curve obtained for the test 
sample corresponds to a sigmoid curve with 
upper and lower plateaus corresponding to ‘cell 
only control’ and ‘cell + TNF-alpha control’, 
respectively. 
NM 
Coefficient of determination calculated for each 
standard/sample curve (r2); 
r2 is not less than 0.972 
NM 
Maximum value (cell only) to minimum value 
(TNF-alpha control) ratio. 
Minimum ratio 3.02 
NM 
ICH Q14 Guideline 
 
50 
 
Established conditions 
ICH Q12 Reporting Category 
Justification/rationale 
Assessment of similarity/ parallelism: 
e.g., The upper asymptote ratio (A std/A test): 
0.8-1.22 
The lower asymptote ratio (D std/D test): 0.8-
1.22 
The Hill slope ratio (B std/B test): 0.8-1.22 
The upper to lower asymptote ratio ((D-A) 
std/(D-A) test): 0.8-1.22 
 
NM 
Cell Preparation 
Cell line; 
WEHI-164 cells (ATCC) 
NM 
Based on the understanding of the mode of action (link to CQA) the suitability of the 
responsive cell line will be confirmed by responding to the TNF-alpha (survival of the 
cell in presence of the drug and cell death without drug). 
 
Adherence to ATP ensured by control strategy and defined bridging strategy (see below) 
to assess impact of changes. 
 
Revised system suitability test should ensure the suitability of the cell line and its 
performance (number of passages, confluency, cell counting, cell viability, signal 
amplitude, shape of the response curve) 
Preparation of cells: 
sub culturing 
NL 
Sufficient cell performance to detect changes in the quality of the drug is ensured by: 
- 
System suitability of the method covers the suitability of the cell preparation 
(number of passages, confluency, cell counting, cell viability, signal amplitude, 
shape of the response curve). 
- 
Changes in cell metabolism that impact performance of the method and link to CQA 
will be detected. 
- 
Changes that lead to insufficient cell performance will not be implemented as they 
could have an impact on the defined performance characteristics and would require 
prior approval. 
- 
Adherence to ATP ensured by control strategy and defined bridging strategy (see 
below) to assess impact of changes. 
Medium composition: 
RPMI 1640, L-glutamine, heat-inactivated fetal 
bovine serum,  
and a suitable antibiotic 
NL 
ICH Q14 Guideline 
 
51 
 
Established conditions 
ICH Q12 Reporting Category 
Justification/rationale 
Preparation of a suspension of WEHI-164 cells 
containing 1x106 cells per milliliter, using assay 
medium containing 2µg/mL of actinomycin D.  
 
NL 
TNF-alpha reference standard solution preparation 
Concentration of the TNF-alpha solution:  
Dilute with assay medium to obtain a suitable 
working concentration (e.g., ED80) as 
determined using a TNF-alpha dose response 
curve and meeting the control strategy elements. 
 
Shape of the TNF-alpha dose response curve: 
 
 
 
 
 
 
 
 
 
NL 
The effect of the drug on the TNF-alpha, which is the basis of the mode of action of the 
drug, is demonstrated by:  
 
- 
Adherence to ATP ensured by control strategy and defined bridging strategy (see 
below) to assess impact of changes.  
- 
1/ The dose-response curve obtained for the reference standard curve corresponds to 
a sigmoid curve with upper and lower plateaus corresponding to ‘cell only control’ 
and ‘cell + TNF-alpha control’, respectively. 
- 
2/ The dose-response curve obtained for the test sample corresponds to a sigmoid 
curve with upper and lower plateaus corresponding to ‘cell only control’ and ‘cell + 
TNF-alpha control’, respectively. 
- 
3/ The coefficient of determination calculated for the standard curve (r2) is not less 
than 0.97.2 
- 
4/Maximum value (cell only) to minimum value (TNF-alpha control) ratio: 
minimum 3.0.2 
- 
5/ Adherence to sample suitability assessment criteria 
 
 
 
 
 
 
 
ICH Q14 Guideline 
 
52 
 
Established conditions 
ICH Q12 Reporting Category 
Justification/rationale 
Sample Preparation and product specific reference solution preparation 
Preparation of the test sample and reference 
solutions: suitable amount of the solutions per 
well to meet the control strategy elements  
NL 
The suitability of the readout and of the dose response curve is ensured by the control 
strategy elements:  
- 
1/ The dose-response curve obtained for the reference standard curve corresponds to 
a sigmoid curve with upper and lower plateaus corresponding to ‘cell only control’ 
and ‘cell + TNF-alpha control’, respectively. 
- 
2/ The dose-response curve obtained for the test sample corresponds to a sigmoid 
curve with upper and lower plateaus corresponding to ‘cell only control’ and ‘cell + 
TNF-alpha control’, respectively. 
- 
3/ The coefficient of determination calculated for the standard curve (r2) is not less 
than 0.972. 
- 
4/Maximum value (cell only) to minimum value (TNF-alpha control) ratio: 
minimum 3.02. 
- 
5/ Adherence to sample suitability assessment criteria 
And by:  
- 
Adherence to ATP ensured by bridging strategy and PQS3 
Assay Execution Step 
Preparation of the positive control wells: 
Suitable Amount of TNF-alpha added 
NL 
The suitability of the readout and of the dose response curve is ensured by the control 
strategy elements:  
- 
1/ The dose-response curve obtained for the reference standard curve corresponds to 
a sigmoid curve with upper and lower plateaus corresponding to ‘cell only control’ 
and ‘cell + TNF-alpha control’, respectively. 
- 
2/ The dose-response curve obtained for the test sample corresponds to a sigmoid 
curve with upper and lower plateaus corresponding to ‘cell only control’ and ‘cell + 
TNF-alpha control’, respectively. 
- 
3/ The coefficient of determination calculated for the standard curve (r2) is not less 
than 0.972. 
- 
4/Maximum value (cell only) to minimum value (TNF-alpha control) ratio: 
minimum 3.02. 
Addition of the TNF-alpha solution to the wells:  
Suitable Amount of TNF-alpha solution per well 
NL 
Amount of cells added  
Add suitable amount of the cell suspension to 
each well maintaining the cells in a uniform 
suspension during addition 
 
NL 
Pre-incubation temperature and duration 
allowing to meet the control strategy elements 
Conditions (temperature, duration, %CO2) 
 
NL 
ICH Q14 Guideline 
 
53 
 
Established conditions 
ICH Q12 Reporting Category 
Justification/rationale 
Incubation temperature and duration allowing to 
meet the control strategy elements Condition 
(temperature, duration, %CO2) 
NL 
- 
5/ adherence to sample suitability assessment criteria 
And by:  
- 
Adherence to the ATP ensured by the bridging strategy and PQS3 
 
Dose response curve construction 
Reconstitute the Tetrazolium salt WST-8 (5-
(2,4-disulfophenyl)-3-(2-methoxy-4-nitropheny)-
2-(4-nitrophenyl)-2H-tetrazol-3-ium sodium)  
 
NL 
The suitability of the readout of the quantification of the effect of the drug on the cell is 
ensured by the control strategy elements:  
- 
1/ The dose-response curve obtained for the reference standard curve corresponds to 
a sigmoid curve with upper and lower plateaus corresponding to ‘cell only control’ 
and ‘cell + TNF-alpha control’, respectively. 
- 
2/ The dose-response curve obtained for the test sample corresponds to a sigmoid 
curve with upper and lower plateaus corresponding to ‘cell only control’ and ‘cell + 
TNF-alpha control’, respectively. 
- 
3/ The coefficient of determination calculated for the standard curve (r2) is not less 
than 0.972. 
- 
4/Maximum value (cell only) to minimum value (TNF-alpha control) ratio: 
minimum 3.02. 
- 
5/ adherence to sample suitability assessment criteria 
 
And by:  
- 
Adherence to ATP ensured by control strategy and defined bridging strategy (see 
below) to assess impact of changes3 
Add a suitable amount of the reconstituted 
tetrazolium salt to each well to meet the control 
strategy elements 
NL 
Incubation conditions (temperature, duration) 
allowing to meet the control strategy 
requirements: 
NL 
Wavelength: 
450 nm and 650 nm  
NL 
Four parameter logistic curve model 
NL 
PA: Prior Approval, NM: notification moderate; NL: notification low (as per ICH Q12 definitions) 
1499 
1 NM if no impact of the change on specification, PA if there is an impact on the specification (see case 1 and 2 below). Note, however, that regulatory agreement may differ by region.  
1500 
2 Individual values are just an example and can differ from product to product. 
1501 
3 Reporting category was initially NM but has been downgraded to NL based on the justification provided 
1502 
ICH Q14 Guideline 
 
54 
 
The following parameters are not ECs: 
1503 
• Preparation of the negative control wells 
1504 
• Plating format 
1505 
Change assessment and bridging strategy 
1506 
 
1507 
The assumption is that the information in the table above (ECs and reporting categories) has 
1508 
been agreed upon up front with the regulatory agency. 
1509 
 
1510 
For every change, the MAH will perform a structured risk assessment to evaluate potential 
1511 
impact on the performance characteristics and the link to CQA (biological activity) as defined 
1512 
in the respective ATP. As a potential outcome of the risk assessment, experimental bridging 
1513 
studies to demonstrate adherence to the performance characteristics and associated criteria will 
1514 
be performed. These can include, if necessary, partial or full (re-)validation of the analytical 
1515 
procedure performance characteristics affected by the change and/or comparative analysis of 
1516 
representative samples and standards. 
1517 
 
1518 
The MAH commits to not implement the modified analytical procedure using the predefined 
1519 
reporting category if adherence to the performance characteristics and associated criteria 
1520 
defined in the ATP cannot be demonstrated during the bridging studies. 
1521 
. 
1522 
Change Description and Management 
1523 
The following scenarios illustrate examples of post- approval changes and illustrate the steps a 
1524 
MAH would follow when actually implementing the change. 
1525 
 
1526 
Change #1: from classical cell culture (continuous cell culture) to ready to use cells (frozen 
1527 
cells) 
1528 
i) Background of change 
1529 
Change from continuous cell culture to ready to use cells for cell-based potency assay using the 
1530 
same cell line. This change affects only the analytical procedure step cell preparation. 
1531 
Conditions of freezing and thawing of the cells are the key parameters to control (cell 
1532 
metabolism of responsive cell line) for the success of this change, while the rest of the analytical 
1533 
procedure is unchanged. This change is inside the technology and is not expected to have an 
1534 
impact on the specifications. 
1535 
ii) Summary of structured risk assessment: 
1536 
The relevance of the test is classified as high as there is a direct link to the CQA potency, 
1537 
which is key for ensuring the efficacy of the drug. The change is not expected to impact the link 
1538 
to the CQA (same cell line used, same readout) and has low criticality in this respect. 
1539 
The cell-based assay used for the measurement of potency represents a complex technology as 
1540 
such assays have multiple sources of variability. Factors contributing to variability are well 
1541 
understood (based on prior knowledge and enhanced development data) and addressed in the 
1542 
analytical procedure control strategy. 
1543 
The extent of the change is restricted to the preparation of the cells (change in analytical 
1544 
procedure step cell preparation), with potential impact on only one analytical procedure 
1545 
ICH Q14 Guideline 
 
55 
 
attribute (cell metabolism). Factors contributing to the cell performance are understood, 
1546 
investigated as part of development of the ready to use cell preparation and monitored by the 
1547 
SST.  
1548 
The initial risk assessment proposed a moderate risk. Further evaluation was performed 
1549 
following Step 2 of ICH Q14 Figure 2. 
1550 
iii) Adherence to criteria for relevant performance characteristics 
1551 
The understanding of the analytical procedure and link to the CQA allowed the definition of 
1552 
criteria for relevant performance characteristics which ensure the post change quality of the 
1553 
measured result after the change (please refer to Table 4). The change can potentially affect cell 
1554 
metabolism and hence the method performance characteristics accuracy and precision. Before 
1555 
implementation of the change, adherence to these performance characteristics should be 
1556 
demonstrated. This change does not impact the performance characteristics specificity and 
1557 
reportable range as the same cell line is used and the potency is measured against the same 
1558 
reference standard. 
1559 
 
1560 
iv) Demonstration of Analytical Procedure performance after change 
1561 
Evaluation of impact on performance characteristics 
1562 
Based on analytical procedure understanding the following parameters that could potentially 
1563 
impact the performance have been evaluated and defined in the analytical procedure 
1564 
description: Cell freezing and thawing conditions/cell metabolism are the key parameters to 
1565 
control (freezing medium, freezing conditions, growth/assay medium). The SST of the method 
1566 
covers the suitability of the cell preparation (e.g., confluency, cell density, cell viability, signal 
1567 
amplitude, shape of the response curve). 
1568 
Experimental Bridging Study Results 
1569 
In accordance to Table 2 of ICH Q14 a partial revalidation of the analytical procedure was 
1570 
performed to demonstrate the affected analytical procedure attributes are met after the change. 
1571 
Comparative analysis of a set of representative samples with pre- and post-change analytical 
1572 
procedure will be performed to ensure that the achieved results are comparable or that observed 
1573 
differences are acceptable and do not impact the established specification. 
1574 
v) Conclusion 
1575 
Evaluation of performance characteristics demonstrated that defined criteria could be met. The 
1576 
result of the studies confirmed the expected cell performance post change. The purpose of the 
1577 
method has not changed and its capability to generate the reportable result is unchanged. 
1578 
Method bridging was successfully performed. The risk associated with the change is considered 
1579 
low taking into account the outcome of the initial risk assessment, the evaluation of the 
1580 
performance characteristics and the bridging study results.  
1581 
vi) Regulatory reporting: 
1582 
The original EC with associated reporting category as agreed upon with the regulator per Table 
1583 
6 was confirmed as a result of the steps performed, thus the change is proposed as notification 
1584 
low. The revised analytical procedure description together with the analytical validation report 
1585 
and the outcome of the bridging study will be submitted accordingly. The SST criteria of the 
1586 
analytical procedure including those ensuring sufficient cell performance remain unchanged. 
1587 
ICH Q14 Guideline 
 
56 
 
Appropriated development data demonstrating suitable absence of impact on cell performance 
1588 
upon preparation and handling frozen cell will be provided. 
1589 
 
1590 
Change #2: from binding ELISA to cell-based assay 
1591 
Another example considers a development scenario where the MAH has initially developed a 
1592 
binding assay (ELISA) to determine the relative potency of the anti TNF alpha recombinant 
1593 
protein and plans to implement a cell-based assay post approval. The measurement requirement 
1594 
as defined in the ATP (Table 4) and included in the initial marketing authorization remained 
1595 
unchanged and were used to support assay development and implementing the change. 
1596 
i) Background of change:  
1597 
Change from binding ELISA to cell-based assay. Both methodologies evaluate the relative 
1598 
potency of the drug in comparison to a reference standard. However, the evaluation of the 
1599 
mechanism of action is usually different: Binding ELISA targets early-stage event (binding 
1600 
activity only), while cell-based assay targets late stage event, i.e., downstream event in the 
1601 
signaling cascade. The change from ELISA to a cell-based assay is outside the technology and 
1602 
a potential impact on the specifications acceptance criteria cannot be excluded. 
1603 
ii) Summary of structured risk assessment: 
1604 
The relevance of the test is classified as high as there is a direct link to the CQA potency, which 
1605 
is key for ensuring the efficacy of the drug. The change could impact the measurement of the 
1606 
CQA potency as the change is from an immunochemical binding assay to a cell-based assay 
1607 
where also downstream event cascades can be targeted. However, this change is expected to 
1608 
better reflect the mode of action of the product. 
1609 
The cell-based assay proposed to be used for the measurement of potency represents a complex 
1610 
technology as it is related to multiple sources of variability. Analytical procedure parameters 
1611 
have been evaluated following a risk-based approach and it could be demonstrated that factors 
1612 
contributing to variability are well understood (based on prior knowledge and enhanced 
1613 
development data) and addressed in the analytical procedure control strategy. 
1614 
The extent of the change is high as a change in technology from an immunochemical binding 
1615 
assay to a cell-based assay is foreseen. The functional properties of the molecule and related 
1616 
mode of action are well understood and supported by preclinical and clinical data. Different 
1617 
responsive cell line candidates have been screened. The WEHI 164 cell line and the assay 
1618 
format (cell proliferation) have been chosen based on predefined selection criteria and the mode 
1619 
of action of the molecule. To address the mode of action of the molecule (anti-TNF), a TNF-
1620 
alpha standard is used to measure the impact of its addition on the proliferation of the cells in 
1621 
presence of the drug. Optimal amounts of TNF-alpha and of drug have been identified and are 
1622 
described in the analytical procedure. Relevant SST criteria have been defined to ensure the 
1623 
proper control of the analytical procedure (refer to analytical procedure description). The initial 
1624 
risk assessment proposed a high risk. Further evaluation was performed following Step 2 of 
1625 
ICH Q14 Figure 2. 
1626 
iii) Adherence to criteria for relevant performance characteristics 
1627 
The understanding of the analytical procedure and link to the CQA allowed the definition of 
1628 
criteria for relevant performance characteristics which ensure the quality of the measured result 
1629 
after the change (please refer to ATP table above). In spite of analytical method principle being 
1630 
ICH Q14 Guideline 
 
57 
 
different between the immunochemical binding ELISA and the cell-based assay methods, in 
1631 
both procedures the reportable result is measured and calculated relative to the same reference 
1632 
standard allowing data normalisation (RS used as “internal calibrator”). Consequently, the 
1633 
reportable result is expressed using the same approach (% relative potency). However, based 
1634 
on the extent of change a validation of the new procedure including data driven assessment of 
1635 
adherence to the performance characteristics as defined in ATP is required.  
1636 
 
1637 
iv) Demonstration of Analytical Procedure performance after change 
1638 
The cell-based assay was developed based on the criteria defined in the ATP. After development, 
1639 
validation of the analytical procedure was performed.  
1640 
If adherence to the performance characteristics as defined in the ATP can be demonstrated and 
1641 
no change to the specification acceptance criteria is needed, then the bridging studies will be 
1642 
initiated. 
1643 
However, due to the complex nature of the cell-based assay, the performance characteristics 
1644 
may be affected compared to the binding ELISA (e.g., precision). An assessment should be 
1645 
done to determine if the performance of the assay still meets the criteria described in the ATP 
1646 
and supports the specification acceptance criteria. In case a change of the performance criteria 
1647 
described in the ATP and/or the specification acceptance criteria is needed, the change should 
1648 
follow a pre-approval pathway. 
1649 
Experimental Bridging Study Results  
1650 
In accordance to Table 2 of ICH Q14 a full validation of the cell-based procedure was performed 
1651 
to demonstrate the suitability for its intended purpose. The cell-based procedure was found to 
1652 
satisfy the requirements of the ATP. Comparative analysis of a set of representative samples 
1653 
with the ELISA and cell-based analytical procedures was performed including representative 
1654 
degraded samples (forced degraded samples able to detect a loss of potency or end of shelf-life 
1655 
samples). The studies were designed to demonstrate continuity of the results generated with the 
1656 
two methods (e.g., abnormal results should be detected as non-conforming by both methods).  
1657 
v) Conclusions  
1658 
Validation of the cell-based procedure and evaluation of performance characteristics 
1659 
demonstrated that the defined criteria were met. The result of the studies demonstrated the 
1660 
ability of both the ELISA and cell-based procedures to measure relative potency with the 
1661 
required levels of accuracy, precision and specificity. The purpose of the analytical procedure 
1662 
had not changed and its capability to generate the reportable result was unchanged.  
1663 
Method bridging was successfully performed. The change evaluation showed that the extent of 
1664 
change had no impact on the ATP nor on specifications. In addition, the bridging evaluation of 
1665 
the two methods had confirmed that the relative potency specification remained unchanged. 
1666 
The risk associated with the change was considered moderate taking into account the outcome 
1667 
of the initial risk assessment, the evaluation of the performance characteristics and the bridging 
1668 
strategy.  
1669 
vi) Regulatory reporting  
1670 
The original EC with associated reporting category as agreed upon with the regulator per Table 
1671 
6 was confirmed as a result of the steps performed, thus the implementation of the change will 
1672 
ICH Q14 Guideline 
 
58 
 
be submitted to the relevant regulatory authorities using “Notification moderate” category. The 
1673 
revised analytical procedure description together with the analytical validation report and the 
1674 
outcome of the bridging study will be submitted. 
1675 
 
 
1676 
ICH Q14 Guideline 
 
59 
 
13.2 
Annex B: Validation Strategies for MODRs 
1677 
This annex describes validation strategies for MODRs and includes an example table to present 
1678 
the performance characteristics combined with the attribute acceptance criteria, parameter 
1679 
ranges, control strategy and validation strategy. 
1680 
ICH Q2 provides the concepts for analytical procedure validation. Generally, the operating 
1681 
space needs to be covered by validation data. The extent of validation activities and the 
1682 
respective operational flexibility associated requires to be assessed and justified on a case-by-
1683 
case basis. Performance characteristics whose validation is already comprised by development 
1684 
are not considered. Two options below represent examples of typical approaches, allowing also 
1685 
in-between solutions.  
1686 
 
1687 
Option 1: 
For validation, at minimum, a single set of univariate operating parameters of 
1688 
the MODR is selected (typically the intended operational conditions or the set 
1689 
point). For future changes of the parameters within the MODR an assessment 
1690 
with regard to additional validation activities should be performed. The strategy 
1691 
for determining the extent of additional validation should be described in the 
1692 
submission  
1693 
Option 2: 
The validation of the set point, e.g., center point, and the extrema of the MODR 
1694 
allows full operational flexibility within the MODR without demand for further 
1695 
validation activities. 
1696 
Figure 1 gives an overview on the lifecycle steps of an analytical procedure showing the impact 
1697 
of the two different validation options. 
1698 
DoE
AP Development
Risk Assessment
MODR
AP Control Strategy
AP Validation 
Strategy
Validation Option I
at Set-Point
Validation Option II
at Set-Point +
Additional Validation 
at Extrema
Update of AP Control 
Strategy
Update of AP Control 
Strategy
Additional Validation
Update of AP Control 
Strategy
Assessment of 
required Validation 
Activities
No Validation Activity 
required
Move AP Parameters 
within MODR
Move AP Parameters 
within MODR
 
1699 
Figure 1: Analytical Procedure Lifecycle following different validation options 
1700 
Table 1 represents an approach to summarize the basic knowledge on an analytical procedure 
1701 
and can be used as a consulting resource for changes. It is an example how to compile the core 
1702 
information of an analytical procedure based on the ATP (col. B) and the DoE results (columns 
1703 
ICH Q14 Guideline 
 
60 
 
D, E, F), leading to the definition of the MODR (col. D) as well as the individual ranges which 
1704 
are shown to fulfil the criteria of specific analytical procedure attributes (col. E). The MODR 
1705 
(col. D) originates common overlap of these individual ranges (col. E), whereas the existing 
1706 
information (col. F) defines the entire investigated range covered by the experiments. At the 
1707 
same time, Table 1 allows to align the acceptance criteria of the analytical procedure attributes 
1708 
(col. B) with the analytical procedure control strategy (col. G) and even to set up an analytical 
1709 
procedure validation strategy (col. H) for the analytical procedure performance characteristics 
1710 
(col. A) derived from ICH Q2. The experimental scheme for future movements of parameters 
1711 
within an MODR can be predefined in the analytical procedure control strategy (col. G). 
1712 
 
1713 
Table 1: Comprehensive compilation of analytical procedure information 
1714 
 
1715 
 
1716 
ICH Q14 Guideline 
 
61 
 
13.3 
Annex C: Example of Multivariate Model Lifecycle Components 
1717 
Model 
Description 
On-line NIR to determine blending 
ranges to achieve blend uniformity 
during development 
Measurement of Content Uniformity 
and Assay of uncoated tablets by NIR 
used for product release 
Glucose Raman model used for qualitative 
identification testing on incoming raw 
material release for GMP use 
Model Category – Low Impact 
Model Category - High impact 
Model Category – High impact 
User requirements 
Defined model requirements (e.g., 
ATP) 
Defined model requirements (e.g., ATP) 
Risk 
Assessment 
Initial assessment based on existing 
knowledge, laboratory and pilot 
studies, or DOE, as appropriate. 
Formal risk assessment based on 
knowledge gained during initial 
development.  
Formal risk assessment with knowledge 
gained during initial development 
Model 
Development -  
Calibration 
Scientifically sound approach based 
on laboratory and pilot data and 
previous experience. 
Formal design-based approach (e.g., 
DOE) covering appropriate ranges of 
relevant variability sources with 
established acceptance criteria that are 
suitable for intended use. 
Formal design-based approach covering 
appropriate ranges of relevant variability 
sources (raw material, lots, packaging, 
instruments-to-instrument, user, software 
limitation) with established acceptance 
criteria that are suitable for intended use. 
Establish an identification threshold that 
has the same probability of detection as 
the existing method and a suitable 
alternative testing method should the 
Raman method fail. 
Validation 
(Verification) 
Assess specificity and robustness, 
optionally assess linearity and/or 
precision 
Full validation covering applicable 
performance characteristics across 
reportable ranges with established 
acceptance criteria (ICH Q2). 
Full validation covering applicable 
performance characteristics across 
reportable ranges with established 
acceptance criteria (ICH Q2). Include 
establishing suitable comparability of 
Raman method to existing method for 
release (can be reference method) 
Performance 
Monitoring 
Routine monitoring – maintain data 
sources (instruments), automation 
connectivity, and data integrity. 
Routine monitoring – maintain data 
sources (instruments), automation 
connectivity, and data integrity. 
Routine monitoring – maintain data 
sources (instruments), automation 
connectivity, and data integrity. 
Real-time diagnostics – implement 
initial diagnostics to confirm model 
performance in real-time. 
Real-time diagnostics – implement 
routine diagnostics to confirm model 
performance in real-time. 
Real-time diagnostics – implement routine 
diagnostics to confirm model performance 
in real-time. 
Periodic monitoring – if applicable, 
compare model predicted results to 
reference method at a frequency that 
is scientifically justified or on an 
event driven basis as needed. 
Periodic monitoring – compare model 
predicted results to reference method 
at a frequency that is scientifically and 
statistically justified or on an event 
driven basis. 
Periodic monitoring – compare model 
predicted results to reference method at a 
frequency that is scientifically and 
statistically justified or on an event driven 
basis. 
Model 
Maintenance 
Model Update - updates are common 
during the process development 
stage as new experimental data 
becomes available 
Model Update - updates should be 
triggered based on Model Monitoring 
and Maintenance Strategy. 
Model Update - updates should be 
triggered based on Model Monitoring and 
Maintenance Strategy. 
Change Management per PQS 
Change Management per PQS 
Change Management per PQS 
 
1718 