22 July 2021 
EMA/CHMP/QWP/BWP/259165/2019 
Committee for Medicinal Products for Human Use (CHMP) 
 
Guideline on quality documentation for medicinal products 
when used with a medical device 
 
Draft agreed by Quality Working Party, Biologics Working Party and 
Committee for Advanced Therapies 
May 2019 
Adopted by CHMP for release for consultation 
29 May 2019 
Start of public consultation 
03 June 2019 
End of consultation (deadline for comments) 
31 August 2019 
Agreed by Biologics Working party 
April 2021 
Agreed by Committee for Advanced Therapies 
April 2021 
Agreed by Quality Working Party 
April 2021 
Adopted by CHMP for publication 
22 July 2021 
Date for coming into effect 
1 January 2022 
 
 
 
Keywords 
Medical device, medicinal products, Article 117, Notified Body opinion, Integral, 
Co-packaged, Referenced 
 
 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 2/22 
 
Guideline on quality documentation for medicinal products 
when used with a medical device 
Table of contents 
Executive summary ..................................................................................... 3 
1. Introduction ............................................................................................ 3 
2. Scope....................................................................................................... 5 
3. Legal references, Application of Standards and Guidelines ...................... 6 
4. General Considerations............................................................................ 6 
4.1. Submission of data, location in the dossier and format .............................................. 7 
4.2. Use of supporting data .......................................................................................... 8 
4.3. Scientific and Regulatory advice ............................................................................. 8 
5. Integral Medicinal Products ..................................................................... 8 
5.1. Module 1, Product Information ............................................................................... 8 
5.2. Module 3.2.P, Drug Product ................................................................................... 8 
5.3. Module 3.2.A.2, Adventitious Agents Safety Evaluation ........................................... 13 
5.4. Module 3.2.R, Regional Information, Medical Device ............................................... 13 
6. Medicinal Products with Co-packaged or Referenced Devices ................ 14 
6.1. Medicinal Products with Co-packaged Devices ........................................................ 14 
6.1.1. Module 1, Product Information .......................................................................... 15 
6.1.2. Module 3.2.P, Drug Product .............................................................................. 15 
6.1.3. Module 3.2.A.2, Adventitious Safety Evaluation ................................................... 17 
6.1.4. Module 3.2.R, Regional Information, Medical Device............................................. 17 
6.2. Medicinal Products with Referenced Devices ........................................................... 17 
7. Bridging to Devices used in Clinical Development ................................. 18 
8. Lifecycle Management ........................................................................... 18 
9. Emerging Technologies.......................................................................... 19 
10. Definitions ........................................................................................... 20 
Abbreviations ............................................................................................ 22 
 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 3/22 
 
Executive summary 
This guideline describes the information that should be presented in the Quality part of a marketing 
authorisation dossier for a medicinal product when it is used with a medical device, or device part, and 
submitted in accordance with Directive 2001/83/EC and/or Regulation (EC) 726/2004. This guideline 
focuses on product-specific quality aspects of a medical device, or device part, that may have an impact 
on the quality, safety and/or efficacy (and hence overall benefit/risk determination) of a medicinal 
product. 
This guideline applies in the following cases: 
• 
Medicinal products where the medical device and/or device part and the medicinal product form 
an integral product that is not reusable (hereafter called integral) and where the action of the 
medicinal product is principal, 
• 
Medicinal products placed on the market by the Marketing Authorisation Holder (MAH), where 
the medical device is packed together with the medicinal product (hereafter called co-packaged), 
or 
• 
Medicinal products, where the product information refers to a specific medical device to be used 
with the medicinal product, and the medical device is obtained separately by the user of the 
medicinal product (hereafter called referenced). 
1.  Introduction 
Increasingly, medicinal products are developed for use with a medical device, or device part; this ranges 
from a simple prefilled syringe to more advanced autoinjectors to medical devices used as sensors 
embedded in tablets. Given the wide range of medical devices, or device parts, that may be used with a 
medicinal product and continuous technological developments, the information provided in submissions 
has been found to be inconsistent and often incomplete. It is therefore appropriate within the scope of 
this guideline to provide guidance to quality assessors and the applicant/MAH of a medicinal product on 
the type of information that should be provided in a submission. 
This guideline considers three different, yet common, configurations of medicinal products used with 
medical devices, and describes the information that should be submitted to a Competent Authority (CA) 
for each of these configurations. 
This guideline also takes into consideration the amendment to Annex I, Directive 2001/83/EC 
(concerning supportive information to be submitted to CAs) introduced by the Medical Devices Regulation 
((EU)2017/745, MDR) by way of Article 117 (see Section 5.4 below). 
As discussed herein, depending on the configuration and the potential impact on the benefit-risk 
assessment of the medicinal product, the information that should be submitted will differ. 
The term ‘device (part)’, as used in this guideline, should be understood to refer to a medical device, or 
parts of a medical device, that are used in an integral or co-packaged configuration. 
In this guideline, the following terminology is used for the three configurations: 
Integral: a medical device (part) that falls under the second sub-paragraph of Article 1(8) or the second 
sub-paragraph of Article 1(9) of Regulation (EU) 2017/745, where the action of the medicinal product is 
principal: 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 4/22 
 
1. Devices that when placed on the market incorporate, as an integral part, a substance that, if used 
separately, would be considered as a medicinal product and has an action that is principal and not 
ancillary to the action of the device (second sub-paragraph of Article 1(8)). Examples include 
medicinal products with an embedded sensor where the sensor is a medical device and its action is 
ancillary to the medicinal product. 
2. Devices intended to administer a medicinal product, where the device and the medicinal product are 
placed on the market in such a way that they form a single integral product intended exclusively for 
use in the given combination and which is not reusable (second sub-paragraph of Article 1(9)). 
Typically, these devices have measuring or delivery functions. Examples of medical devices 
currently authorised for use in integral products include: 
• 
Single-use pre-filled syringes, single-use pre-filled pens and single-use pre-filled injectors 
(including autoinjectors) used for the delivery of one or more doses of medicine and which are 
not intended to be re-used or refilled once the initial doses provided are exhausted. 
• 
Drug-releasing intra-uterine devices and pre-assembled, non-reusable applicators for vaginal 
tablets. 
• 
Dry powder inhalers and pressurised metered dose inhalers that are preassembled with the 
medicinal product and ready for use with single or multiple doses but cannot be refilled when all 
doses are exhausted. 
• 
Implants containing medicinal products whose primary purpose is to release the medicinal product. 
For the integral configuration outlined above, the relevant General Safety and Performance Requirements 
(GSPRs, as set out in Annex I of the MDR) apply with respect to the safety and performance of the device 
(part). Integral products are discussed in Section 5 below. 
The guideline is also applicable to the following configurations where a medicinal product and a medical 
device are not placed on the market as a single integral unit, but the medicinal product is intended for 
use with a device. 
Co-packaged: A medicinal product and a medical device are packed together into a single pack (e.g. 
carton), which is placed on the market by the MAH, and 
Referenced: The product information (SmPC and/or package leaflet) of the medicinal product refers to 
a specific medical device to be used (e.g. identified by its brand name and/or specific description), and 
the specified medical device is obtained separately by the user of the medicinal product. 
In both co-packaged and referenced configurations, the medical device should comply with the 
requirements as laid down by the applicable medical device legal framework. Co-packaged and 
referenced products are discussed in Section 6 below, where the extent of information provided will vary 
according to the risks associated with the use of the device with the medicinal product. 
Some examples of medical devices co-packaged with, or specifically referenced by, authorised medicinal 
products include: 
• 
Oral administration devices (e.g. spoons, syringes). 
• 
Injection needles. 
• 
Refillable/reusable (e.g. using cartridges) pens and injectors (including autoinjectors). 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 5/22 
 
• 
Refillable/reusable dry powder inhalers and metered dose inhalers; spacers for inhalation sprays. 
• 
Nebulisers and vaporisers. 
• 
Single use or reusable pumps for medicinal product delivery. 
2.  Scope 
This guideline provides clarification regarding documentation for medicinal products in respect of a 
marketing authorisation application (MAA) or post-authorisation applications. This guideline considers 
the requirements, as laid down in Directive 2001/83/EC and/or Regulation (EC) 726/2004, where the 
action of the medicinal product is principal. This guideline also reflects current quality assessment 
practices and defines the additional documents/information for the three configurations described in this 
guideline. It applies to the following medicinal product types: chemical, biological or radiopharmaceutical. 
Section 5 of this guideline primarily considers integral products falling within the definition of second 
subparagraph of Article 1(9) of the MDR. However, it is recognised that products defined by the second 
subparagraph of Article 1(8) of the MDR will likely become more common-place as technology develops 
and medicinal products meeting this definition should follow the basic principles defined herein, 
recognising that elements of this guideline specifically referring to administration aspects may not be 
applicable. 
This guideline is not exhaustive, and the applicant/MAH (hereafter applicant) should consider all other 
relevant guidelines related to quality aspects of medicinal products. 
Medicinal products utilising electromechanical devices (including active implantable devices), electronic 
add-ons and digital elements of devices are in the scope of the guideline where they are expected to 
impact, even potentially, the benefit-risk assessment of the medicinal product from a quality perspective; 
however, in this regard, it is acknowledged that more detailed guidance may be required in future. 
Medicinal products intended to be used with a Class I medical device are also in scope, and requirements 
are dealt with specifically in the relevant sub-sections below. 
 
Advanced Therapy Medicinal Products (ATMPs) 
Article 117 does not apply in the case of combined ATMPs as defined under Article 2(1)(d) of Regulation 
(EC) No 1394/2007. 
The information related to the medical device part(s) of combined ATMPs are detailed in ATMP specific 
guidelines. Where ATMP specific guidelines do not indicate the location of the relevant information in the 
dossier, the principles in this guideline should be followed. 
The content of this guideline should be taken into consideration in the following cases: 
• 
Medical devices that are co-packaged with ATMPs. 
• 
Separately obtained devices which are referenced in the medicinal product’s product information 
because of their potential impact on the quality, safety and/or efficacy of the ATMP. 
For ATMPs, the content of the MAA may be adapted, provided that this is justified under a risk-based 
approach. 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 6/22 
 
For medical devices that are used as a container closure system for ATMPs, the applicant should provide 
evidence that the relevant GSPRs are met, for example EU Declaration of Conformity, EU certificate 
issued by a Notified Body (NB) or confirmation from the applicant in the form of summary information. 
Section 3.2.R may also include, if relevant, cross-reference to studies or additional information provided 
in 3.2.P sections. 
Information on medical devices used during surgical procedures for application, implantation or 
administration of an ATMP that may have an impact on the efficacy or safety of the ATMP (as per Annex 
I, Part IV, Section 5.2.1 of Dir2001/83/EC) is expected in the electronic Common Technical Document 
(eCTD) Module 5. 
 
The following are out of scope of this guideline: 
a) Veterinary products. 
b) In-vitro diagnostic devices, including companion diagnostics. 
c) System and procedure packs regulated under Article 22 of the MDR. 
d) General groups of devices where reference is directly made, or inferred, in the product information, 
(e.g. “using a syringe or “an infusion line”, etc…). 
e) Products falling under the first sub-paragraph of MDR Article 1(8). 
3.  Legal references, Application of Standards and Guidelines 
This guideline should be read in conjunction with: 
▪ 
Directive 2001/83/EC (the Medicinal Products Directive) and Regulation (EC) 726/2004 (as amended), 
▪ 
Regulation (EU) 2017/745 on Medical Devices amending Directive 2001/83/EC, Regulation (EC) No 
178/2002 and Regulation (EC) No 1223/2009, and repealing Council Directives 90/385/EEC and 
93/42/EEC. 
In addition, this guideline is presented without prejudice and should be read in conjunction with all other 
relevant directives and regulations forming part of the pharmaceutical acquis, the European 
Pharmacopeia and all relevant European Commission, ICH and CHMP guidelines, Q&A documents and 
other documents as linked to, or published on, the European Medicines Agency (EMA) website. 
4.  General Considerations 
As a general principle for medicinal products falling within the scope of this guideline, the assessment of 
the suitability of a device (part) for its intended purpose should take into account the relevant quality 
aspects of the device (part) in the context of its use with the medicinal product. The complexity of the 
device (part), relevant patient characteristics and user requirements, as well as the clinical setting or 
use environment, are also important aspects of the assessment process. The medicinal product dossier 
should include a discussion of the impact of the device (part) on the Quality Target Product Profile (QTPP), 
Critical Quality Attributes (CQA) and overall control strategy of the medicinal product. 
The amended provisions of point 12 of Section 3.2 of Annex I to Directive 2001/83/EC (introduced by 
Article 117 of the MDR), require applicants for a medicinal product, where a medical device (part) and 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 7/22 
 
the medicinal product form an integral product that falls under the second sub-paragraph Article 1(8) or 
the second sub-paragraph of Article 1(9) of the MDR, to submit the results of the assessment of the 
conformity of the device (part) with the relevant GSPRs set out in Annex I to the MDR (i.e. the EU 
Declaration of Conformity or the relevant EU certificate issued by a Notified Body). If the application 
dossier does not contain these results and where the conformity assessment of the device (part), if used 
separately, requires the involvement of a NB, the applicant is required to provide an opinion on the 
conformity of the device (part) with the relevant GSPRs, issued by a NB (known as Notified Body Opinion 
or NBOp). It is the responsibility of the applicant to ensure that the NB is appropriately designated for 
the issuance of such an opinion by reference to the NANDO website. 
Refer to Section 5.4 (3.2.R) below for further details of the documentation required to demonstrate 
conformity of the device with the relevant GSPRs in Annex I, MDR. 
The core precept of this guideline is that the CA responsible for the regulation of medicines will evaluate 
the device (part) specific aspects relevant to the quality safety and efficacy (and hence overall 
benefit/risk determination) of the medicinal product, and that, as applicable, the NB will assess the 
relevant GSPRs for the device (part). 
This guideline does not define the scope of the NB review, but rather clarifies the type and level of 
information to be provided as part of the medicinal product dossier, in order to minimise duplication of 
effort for regulatory authorities, notified bodies and applicants. 
This guideline clarifies the documentation that should be submitted in relation to the quality of the 
medicinal product, including the manufacturing and control methods thereof. It is not intended to, and 
cannot, address the obligations of the manufacturers of the device (part) in accordance with applicable 
device regulations. Co-packaged and referenced medical devices should be in compliance with the 
applicable medical device legal framework. 
Where requested or required by a CA, samples of the medicinal product to be placed on the market 
should be provided. For emerging technologies, and to aid assessment, provision of a sample at time of 
initial submission is strongly encouraged. 
4.1.  Submission of data, location in the dossier and format 
Relevant information on the device (part) should be presented in a clearly structured manner, following 
the eCTD format (Volume 2B Notice to Applicants Medicinal Products for Human Use – Presentation and 
Format of the Dossier). In Sections 5 and 6 below, guidance is provided on information that should be 
included in specific sections of Modules 1 and 3. To avoid repetition, it is acceptable to cross-reference 
between sections. 
For sections of the eCTD that are not specifically discussed in this guideline (e.g. P.5.2), no specific 
medicinal product-related aspects are foreseen and the information presented should be in line with the 
guidance given in the Notice to Applicants. 
Regarding Module 3, Section 3.2.P should contain information on the product-specific quality aspects 
related to the device (part) that may have an impact on the quality, safety and/or efficacy (and hence 
overall benefit/risk determination) of the medicinal product. Section 3.2.R should include relevant 
information related to the demonstration of compliance of the device (part) with MDR Annex I (see 
Section 5.4 for further guidance). 
Usability studies are multidisciplinary in nature and the location of these data is discussed in Section 5.4. 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 8/22 
 
4.2.  Use of supporting data 
On a case-by-case basis, and where pre-existing data on a medical device (part) used with an already 
approved medicinal product is available in an applicant’s product portfolio, this data could be provided 
as supportive data in a submission. Discussion and comprehensive justification for the use of this 
supportive data should be included in Module 3.2.P.2. 
4.3.  Scientific and Regulatory advice 
This guidance cannot cover all types of devices (parts) and/or future technological developments that 
may raise novel questions and/or require complex scientific assessment. Therefore, consideration should 
be given to seeking scientific and/or regulatory advice within the EU CA (medicines) network early in 
development, particularly for new and/or emerging technologies (see also Section 9). 
5.  Integral Medicinal Products 
5.1.  Module 1, Product Information 
The product information of integral medicinal products should follow the requirements of Directive 
2001/83/EC or Regulation (EC) No 726/2004, as applicable (see QRD [Quality Review of Documents] 
templates) and should not include any administrative information on the device (part). The SmPC, 
labelling and package leaflet should not include details of the device (part) manufacturer/authorised 
representative, CE mark (including NB number), device symbols, Unique Device Identifier (UDI) or 
references to device market surveillance reporting. 
Information of the device (part) which is necessary for the intended use of the integral medicinal 
product should be included in the relevant sections of the package leaflet and SmPC, as applicable (for 
details refer to the SmPC guideline & the QRD annotated template). 
5.2.  Module 3.2.P, Drug Product 
P.1 Description and Composition 
Concise information on integral medicinal products, and brief information on any additional medical 
device (part)/accessories provided and intended for use with the medicinal product, should be submitted, 
and their identity (e.g. type/version), description and function should be stated. 
P.2 Pharmaceutical Development 
This section of the dossier should summarise the information relevant to development of the specific 
medical device (part) integrated into the medicinal product, including the rationale for its selection in the 
specific sections of 3.2.P.2. A risk assessment summary for the medicinal product, aligned with relevant 
risk management principles in ICH Q9, should be presented. 
P.2.2 Drug Product 
The applicant should take into consideration the intended use of the medical device (part) and its 
functionality, suitability for use within the context of the medicinal product, the therapeutic indication 
and the relevant target patient population. 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 9/22 
 
The functional aspects of the medical device (part) should be qualified in line with its complexity and 
should include the rationale for the choice and optimisation of the relevant aspects of design and 
performance (such as dose-delivery performance and mechanical functionality of the device (part)). 
Where required (e.g. due to changes in medical device (part) design during development), summary 
bridging data (see Section 7) should be provided in this section of the dossier, with cross-references to 
relevant data in other modules of the eCTD, as appropriate. Data should be provided to demonstrate 
and justify the equivalence of the overall performance of the medicinal product prototype(s) used during 
pivotal clinical development with the medicinal product intended for marketing. 
P.2.3 Manufacturing Process Development 
A concise description of manufacturing process development for the integral medicinal product should 
be provided in line with relevant guidance; development and manufacturing processes for the medical 
device (part) are out of scope. The method of sterilisation of the integral medicinal product, and where 
applicable, the device (part) should be explained and justified. 
A clear narrative of the development history as it pertains to the integral medicinal product including a 
comparison of the manufacturing process of integral medicinal products from pivotal or bridging clinical 
studies to the commercial integral medicinal product should be presented. 
The development of the control strategy for the integral medicinal product manufacturing process should 
be described. 
P.2.4 Container Closure System (CCS) 
The following aspects of the development of the container closure system should be considered: 
Description and rationale for integral medicinal product 
A brief description of the container closure system should be presented, including, for example: 
▪ 
Rationale for the container/medical device (part) and materials of construction. 
▪ 
Critical functional parts e.g. dose-setting mechanism. 
▪ 
Features to enhance user safety for dose delivery (e.g. audible click), sharps injury prevention 
features, safety/lock-out features to prevent over-dosage, safe disposal information, etc. 
▪ 
For implantable/transdermal device (part), information on the matrix and reservoir, including 
mechanism of drug release etc. 
▪ 
If the device (part) includes a graduation marking, the requirements of the Quality of Medicines, 
Questions and Answers, on the EMA website should be considered. 
Functional Performance 
Those factors that may impact the functional performance of the device (part) should be discussed and 
considered, e.g. posology, method of administration, physiological factors of the patient population, 
requirements in other EU guidelines, CHMP recommendations, etc. 
Functional performance aspects of the integral medicinal product should be limited to those aspects that 
impact quality, safety and/or efficacy (and hence overall benefit/risk determination) e.g. dose accuracy 
and precision over the range of (re)use, mechanical functionality and/or other aspects directly relevant 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 10/22 
 
to the intended use of the device (part) as part of the medicinal product. In particular, the ability of the 
device (part) to deliver/administer the medicinal product in accordance with the posology stated in 
Section 4.2 of the SmPC in an accurate and reproducible manner should be demonstrated. In this regard, 
the following should be considered: 
▪ 
As far as possible, test conditions should simulate the use of the integral medicinal product under 
relevant (in-use) storage conditions. This need not necessarily include simulation of interaction with 
the human body. 
▪ 
Consistency of dosing (demonstrating that the medical device (part) accurately and precisely delivers 
the intended dose) should be guaranteed from release until the end of the shelf life, and also during 
the use of the particular integral medicinal product under the conditions recommended in the SmPC 
(in-use stability testing). Where transportation studies are not presented in P.3.5, justification should 
be provided. 
▪ 
The effects of simulated real-world in-use conditions should be discussed, and data provided as 
appropriate e.g. shaking, priming, dropping test. 
For the above, such studies should be conducted using the intended medicinal product. If the intended 
medicinal product is not used, the approach should be justified. Furthermore, the studies should be 
conducted on the device (part) variant to be commercialised, unless otherwise justified. As applicable, 
conformance to relevant standards should be confirmed. 
For usability studies, see Section 5.4 (3.2.R), below. 
P.2.5 Microbiological Attributes 
For sterile products, the integrity of the integral medicinal product throughout use and shelf-life, as it 
relates to preventing microbial contamination, should be demonstrated. 
P.2.6 Compatibility 
Compatibility between all materials in contact with the drug product, including any diluents for 
reconstitution, should be investigated to provide appropriate and supportive information. The following 
aspects should be considered: 
▪ 
All materials in contact with the medicinal product during administration/delivery. Unless otherwise 
justified, interaction studies should be performed, assessing physical and chemical compatibility (e.g. 
sorption, precipitation of drug substance in solution, stability, extractable and leachables, etc., as 
appropriate). Studies should demonstrate no impact on the safety, efficacy and quality of the 
medicinal product. 
▪ 
If processing aids (e.g. lubricants, glue/adhesive from labels etc.) are used with the medical device 
(part) and come into direct contact with the drug product, interaction studies should be performed to 
evaluate their effects on the drug product as well as on the performance of the device (part), unless 
otherwise justified. For example, silicone oils released from the device (part) can nucleate the 
formation of proteinaceous particles/aggregates with protein products. Toxicological assessments of 
processing aids that are in direct contact with the drug product should be performed, as necessary. 
▪ 
Compatibility should be considered from a chemical and physical stability perspective i.e. under 
justified worst-case orientation(s), in-use conditions and during simulated transportation studies. 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 11/22 
 
▪ 
The suitability of the device (part) for the particular drug product (e.g. considering the rheological 
properties of the drug product) should be discussed and justified. 
P.3 Manufacture 
P.3.1 Manufacturers 
Names, addresses and responsibilities for the integral medicinal product manufacturer and batch release 
sites should be stated, including, as appropriate, sites for assembly of the integral medicinal product, 
packaging, sterilisation, labelling and quality control. Suppliers of the device (part), and the sites 
responsible for manufacture or commercial sub-assembly of the device (part) alone need not be stated 
in this section. 
P.3.3 Description of manufacturing process and process controls 
The description of the manufacturing process of the integral medicinal product should include operations 
relating to the integration of device (part) and drug product. Critical processes, technologies and/or 
packaging operations that directly affect product quality should be described in detail. 
The following information should be included: 
▪ 
Appropriate description of any manufacturing operations that are performed by the integral medicinal 
product manufacturer to prepare the medical device (part) for the final assembly of the integral 
medicinal product such as subassembly steps, washing, coating, sterilisation, depyrogenation etc. 
▪ 
Description of the integral medicinal product manufacturer sterilisation methods and conditions, 
where relevant. Information on sterilisation of medical device (part) should be presented in this 
section of the dossier or reference given to another appropriate section in line with relevant EMA 
guidelines. 
▪ 
When empty, sterile, ready-to-use container closure components are purchased, information should 
be provided in line with relevant EMA guidelines. The inclusion of an EU certificate issued by a Notified 
Body or NBOp including review of sterility is sufficient to assure sterility where applicable. This should 
be presented in 3.2.R. 
▪ 
A description of the filling steps and the final assembly of the integral medicinal product, as performed 
by the manufacturer should be detailed together with critical process parameters and in-process 
controls. 
▪ 
Applied labels with a functional role should be described and, when applicable and considered critical, 
acceptable tolerances for label positioning should be defined. 
P.3.4 Controls of critical steps and intermediates 
Any critical steps should be justified and critical process parameters, process parameters whose impact 
on quality cannot reasonably be excluded and in-process controls for the manufacture of the integral 
medicinal product should be stated, with acceptance criteria. Any device-(part) specific intermediates 
(sub-assemblies) produced during manufacture of the integral medicinal product should be defined, 
along with relevant specifications, test methods and their validation. Any holding times should be defined 
and justified. 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 12/22 
 
P.3.5 Process validation and/or evaluation 
Process validation for the integral medicinal product manufacturing process should be performed, as 
appropriate, in line with relevant European guidelines, including the assembly and sterilisation of the 
device (part) (if applicable) and any filling steps. 
Actual transportation or simulated transport studies that encompass chemical (e.g. degradation under 
different conditions) and physical (e.g. vibration) aspects to demonstrate stability during transportation 
should be investigated, where relevant. 
P.5 Control of drug product 
P.5.1 Specification(s) 
The specification should include those parameters that have been identified as CQAs of the medicinal 
product and that are controlled in the integral medicinal product. When appropriate, the specification 
should include the following: 
▪ 
Description/appearance. 
▪ 
Performance and functionality tests relevant to the intended use of the integral medicinal product 
e.g. delivered dose uniformity (at release and shelf life) or extractable volume. 
▪ 
Other critical test parameters related to CQAs of the integral medicinal product where earlier controls 
are not in place, e.g. glide force, needle penetration force, delivery time, exposed needle length after 
activation of device (part), activation force, lock-out system control to prevent over-dosing, etc. 
P.7 Container closure system (CCS) 
Where the device (part) is part of the primary container closure system (immediate packaging) as 
intended for marketing (e.g. staked-in needle) or forms part of the functional secondary packaging (e.g. 
pen subassemblies enclosing a pre-filled syringe), the following information should be provided: 
▪ 
A description of the container closure system, including primary packaging and secondary packaging 
devices. 
▪ 
Suitable quality control specifications of the device (part). Specifications should reflect the complexity 
of the device (part), such that equivalence between devices (parts) from different suppliers can be 
assured. Unless justified, the supplier of device (part) should be defined in this section. 
▪ 
Detailed specifications (including description, identification and functional tests as relevant) as well 
as critical dimensions (with drawings and photographs where appropriate) of primary and functional 
secondary packaging materials should be provided. Test procedures should be presented, as 
appropriate.  
▪ 
For device (part) not in direct contact with the drug product, information commensurate with its 
functionality for the correct use of the integral medicinal product should be provided. The secondary 
packaging should take into account the intended user population of the integral medicinal product. 
▪ 
Primary packaging materials of construction should be described and comply with the relevant Ph. 
Eur. monographs, if applicable, and food contact directives, as appropriate (e.g. evidenced by 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 13/22 
 
declarations of compliance from suppliers). Where a Ph. Eur. monograph is not available, alternative 
standards may be referenced with justification. 
▪ 
Where a material of construction does not meet compendial standards and is used for the first time 
in a medicinal product, a specification should be included and justified with safety data (as 
appropriate). 
P.8 Stability 
Stability studies for the integral medicinal product (or variant, where justified) should include the 
following tests/studies: 
▪ 
Functionality tests determined as stability-indicating CQAs for the medicinal product (refer to P.2.4). 
▪ 
In-use stability testing (including relevant functionality tests) performed under the conditions of use 
as stated in the SmPC, unless otherwise justified. 
▪ 
Tests for identified stability-indicating CQAs (e.g. microbial quality, sterility, container closure 
integrity, content/potency and purity) over the shelf-life and in-use period, as appropriate for the 
integral medicinal product in question. If necessary, appropriate and scientifically justified alternatives 
for sterility (such as container closure integrity testing) may be used. 
5.3.  Module 3.2.A.2, Adventitious Agents Safety Evaluation 
All materials of human or animal origin used in the manufacturing process of the integral medicinal 
product should be identified. Information assessing the risk with respect to potential contamination with 
adventitious agents of human or animal origin should be provided. 
TSE agents 
Where appropriate, a TSE statement confirming compliance of the part(s) of the integral medicinal 
product with current, relevant European guidance and standards should be provided. 
Viral safety 
Where applicable, an assessment of the risk to the integral medicinal product with respect to potential 
viral contamination should be provided. The viral risk assessment should be performed in accordance 
with the current, relevant European guidance and standards. 
For substances from human blood/plasma, compliance with relevant EU directives, Ph. Eur. and EMA 
guidelines should be verified.  
Other adventitious agents 
Detailed information regarding other adventitious agents, such as bacteria, mycoplasma and fungi should 
be provided in relevant sections pertaining to the device (part) within the core dossier, as appropriate. 
5.4.  Module 3.2.R, Regional Information, Medical Device 
In accordance with Article 117 of the MDR, all applications for an integral medicinal product should 
include evidence of the conformity of the device (part) with the relevant GSPRs set out in Annex I of 
Regulation (EU) 2017/745. Refer to the EMA Q&A on Implementation of the Medical Devices and In Vitro 
Diagnostic Medical Devices Regulations ((EU) 2017/745 and (EU) 2017/746) for further details. 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 14/22 
 
Notified Body Opinion (NBOp) 
The processes by which a NB derives their opinion are not within the scope of this guideline. 
The assessment of the integral medicinal product can be facilitated when the NBOp is presented as a 
summary technical report. This enables both assessor and applicant to determine how the opinion was 
derived, minimise duplication and avoid gaps in assessment, and to identify aspects to be considered 
during the MAA/variation. 
Usability studies 
A usability study should be provided in the following situations: 
1. Where supporting information is not included in the dossier, and the device (part) has not been used 
in the intended user population before, taking specific limitations due to the indication and/or disease 
into account, or 
2. Where other aspects of the intended use, including changes to the clinical setting or use environment 
are new or different from the intended use as confirmed by the EU certificate issued by a Notified 
Body or NBOp. 
In these cases, and in the cases where no NBOp is needed, detailed information on use-related risks and 
results of usability studies (or justification for their absence) should be presented in Module 3.2.R (cross-
referencing data in Module 5, if relevant). Usability is considered a multidisciplinary topic and may also 
be assessed outside of quality considerations. 
Where evidence of usability is required, this may be supported by published and/or other relevant data 
for an identical/equivalent device (part) on the market. In this case, relevant data could be bridging data 
to an identical/equivalent device (part) used in different patient populations. However, if usability cannot 
be satisfactorily demonstrated in this way, a formal usability study is required to demonstrate safe and 
effective use of the integral medicinal product by the intended user population. Applicants are 
encouraged to follow/use relevant EU harmonised standards to demonstrate compliance. 
Use of Supportive Data 
Detailed information pertaining to the discussion presented in Section 4.2 above should be presented in 
this section. 
6.  Medicinal Products with Co-packaged or Referenced 
Devices 
Where a medical device is co-packaged or referenced in the product information of the medicinal product, 
additional information on the device may be needed. The extent of the additional information will depend 
on the specifics of the device, its intended use and the risks thereof to the quality, safety and/or efficacy, 
and hence overall benefit/risk determination of the medicinal product under consideration (e.g. 
compatibility, extractables and leachables, etc). There are separate sections for devices that are co-
packaged and for those that are referenced. 
6.1.  Medicinal Products with Co-packaged Devices 
This configuration includes products where a co-packaged medical device is intended to be used only 
for the administration and/or application of the co-packaged medicinal product. 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 15/22 
 
6.1.1.  Module 1, Product Information 
The product information of a medicinal product including a co-packaged device should follow the 
requirements of Directive 2001/83/EC or Regulation (EC) No 726/2004, as applicable (see QRD 
templates) and should not include any administrative information on the medical device. The SmPC, 
labelling and package leaflet should not include details of the device manufacturer/authorised 
representative, CE mark (incl. NB number), device symbols, UDI or references to device market 
surveillance reporting.  
Relevant information for the use of the co-packaged device, especially if necessary for the intended use 
of the medicinal product with the device should be included in the appropriate sections of the medicinal 
product package leaflet and SmPC, as applicable (for details, refer to the SmPC guideline & the QRD 
annotated template). 
The MAH of the medicinal product is responsible for the co-packaged medicinal product and its 
traceability (including the co-packaged device), and hence only the contact details of the MAH should 
be on the package leaflet/labelling. 
The co-packaged device must also be in conformity with the MDR, which includes compliance with 
labelling requirements specific to the device. 
6.1.2.  Module 3.2.P, Drug Product 
P.1 Description and Composition 
A brief description, including function and identity (e.g. type/version) of any device(s) with which the 
medicinal product will be used or administered and which will be supplied along with the medicinal 
product should be stated. 
P.2 Pharmaceutical Development 
This section of the dossier should summarise the information on the proposed medical device relevant 
to development of the medicinal product, in the appropriate sections of 3.2.P.2. The use of a medicinal 
product with a specified device should be demonstrated to be safe and effective. The amount of 
information provided in this section should reflect the risk of the device to impact the quality, safety 
and/or efficacy (and hence overall benefit/risk determination) of the medicinal product. In certain cases, 
the impact of the method of sterilisation of the device (e.g. ethylene oxide) on product quality should be 
understood. 
The information provided in P.2 is not expected to be as detailed as the information provided in 3.2.R 
for the device (i.e. cross-referencing with relevant sections of 3.2.R is possible). 
P.2.2 Drug Product 
A general discussion on the choice of device should be provided, including the intended use, rationale 
for choice of device, etc. 
The functional aspects of the device should be qualified in line with its complexity and should include the 
rationale for the choice and optimisation of the design and performance (such as dose-delivery 
performance and mechanical functionality of the device). Dose accuracy/delivered dose uniformity should 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 16/22 
 
be demonstrated with the medicinal product. Any markings/graduation should be justified in line with 
the posology stated in Section 4.2 of the SmPC. Details of the cleaning of the device(s) should be stated 
in Section 6.6. of the SmPC, where relevant and in the Package Leaflet. 
Where changes are made to the device during clinical development, refer to Section 7. 
P.2.5 Microbiological Attributes 
For medicinal products intended to be used in a sterile manner, the sterility of the device is assumed 
as long as the device is supplied in its primary packaging with the sterile barrier intact. Maintenance of 
sterility of the medicinal product throughout its in-use period, and under the conditions of its intended 
use, should be discussed and justified. 
P.2.6 Compatibility 
Unless otherwise justified, compatibility between device and drug product should be investigated: 
▪ 
Compatibility should be considered from an in-use stability perspective and the physical and chemical 
compatibility of the drug product with the device(s) should be demonstrated (e.g. sorption, 
precipitation of drug substance in solution, stability, etc.). Interaction studies should be performed, 
as appropriate, using a risk-based approach. All materials in contact with the drug product should be 
considered in an appropriate manner and having regard of contact time. 
▪ 
The suitability of the device for the particular drug product (e.g. considering the rheological properties 
of the product) should be discussed and justified. 
P.7 Container Closure System 
The device is not part of the container closure system and thus only a brief description of the device 
should be provided in this section (for example; “1 mL glass syringe including 0.05 mL marked 
graduations”, along with the name and/or base-UDI of the device). Where appropriate, a specification, 
as applied to the incoming device upon receipt by the medicinal product manufacturer, should be 
presented. For further details, reference should be made to the information in 3.2.R, including evidence 
of compliance with the applicable medical device legal framework. 
Where the co-packaged medical device becomes part of the container closure system prior to 
administration or application (for example a finished product transfer device attached to a vial) the device 
should prevent microbial ingress and maintain the sterility of the medicinal product throughout its use. 
P.8 Stability 
If relevant, in-use stability data should be provided for the drug product in contact with the device, 
including parameters for device functionality that may impact the quality, safety and/or efficacy (and 
hence overall benefit/risk determination) of the medicinal product. The shelf-life of the medicinal product 
should be set based on the shortest expiry date of all components of the medicinal product and a co-
packaged administration or application device(s) marketed as a single unit. 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 17/22 
 
6.1.3.  Module 3.2.A.2, Adventitious Safety Evaluation 
An EU certificate or EU Declaration of Conformity can be accepted as evidence of compliance of the 
device (part) with EU requirements. 
6.1.4.  Module 3.2.R, Regional Information, Medical Device 
Evidence should be provided that relevant standards have been met e.g. EU Declaration of Conformity 
or, where applicable, EU certificate, or other appropriate documentation such as summary information 
confirming compliance with relevant GSPRs. 
Where required and applicable (e.g. owing to changes in device design), summary bridging data should 
be provided in this section of the dossier, with cross-reference to relevant data in Module 4 or Module 5, 
as appropriate (see Section 7). 
If the device has not been used in the proposed patient population before, or if the setting of use is new, 
a usability study is expected. Evidence of usability may be supported by published or other relevant data 
for identical/similar devices on the market. Relevant data could also include bridging data to similar 
devices or to the same devices used in different patient populations. However, if usability cannot be 
satisfactorily demonstrated in this way, a formal usability study is required to demonstrate safe and 
effective use of the medicinal product by the intended user population. Applicants are encouraged to 
follow/use relevant EU harmonised standards to demonstrate compliance. 
A summary should be provided in Module 3.2.R, cross-referring to Module 5, where information on 
usability studies (or justification for their absence) should be presented. 
Discussion of, and justification for the use of supportive data should be included in this section (for 
further detail, see Section 4.2 above). 
6.2.  Medicinal Products with Referenced Devices 
Where a medical device is referenced in the product information of the medicinal product and may have 
a potential impact on the quality, safety and/or efficacy (and hence overall benefit/risk determination) 
of the medicinal product, additional information may be required. The requirements in this section do 
not apply where reference is made to a general group of devices (e.g. “using a syringe” or “an infusion 
line”, etc.). 
The impact of the referenced device on the medicinal product when used together should be considered 
using a risk-based approach, taking into account the need for a usability study. Evidence of safety and/or 
bioequivalence/efficacy for the medicinal product used with the device should be provided as appropriate. 
The product information should be sufficiently detailed to ensure correct use of the medicinal product 
with the referenced device. Refer to Section 6.1.1 above. 
In section 3.2.P.2, data on compatibility, dosing accuracy, functionality, handling, manipulation, etc., 
should be presented as appropriate. 
In section 3.2.P.8, in-use stability data should be presented, if applicable. 
Information on usability studies should be presented, unless otherwise justified (see Section 6.1.4 above). 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 18/22 
 
7.  Bridging to Devices used in Clinical Development 
While authorisation of clinical trials is within the competence of national CAs (and hence outside the 
remit of this guideline), in the context of a MAA, the following guidance is provided: 
• 
Given the (often) critical contribution that a device makes to the safe and effective administration or 
application of a medicinal product, the device (part) should be as advanced as possible in the 
development process (e.g. meets relevant GSPRs) by the time pivotal clinical trials (that include the 
device (part)) commence. This will subsequently reduce the potential need for extensive bridging 
data at time of marketing authorisation. 
• 
The impact of any changes in the device during pivotal clinical trials should be described, evaluated 
and justified in terms of any potential impact on the quality, safety and/or efficacy (and hence overall 
benefit/risk determination) of the medicinal product as applicable for the particular configuration 
(e.g. delivered dose, needle penetration force for subcutaneous/intramuscular injection, usability 
factors). Where changes are made, data to bridge the changes from a quality, safety and/or efficacy 
perspective may be required. Appropriate data should be provided to demonstrate and justify the 
equivalence of the overall performance during clinical phases with that at time of MAA. 
• 
Quality-relevant aspects should be discussed in Module 3.2.P.2.2, and should describe the changes, 
the batches used and trial(s) affected, as well as the mitigation measures applied to ensure that the 
impact on product quality was minimal.  
8.  Lifecycle Management 
The MAH should ensure that they are aware of any changes made to the device (part) used in an integral 
or co-packaged medicinal product. Depending on the nature of the change, the MAH should determine 
whether updates to relevant documentation in the eCTD (e.g. specifications, etc.) are required. 
Variations to the dossier related to a device (part) should be submitted in accordance with Commission 
Regulation (EC) No 1234/2008 concerning the examination of variations to the terms of marketing 
authorisations for medicinal products for human use, and associated variation guidelines. 
The category of variation should take into consideration the impact of the change, particularly where a 
change impacts any medicinal product CQA and/or any element of the overall medicinal product control 
strategy. In cases where the need for a variation is unclear and/or the category of the change is unclear, 
it is recommended that the medicines CA that issued the marketing authorisation be consulted to agree 
the category prior to submission of the variation application. 
The MAH should also consider whether changes to the medicinal product (e.g. changes to volume, 
viscosity, etc.) may impact the performance of the device (part), in such a way that its use may require 
further verification and/or validation. In addition, changes in the intended use or target population may 
require an additional usability study. 
Additional considerations 
In cases where a variation application is submitted to change, replace or add a device (part) or other 
aspect of the user interface, consideration should be given to whether there is an impact on the product 
information that may affect the potential risk of medication errors. The overall risk assessment of the 
medicinal product should be updated accordingly, with consideration given to the following: 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 19/22 
 
▪ 
The need for communication with patients and/or health care professionals regarding the change. 
▪ 
Where differences exist, the risks of potential medication errors should be assessed. Additional 
usability studies may be required, as determined by the outcome of this risk assessment. 
▪ 
If there is a risk of a medication error because of the change, this may need to be captured in the 
Risk Management Plan (RMP). 
9.  Emerging Technologies 
It is recognised that there are rapid developments in science and technology for both medical device 
(part) and medicinal products. Whilst this guideline discusses information that should be provided in an 
MAA or post-authorisation application, alternative approaches for emerging technologies may be 
acceptable, where adequately justified. 
For a medicinal product utilising emerging technologies, it is recommended to engage with the relevant 
medicines CAs in a timely manner, e.g. by requesting formal scientific advice, or through an Innovation 
Office, etc. It is also recommended to identify a NB and engage in discussions in a timely manner. 
 
 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 20/22 
 
10.  Definitions 
Applicant 
The commercial entity responsible for the marketing authorisation application or post-approval 
application in the EU. 
ATMP 
Advanced Therapy Medicinal Products (ATMPs, as per Regulation 2007/1394 EC) are medicines for 
human use that are based on genes, tissues or cells. 
Control Strategy 
A planned set of controls derived from current product and process understanding that ensures process 
performance and product quality. The controls can include parameters and attributes related to drug 
substance, drug product, medical device (part), facility and equipment operating conditions, in-process 
controls, finished product specifications, and the associated methods and frequency of monitoring and 
control. 
Container Closure System (CCS) 
The sum of parts that together contain and/or protect the drug product. This includes immediate and 
secondary packaging, where the latter is intended to provide additional protection to the drug product. 
The container closure system may include medical devices (or device parts), as defined in Section 1 of 
this guideline. 
Device (synonym: Medical Device) 
A device that fulfils the definition of a medical device, as defined in the relevant EU legal framework, and 
is intended to be placed on the EU market or made available on the EU market. This also applies to part 
of a device that fulfils the definition of Article 2(1) MDR. 
Dossier 
The complete body of data submitted for regulatory assessment. In this case, the dossier refers to the 
administrative and quality constituents of the eCTD, i.e. Module 1 (administrative), Module 2 (Overall 
Summaries) and Module 3 (quality) respectively, and (typically) more specifically to the content of 
Module 3. 
Marketing Authorisation Holder (MAH) 
The company that has been granted a marketing authorisation for a medicinal product by a competent 
authority of a member state in accordance with Directive 2001/83/EC (as amended) or by the European 
Commission in accordance with Regulation (EC) No 726/2004, and is responsible for marketing the 
medicinal product. 
Medical Device Manufacturer 
The commercial entity manufacturing and supplying sterile/non-sterile devices and/or device parts to 
the medicinal product manufacturer. 
Medicinal Product Manufacturer 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 21/22 
 
The physical location and commercial entity legally responsible for the manufacture of the integral and/or 
co-packaged medicinal product. 
Notified Body Opinion (NBOp) 
An opinion provided by a Notified Body on the conformity of a device (part) with the relevant GSPRs set 
out in Annex I of Regulation 2017/745, as required by Article 117 of the MDR. 
Usability 
The level to which a medicinal product can be handled in accordance with the product information in the 
different settings where it may be used, taking into account the variety of patient characteristics, the 
risk for medication errors and the impact to the patient and caregiver’s quality of life, etc. Therefore, 
this definition includes (and is not limited) to that presented in IEC 62366. 
 
 
 
 
EMA/CHMP/QWP/BWP/259165/2019 
Page 22/22 
 
Abbreviations 
ATMP 
 
Advanced Therapy Medicinal Product 
CA 
 
Competent Authority (for the regulation of medicines, either National or EMA) 
CCS 
 
Container Closure System 
CHMP  
Committee for Human Medicinal Products 
CQA 
 
Critical Quality Attribute 
eCTD 
 
electronic Common Technical Document 
EMA 
 
European Medicines Agency 
GSPR 
 
General Safety and Performance Requirement 
ICH 
International Council for Harmonisation of Technical requirements for Pharmaceuticals 
for Human Use 
MAA 
 
Marketing Authorisation Application 
MAH 
 
Marketing Authorisation Holder 
MDR 
 
Medical Device Regulation (EC 2017/745) 
NB 
 
Notified Body 
NBOp 
 
Notified Body Opinion 
Q&A 
 
Questions and Answers 
QRD 
 
Quality Review of Documents group, EMA 
QTTP 
 
Quality Target Product Profile 
SmPC  
Summary of Product Characteristics 
TSE 
 
Transmissible Spongiform Encephalopathy 
UDI 
 
Unique Device Identifier 