EMA/CHMP/ICH/172948/2019 
Committee for Medicinal Products for Human Use 
ICH guideline M10 on bioanalytical method validation and 
study sample analysis 
Step5 
Transmission to CHMP 
28 February 2019 
Adoption by CHMP 
28 February 2019 
Release for public consultation 
14 March 2019 
End of consultation (deadline for comments) 
1 September 2019 
Final adoption by CHMP 
21 July 2022 
Date of coming into effect 
21 January 2023 
 
 
 
 
ICH guideline M10 on bioanalytical method validation and study sample analysis  
 
EMA/CHMP/ICH/172948/2019  
Page 2/45 
 
M10 Document History 
Code 
History 
Date 
M10 
Endorsement by the Members of the ICH Assembly under 
Step 2 and release for public consultation (document dated 
15 January 2019). 
26 February 2019 
M10 
Endorsement by the Regulatory Members of the ICH 
Assembly under Step 4, including a title change from 
“Bioanalytical Method Validation” to “Bioanalytical Method 
Validation and Study Sample Analysis” 
24 May 2022 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Legal notice: This document is protected by copyright and may, with the exception of the ICH logo, 
be used, reproduced, incorporated into other works, adapted, modified, translated or distributed under 
a public license provided that ICH's copyright in the document is acknowledged at all times. In case of 
any adaption, modification or translation of the document, reasonable steps must be taken to clearly 
label, demarcate or otherwise identify that changes were made to or based on the original document. 
Any impression that the adaption, modification or translation of the original document is endorsed or 
sponsored by the ICH must be avoided. 
The document is provided "as is" without warranty of any kind. In no event shall the ICH or the 
authors of the original document be liable for any claim, damages or other liability arising from the use 
of the document. 
The above-mentioned permissions do not apply to content supplied by third parties. Therefore, for 
documents where the copyright vests in a third party, permission for reproduction must be obtained 
from this copyright holder. 
 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 3/45 
 
ICH guideline M10 on bioanalytical method validation 
Table of contents 
1. Introduction ............................................................................................ 5 
1.1. Objective............................................................................................................. 5 
1.2. Background ......................................................................................................... 5 
1.3. Scope ................................................................................................................. 5 
2. General .................................................................................................... 6 
2.1. Method development ............................................................................................ 6 
2.2. Method validation ................................................................................................. 6 
2.2.1. Full validation ................................................................................................... 6 
2.2.2. Partial validation ................................................................................................ 7 
2.2.3. Cross validation ................................................................................................. 7 
3. Chromatography ...................................................................................... 7 
3.1. Reference standards ............................................................................................. 7 
3.2. Validation ............................................................................................................ 8 
3.2.1. Selectivity ........................................................................................................ 8 
3.2.2. Specificity ......................................................................................................... 8 
3.2.3. Matrix effect ..................................................................................................... 9 
3.2.4. Calibration curve and range ................................................................................ 9 
3.2.5. Accuracy and precision ..................................................................................... 10 
3.2.6. Carry-over ...................................................................................................... 11 
3.2.7. Dilution integrity ............................................................................................. 11 
3.2.8. Stability ......................................................................................................... 11 
3.2.9. Reinjection reproducibility................................................................................. 13 
3.3. Study sample analysis ........................................................................................ 14 
3.3.1. Analytical run .................................................................................................. 14 
3.3.2. Acceptance criteria for an analytical run ............................................................. 14 
3.3.3. Calibration range ............................................................................................. 15 
3.3.4. Reanalysis of study samples ............................................................................. 16 
3.3.5. Reinjection of study samples ............................................................................. 17 
3.3.6. Integration of chromatograms ........................................................................... 17 
4. Ligand binding assays ........................................................................... 17 
4.1. Key reagents ..................................................................................................... 17 
4.1.1. Reference standard .......................................................................................... 17 
4.1.2. Critical reagents .............................................................................................. 17 
4.2. Validation .......................................................................................................... 18 
4.2.1. Specificity ....................................................................................................... 18 
4.2.2. Selectivity ...................................................................................................... 19 
4.2.3. Calibration curve and range .............................................................................. 19 
4.2.4. Accuracy and precision ..................................................................................... 20 
4.2.5. Carry-over ...................................................................................................... 21 
4.2.6. Dilution linearity and hook effect ....................................................................... 21 
4.2.7. Stability ......................................................................................................... 21 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 4/45 
 
4.3. Study sample analysis ........................................................................................ 22 
4.3.1. Analytical run .................................................................................................. 22 
4.3.2. Acceptance criteria for an analytical run ............................................................. 23 
4.3.3. Calibration range ............................................................................................. 23 
4.3.4. Reanalysis of study samples ............................................................................. 23 
5. Incurred sample reanalysis (ISR) .......................................................... 24 
6. Partial and cross validation ................................................................... 26 
6.1. Partial validation ................................................................................................ 26 
6.2. Cross validation ................................................................................................. 27 
7. Additional considerations ...................................................................... 27 
7.1. Methods for analytes that are also endogenous molecules ....................................... 27 
7.1.1. Quality control samples for methods for analytes that are also endogenous molecules
 .............................................................................................................................. 28 
7.1.2. Selectivity, recovery and matrix effects for methods for analytes that are also 
endogenous molecules .............................................................................................. 29 
7.1.3. Parallelism for methods for analytes that are also endogenous molecules ............... 29 
7.1.4. Accuracy and precision for methods for analytes that are also endogenous molecules
 .............................................................................................................................. 30 
7.1.5. Stability for methods for analytes that are also endogenous molecules ................... 30 
7.2. Parallelism ......................................................................................................... 30 
7.3. Recovery ........................................................................................................... 30 
7.4. Minimum required dilution ................................................................................... 31 
7.5. Commercial and diagnostic kits ............................................................................ 31 
7.6. New or alternative technologies ........................................................................... 32 
7.6.1. Dried matrix methods ...................................................................................... 32 
8. Documentation ...................................................................................... 32 
8.1. Summary information ......................................................................................... 33 
8.2. Documentation for validation and bioanalytical reports ............................................ 33 
9. Glossary ................................................................................................ 41 
 
 
 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 5/45 
 
1.  Introduction 
1.1.  Objective 
This guideline is intended to provide recommendations for the validation of bioanalytical methods for 
chemical and biological drug quantification and their application in the analysis of study samples. 
Adherence to the principles presented in this guideline will ensure the quality and consistency of the 
bioanalytical data in support of the development and market approval of both chemical and biological 
drugs.  
The objective of the validation of a bioanalytical method is to demonstrate that it is suitable for its 
intended purpose. Changes from the recommendations in this guideline may be acceptable if 
appropriate scientific justification is provided. Applicants are encouraged to consult the regulatory 
authority(ies) regarding significant changes in method validation approaches when an alternate 
approach is proposed or taken. 
1.2.  Background 
Concentration measurements of chemical and biological drug(s) and their metabolite(s) in biological 
matrices are an important aspect of drug development. The results of studies employing such methods 
contribute to regulatory decisions regarding the safety and efficacy of drug products. It is therefore 
critical that the bioanalytical methods used are well characterised, appropriately validated and 
documented in order to ensure reliable data to support regulatory decisions.  
This guideline intends to facilitate development of drugs in accordance with the principles of 3Rs 
(Reduce, Refine, Replace) for animal studies, where valid. 
1.3.  Scope 
This guideline describes the validation of bioanalytical methods and study sample analysis that are 
expected to support regulatory decisions. The guideline is applicable to the bioanalytical methods used 
to measure concentrations of chemical and biological drug(s) and their metabolite(s) in biological 
samples (e.g., blood, plasma, serum, other body fluids or tissues) obtained in nonclinical toxicokinetic 
(TK) studies conducted according to the principles of GLP, nonclinical pharmacokinetic (PK) studies 
conducted as surrogates for clinical studies, and all phases of clinical trials, including comparative 
bioavailability/bioequivalence (BA/BE) studies, in regulatory submissions. Full method validation is 
expected for the primary matrix intended to support regulatory submissions. Additional matrices 
should be validated as necessary.  
For studies that are not submitted for regulatory approval or not considered for regulatory decisions 
regarding safety, efficacy or labelling (e.g., exploratory investigations), applicants may decide on the 
level of qualification that supports their own internal decision making.  
The information in this guideline applies to the quantitative analysis by ligand binding assays (LBAs) 
and chromatographic methods such as liquid chromatography (LC) or gas chromatography (GC), which 
are typically used in combination with mass spectrometry (MS) detection. 
For studies that are subject to Good Laboratory Practice (GLP) or Good Clinical Practice (GCP) the 
bioanalysis of study samples should also conform to their requirements. 
The bioanalysis of biomarkers and bioanalytical methods used for the assessment of immunogenicity 
are not within the scope of this guideline. 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 6/45 
 
2.  General  
2.1.  Method development 
The purpose of bioanalytical method development is to define the design, operating conditions, 
limitations and suitability of the method for its intended purpose and to ensure that the method is 
ready for validation. 
Before or during the development of a bioanalytical method, the applicant is encouraged to, if feasible, 
understand the analyte of interest (e.g., the physicochemical properties of the drug, in vitro and in 
vivo metabolism, preferential distribution between red blood cells and plasma, and protein binding) 
and consider aspects of any prior analytical methods that may be applicable. 
Method development involves identifying the procedures and conditions involved with quantifying the 
analyte. Method development can include the characterisation of the following bioanalytical elements: 
reference standards, critical reagents, calibration curve, quality control samples (QCs), selectivity and 
specificity, sensitivity, accuracy, precision, recovery, stability of the analyte, minimum required dilution 
(MRD). 
Bioanalytical method development does not require extensive record keeping or notation. Once the 
method has been developed, bioanalytical method validation proves that the method is suited to the 
analysis of the study samples.  
If a problem is encountered with the method during the analysis of nonclinical or clinical study samples 
that requires that the analysis be stopped, any changes to the method and the rationale should be 
documented. 
2.2.  Method validation 
2.2.1.  Full validation 
Bioanalytical method validation is essential to ensure the acceptability of assay performance and the 
reliability of analytical results. A bioanalytical method is defined as a set of procedures used for 
measuring analyte concentrations in biological samples. A full validation of a bioanalytical method 
should be performed when establishing a bioanalytical method for the quantification of an analyte in 
clinical and in applicable nonclinical studies. Full validation should also be performed when 
implementing an analytical method that is reported in the literature and when a commercial kit is 
repurposed for bioanalytical use in drug development. Usually, one analyte has to be determined, but 
on occasion it may be appropriate to measure more than one analyte. This may involve two different 
drugs, a parent drug with its metabolites or the enantiomers or isomers of a drug. In these cases, the 
principles of validation and analysis apply to all analytes of interest. For chromatographic methods a 
full validation should include the following elements, unless otherwise justified: selectivity, specificity, 
matrix effect, calibration curve (response function), range (lower limit of quantification (LLOQ) to 
upper limit of quantification (ULOQ)), accuracy, precision, carry-over, dilution integrity, stability and 
reinjection reproducibility.  
For LBAs the following elements should be evaluated, unless otherwise justified: specificity, selectivity, 
calibration curve (response function), range (LLOQ to ULOQ), accuracy, precision, carry-over, dilution 
linearity, and stability. If necessary, parallelism can be conducted when appropriate study samples are 
available.  
The assessments that are performed during validation should be relevant to the sample analysis 
workflow. The matrix used for bioanalytical method validation should be the same as the matrix of the 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 7/45 
 
study samples, including anticoagulants and additives. In some cases, it may be difficult to obtain an 
identical matrix to that of the study samples (e.g., rare matrices such as tissue, cerebrospinal fluid, 
bile or in cases where free drug is measured). In such cases, surrogate matrices may be acceptable for 
analytical method validation.  
The choice of surrogate matrix should be scientifically justified. Matrix differences within species (e.g., 
age, ethnicity, gender) are generally not considered different when validating a method. 
A specific, detailed, written description of the bioanalytical method and validation procedure should be 
established a priori. This description may be in the form of a protocol, study plan, report, notebook or 
Standard Operating Procedure (SOP). 
2.2.2.  Partial validation 
Modifications to a fully validated analytical method may be evaluated by partial validation. Partial 
validation can range from as little as one accuracy and precision determination to a nearly full 
validation (refer to Section 6.1). The items in a partial validation should be determined according to 
the extent and nature of the changes made to the method. 
2.2.3.  Cross validation 
Cross validation is required to demonstrate how the reported data are related when multiple 
bioanalytical methods and/or multiple bioanalytical laboratories are involved. (Refer to Section 6.2) 
3.  Chromatography 
3.1.  Reference standards 
During method validation and the analysis of study samples, a blank biological matrix is spiked with 
the analyte(s) of interest using solutions of reference standard(s) to prepare calibration standards and 
QCs. Calibration standards and QCs should be prepared from separate stock solutions. However, 
calibration standards and QCs may be prepared from the same stock solution provided the accurate 
preparation and stability of the stock solution should have been verified.  
A suitable internal standard (IS) should be added to all calibration standards, QCs and study samples 
during sample processing. The absence of an IS should be justified.  
It is important that the reference standard is well characterised and the quality (e.g., purity, identity) 
of the reference standard and the suitability of the IS is ensured, as the quality will affect the outcome 
of the analysis and, therefore, the study data. The reference standard used during validation and study 
sample analysis should be obtained from an authentic and traceable source. The reference standard 
should be identical to the analyte. If this is not possible, an established form (e.g., salt or hydrate) of 
known quality should be used. 
Suitable reference standards include compendial standards, commercially available standards or 
sufficiently characterised standards prepared in-house or by an external organisation. A certificate of 
analysis (CoA) or an equivalent alternative is required to ensure quality and to provide information on 
the purity, storage conditions, retest/expiration date and batch number of the reference standard.  
A CoA is not required for the IS as long as the suitability for use is demonstrated, e.g., a lack of 
analytical interference is shown for the substance itself or any impurities thereof.  
When MS detection is used, the use of the stable isotope-labelled analyte as the IS is recommended 
whenever possible. However, it is essential that the labelled standard is of high isotope purity and that 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 8/45 
 
no isotope exchange reaction occurs. The presence of unlabelled analyte should be checked and if 
unlabelled analyte is detected, the potential influence should be evaluated during method validation. 
Stock and working solutions should only be prepared from reference standards that are within the 
stability period as documented in the CoA (either expiration date or the retest date). 
3.2.  Validation 
3.2.1.  Selectivity 
Selectivity is the ability of an analytical method to differentiate and measure the analyte in the 
presence of potential interfering substances in the blank biological matrix.  
Selectivity should be evaluated using blank samples (matrix samples processed without addition of an 
analyte or IS) obtained from at least 6 individual sources/lots (non-haemolysed and non-lipaemic). Use 
of fewer sources may be acceptable in the case of rare matrices. Selectivity for the IS should also be 
evaluated. 
The evaluation of selectivity should demonstrate that no significant response attributable to interfering 
components is observed at the retention time(s) of the analyte or the IS in the blank samples. 
Responses attributable to interfering components should not be more than 20% of the analyte 
response at the LLOQ and not more than 5% of the IS response in the LLOQ sample for each matrix. 
For the investigation of selectivity in lipaemic matrices at least one source of matrix should be used. To 
be scientifically meaningful, the matrix used for these tests should be representative as much as 
possible of the expected study samples. A naturally lipaemic matrix with abnormally high levels of 
triglycerides should be obtained from donors. Although it is recommended to use lipaemic matrix from 
donors, if this is difficult to obtain, matrix can be spiked with triglycerides even though it may not be 
representative of study samples. However, if the drug impacts lipid metabolism or if the intended 
patient population is hyperlipidaemic, the use of spiked samples is discouraged. This evaluation is not 
necessary for nonclinical studies unless the drug impacts lipid metabolism or is administered in a 
particular animal strain that is hyperlipidaemic.  
For the investigation of selectivity in haemolysed matrices at least one source of matrix should be 
used. Haemolysed matrices should be obtained by spiking matrix with haemolysed whole blood (at 
least 2% V/V) to generate a visibly detectable haemolysed sample. 
3.2.2.  Specificity 
Specificity is the ability of a bioanalytical method to detect and differentiate the analyte from other 
substances, including its related substances (e.g., substances that are structurally similar to the 
analyte, metabolites, isomers, impurities, degradation products formed during sample preparation, or 
concomitant medications that are expected to be used in the treatment of patients with the intended 
indication). 
If the presence of related substances is anticipated in the biological matrix of interest, the impact of 
such substances should be evaluated during method validation, or alternatively, in the pre-dose study 
samples. In the case of LC-MS based methods, to assess the impact of such substances, the evaluation 
may include comparing the molecular weight of a potential interfering related substance with the 
analyte and chromatographic separation of the related substance from the analyte.  
Responses detected and attributable to interfering components should not be more than 20% of the 
analyte response at the LLOQ and not more than 5% of the IS response in the LLOQ sample. 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 9/45 
 
The possibility of back-conversion of a metabolite into the parent analyte during the successive steps 
of the analysis (including extraction procedures or in the MS source) should also be evaluated when 
relevant (e.g., potentially unstable metabolites such as ester analytes to ester/acidic metabolites, 
unstable N-oxides or glucuronide metabolites, lactone-ring structures). It is acknowledged that this 
evaluation will not be possible in the early stages of drug development of a new chemical entity when 
the metabolism is not yet evaluated. However, it is expected that this issue should be investigated, 
and partial validation performed if needed. The extent of back-conversion, if any, should be established 
and the impact on the study results should be discussed in the Bioanalytical Report. 
3.2.3.  Matrix effect 
A matrix effect is defined as an alteration of the analyte response due to interfering and often 
unidentified component(s) in the sample matrix. During method validation the matrix effect between 
different independent sources/lots should be evaluated. 
The matrix effect should be evaluated by analysing at least 3 replicates of low and high QCs, each 
prepared using matrix from at least 6 different sources/lots. For each individual matrix sources/lots 
evaluated, the accuracy should be within ±15% of the nominal concentration and the precision (per 
cent coefficient of variation (%CV)) should not be greater than 15%. Use of fewer sources/lots may be 
acceptable in the case of rare matrices.  
The matrix effect should also be evaluated in relevant patient populations or special populations (e.g., 
hepatically impaired or renally impaired) when available. An additional evaluation of the matrix effect 
is recommended using haemolysed or lipaemic matrix samples during method validation on a case-by-
case basis, especially when these conditions are expected to occur within the study. 
3.2.4.  Calibration curve and range 
The calibration curve demonstrates the relationship between the nominal analyte concentration and 
the response of the analytical platform to the analyte. Calibration standards, prepared by spiking 
matrix with a known quantity of analyte(s), span the calibration range and comprise the calibration 
curve. Calibration standards should be prepared in the same biological matrix as the study samples. 
The calibration range is defined by the LLOQ, which is the lowest calibration standard, and the ULOQ, 
which is the highest calibration standard. There should be one calibration curve for each analyte 
studied during method validation and for each analytical run. 
A calibration curve should be generated with a blank sample, a zero sample (blank sample spiked with 
IS), and at least 6 concentration levels of calibration standards, including the LLOQ and the ULOQ.  
A simple regression model that adequately describes the concentration-response relationship should be 
used. The selection of the regression model should be directed by written procedures. The regression 
model, weighting scheme and transformation should be determined during the method validation. 
Blank and zero samples should not be included in the determination of the regression equation for the 
calibration curve. Each calibration standard may be analysed in replicate, in which case data from all 
acceptable replicates should be used in the regression analysis. 
The calibration curve parameters should be reported (e.g., slope and intercept in the case of a linear 
model). The back-calculated concentrations of the calibration standards should be presented together 
with the calculated mean accuracy and precision values. All acceptable curves obtained during 
validation, based on a minimum of 3 independent runs over several days, should be reported. The 
accuracy of the back-calculated concentrations of each calibration standard should be within ±20% of 
the nominal concentration at the LLOQ and within ±15% at all the other levels. At least 75% of the 
calibration standards with a minimum of 6 calibration standard levels should meet the above criteria.  
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 10/45 
 
In the case that replicates are used, the criteria (within ±15% or ±20% for LLOQ) should also be 
fulfilled for at least 50% of the calibration standards tested per concentration level. In the case that a 
calibration standard does not comply with these criteria, this calibration standard sample should be 
rejected, and the calibration curve without this calibration standard should be re-evaluated, including 
regression analysis. For accuracy and precision runs, if all replicates of the LLOQ or the ULOQ 
calibration standard in a run are rejected, then the run should be rejected, the possible source of the 
failure should be determined and the method revised, if necessary. If the next validation run also fails, 
then the method should be revised before restarting validation.  
The calibration curve should be prepared using freshly spiked calibration standards in at least one 
assessment. Subsequently, frozen calibration standards can be used within their defined period of 
stability. 
3.2.5.  Accuracy and precision 
3.2.5.1.  Preparation of quality control samples 
The QCs are intended to mimic study samples and should be prepared by spiking matrix with a known 
quantity of analyte, storing them under the conditions anticipated for study samples and analysing 
them to assess the validity of the analytical method. 
Calibration standards and the QCs should be prepared from separate stock solutions in order to avoid 
biased estimations which are not related to the analytical performance of the method. If calibration 
standards and the QCs may be prepared from the same stock solution, the accuracy and stability of 
the stock solution should be verified. A single source of blank matrix may be used, which should be 
free of interference or matrix effects, as described in Section 3.2.3. 
During method validation the QCs for accuracy and precision runs should be prepared at a minimum of 
4 concentration levels within the calibration curve range: the LLOQ, within three times of the LLOQ 
(low QC), around 30 - 50% of the calibration curve range (medium QC) and at least 75% of the ULOQ 
(high QC).  
For non-accuracy and precision validation runs, low, medium and high QCs may be analysed in 
duplicate. These QCs, along with the calibration standards, will provide the basis for the acceptance or 
rejection of the run. 
3.2.5.2.  Evaluation of accuracy and precision 
Accuracy and precision should be determined by analysing the QCs within each run (within-run) and in 
different runs (between-run). Accuracy and precision should be evaluated using the same runs and 
data. 
Within-run accuracy and precision should be evaluated by analysing at least 5 replicates at each QC 
concentration level in each analytical run. Between-run accuracy and precision should be evaluated by 
analysing each QC concentration level in at least 3 analytical runs over at least two days. To enable the 
evaluation of any trends over time within one run, it is recommended to demonstrate accuracy and 
precision of the QCs over at least one of the runs in a size equivalent to a prospective analytical run of 
study samples. Reported method validation data and the determination of accuracy and precision 
should include all results obtained, including individual QCs outside of the acceptance criteria, except 
those cases where errors are obvious and documented. Within-run accuracy and precision data should 
be reported for each run. If the within-run accuracy or precision criteria are not met in all runs, an 
overall estimate of within-run accuracy and precision for each QC level should be calculated. Between-
run (intermediate) precision and accuracy should be calculated by combining the data from all runs.  
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 11/45 
 
The calibration curves for these assessments should be prepared using freshly spiked calibration 
standards in at least one run. If freshly spiked calibration standards are not used in the other runs, 
stability of the frozen calibration standards should be demonstrated. 
The accuracy at each concentration level should be within ±15% of the nominal concentration, except 
at the LLOQ, where it should be within ±20%. The precision (%CV) of the concentrations determined 
at each level should not exceed 15%, except at the LLOQ, where it should not exceed 20%. For non-
accuracy and precision validation runs, at least 2/3 of the total QCs and at least 50% at each 
concentration level should be within ±15% of the nominal values. 
3.2.6.  Carry-over 
Carry-over is an alteration of a measured concentration due to residual analyte from a preceding 
sample that remains in the analytical instrument.  
Carry-over should be assessed and minimised during method development. During validation carry-
over should be assessed by analysing blank samples after the calibration standard at the ULOQ. Carry-
over in the blank samples following the highest calibration standard should not be greater than 20% of 
the analyte response at the LLOQ and 5% of the response for the IS. If it appears that carry-over is 
unavoidable, study samples should not be randomised. Specific measures should be considered, 
validated and applied during the analysis of the study samples, so that carry-over does not affect 
accuracy and precision. This could include the injection of blank sample(s) after samples with an 
expected high concentration, before the next study sample. 
3.2.7.  Dilution integrity 
Dilution integrity is the assessment of the sample dilution procedure, when required, to confirm that it 
does not impact the accuracy and precision of the measured concentration of the analyte. The same 
matrix from the same species used for preparation of the QCs should be used for dilution. 
Dilution QCs should be prepared with analyte concentrations in matrix that are greater than the ULOQ 
and then diluted with blank matrix. At least 5 replicates per dilution factor should be tested in one run 
to determine if concentrations are accurately and precisely measured within the calibration range. The 
dilution factor(s) and concentrations applied during study sample analysis should be within the range 
of the dilution factors and concentrations evaluated during validation. The mean accuracy of the 
dilution QCs should be within ±15% of the nominal concentration and the precision (%CV) should not 
exceed 15%. 
In the cases of rare matrices, use of a surrogate matrix for dilution may be acceptable. It should be 
demonstrated that this does not affect precision and accuracy. 
3.2.8.  Stability 
Stability evaluations should be carried out to ensure that every step taken during sample preparation, 
processing and analysis as well as the storage conditions used do not affect the concentration of the 
analyte.  
The storage and analytical conditions applied to the stability tests, such as the sample storage times 
and temperatures, sample matrix, anticoagulant and container materials, should reflect those used for 
the study samples. Reference to data published in the literature is not considered sufficient. Validation 
of storage periods should be performed on QCs that have been stored for a time that is equal to or 
longer than the study sample storage periods. 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 12/45 
 
Stability of the analyte in the matrix is evaluated using low and high concentration QCs. Aliquots of the 
low and high QCs are analysed at time zero and after the applied storage conditions that are to be 
evaluated. One bulk QC should be prepared at each concentration level. For each concentration tested, 
the bulk sample should be divided into a minimum of 3 aliquots that will be stored, stressed and 
analysed.  
The QCs should be analysed against a calibration curve, obtained from freshly spiked calibration 
standards in a run with its corresponding freshly spiked QCs or QCs for which stability has been 
proven. The mean concentration at each QC level should be within ±15% of the nominal concentration. 
If the concentrations of the study samples are consistently higher than the ULOQ of the calibration 
range, the concentration of the high QC should be adjusted to reflect these higher concentrations. It is 
recognised that this may not be possible in nonclinical studies due to solubility limitations. 
For fixed dose combination products and specifically labelled drug regimens, the freeze-thaw, bench-
top and long-term stability tests of an analyte in matrix should be conducted with the matrix spiked 
with all of the dosed compounds.  
The following stability tests should be evaluated: 
1. Stability of the analyte in matrix  
Freeze-thaw stability in matrix  
To assess the impact of repeatedly removing samples from frozen storage, the stability of the analyte 
should be assessed after multiple cycles of freezing and thawing. Low and high QCs should be thawed 
and analysed according to the same procedures as the study samples. QCs should be kept frozen for at 
least 12 hours between the thawing cycles. QCs for freeze-thaw stability should be assessed using 
freshly prepared calibration standards and QCs, or QCs for which stability has been proven. The 
number of freeze-thaw cycles validated should equal or exceed that of the freeze-thaw cycles 
undergone by the study samples, but a minimum of three cycles should be conducted. 
Bench-top (short-term) stability in matrix  
Bench top matrix stability experiments should be designed and conducted to cover the laboratory 
handling conditions for the study samples. 
Low and high QCs should be thawed in the same manner as the study samples and kept on the bench 
top at the same temperature and for at least the same duration as the study samples. 
The total time on the bench top should be concurrent; it is not acceptable to use additive exposure to 
bench top conditions (i.e., time from each freeze-thaw evaluation should not be added up). 
Long-term stability in matrix  
The long-term stability of the analyte in matrix stored in the freezer should be established. Low and 
high QCs should be stored in the freezer under the same storage conditions and at least for the same 
duration as the study samples.  
For chemical drugs, the stability at one temperature (e.g., -20°C) to can be extrapolated to lower 
temperatures (e.g., -70/-80°C). 
For biological drugs, a bracketing approach can be applied, e.g., in the case that the stability has been 
demonstrated at -70/-80°C and at -20°C, then it is not necessary to investigate the stability at 
temperatures in between those two points at which study samples will be stored.  
2. Stability of the analyte in processed samples  
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 13/45 
 
The stability of processed samples, including the time until completion of analysis (in the 
autosampler/instrument), should be determined. For example: 
• 
Stability of the processed sample under the storage conditions to be used during the analysis of 
study samples (dry extract or in the injection phase)  
• 
On-instrument/autosampler stability of the processed sample at injector or autosampler 
temperature.  
The total time that a processed sample is stored must be concurrent (i.e., autosampler and other 
storage times cannot be added together). 
3. Stability of the analyte and IS in stock and working solutions 
The stability of the stock and working solutions of the analyte and IS should be determined under the 
storage conditions used during the analysis of study samples by using the lowest and the highest 
concentrations of these solutions. They should be assessed using the response of the detector. 
Stability of the stock and working solutions should be tested with an appropriate dilution, taking into 
consideration the linearity and measuring range of the detector. If the stability varies with 
concentration, then the stability of all concentrations of the stock and working solutions needs to be 
assessed. If no isotopic exchange occurs for the stable isotopically-labelled IS under the same storage 
conditions as the analyte for which the stability is demonstrated, then no additional stability 
determinations for the IS are necessary. If the reference standard expires, or it is past the retest date, 
the stability of the stock solutions made previously with this lot of reference standard are defined by 
the expiration or retest date established for the stock solution. The practice of making stock and 
working solutions from reference standards solely for extending the expiry date for the use of the 
reference standard is not acceptable.    
In addition, the following test should be performed if applicable: 
4. Stability of the analyte in whole blood  
Sufficient attention should be paid to the stability of the analyte in the sampled matrix (blood) directly 
after collection from subjects and prior to preparation for storage to ensure that the concentrations 
obtained by the analytical method reflect the concentrations of the analyte in the subject’s blood at the 
time of sample collection.  
If the matrix used is plasma, the stability of the analyte in blood should be evaluated during method 
development (e.g., using an exploratory method in blood) or during method validation. The results 
should be provided in the Validation Report. 
3.2.9.  Reinjection reproducibility 
Reproducibility of the method is assessed by replicate measurements of the QCs and is usually 
included in the assessment of precision and accuracy. However, if samples could be reinjected (e.g., in 
the case of instrument interruptions or other reasons such as equipment failure), reinjection 
reproducibility should be evaluated to establish the viability of the processed samples and to support 
their storage prior to reinjection. 
Reinjection reproducibility is assessed by reinjecting a run that is comprised of calibration standards 
and a minimum of 5 replicates of the low, middle and high QCs after storage. The precision and 
accuracy of the reinjected QCs establish the viability of the processed samples. 
The results should be included in the Validation Report or provided in the Bioanalytical Report of the 
study where it was conducted. 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 14/45 
 
3.3.  Study sample analysis 
The analysis of study samples can be carried out after validation has been completed, however, it is 
understood that some parameters may be completed at a later stage (e.g., long-term stability). By the 
time the data are submitted to a regulatory authority, the bioanalytical method validation should have 
been completed. The study samples, QCs and calibration standards should be processed in accordance 
with the validated analytical method. If system suitability is assessed, a predefined specific study plan, 
protocol or SOP should be used. System suitability, including apparatus conditioning and instrument 
performance, should be determined using samples that are independent of the calibration standards 
and QCs for the run. Subject samples should not be used for system suitability. The IS responses of 
the study samples should be monitored to determine whether there is systemic IS variability. Refer to 
Table 1 for expectations regarding documentation. 
3.3.1.  Analytical run 
An analytical run consists of a blank sample (processed matrix sample without analyte and without IS), 
a zero sample (processed matrix with IS), calibration standards at a minimum of 6 concentration 
levels, at least 3 levels of QCs (low, medium and high) in duplicate (or at least 5% of the number of 
study samples, whichever is higher) and the study samples to be analysed. The QCs should be 
interspersed in the run in such a way that the accuracy and precision of the whole run is ensured. 
Study samples should always be bracketed by QCs.  
The calibration standards and QCs should be spiked independently using separately prepared stock 
solutions, unless the accuracy and stability of the stock solutions have been verified. All samples 
(calibration standards, QCs and study samples) should be processed and extracted as one single batch 
of samples in the order in which they are intended to be analysed. Analysing samples that were 
processed as several separate batches in a single analytical run is discouraged. If such an approach 
cannot be avoided, for instance due to bench top stability limitations, each batch of samples should 
include low, medium and high QCs. 
For comparative BA/BE studies, it is advisable to analyse all samples of one subject together in one 
analytical run to reduce variability.  
The impact of any carry-over that occurs during study sample analysis should be assessed and 
reported (refer to Section 3.2.6). If carry-over is detected, its impact on the measured concentrations 
should be mitigated (e.g., non-randomisation of study samples, injection of blank samples after 
samples with an expected high concentration) or the validity of the reported concentrations should be 
justified in the Bioanalytical Report. 
3.3.2.  Acceptance criteria for an analytical run 
Criteria for the acceptance or rejection of an analytical run should be defined in the protocol, in the 
study plan or in an SOP. In the case that a run contains multiple batches, acceptance criteria should be 
applied to the whole run and to the individual batches. It is possible for the run to meet acceptance 
criteria, even if a batch within that run is rejected for failing to meet the batch acceptance criteria. 
Calibration standards in a failed batch cannot be used to support the acceptance of other batches 
within the analytical run. 
The back-calculated concentrations of the calibration standards should be within ±15% of the nominal 
value, except for the LLOQ for which it should be within ±20%. At least 75% of the calibration 
standard concentrations, which should include a minimum of six concentration levels, should fulfil 
these criteria. If more than six calibration standard levels are used and one of the calibration standards 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 15/45 
 
does not meet the criteria, this calibration standard should be rejected and the calibration curve 
without this calibration standard should be re-evaluated and a new regression analysis performed. 
If the rejected calibration standard is the LLOQ, the new lower limit for this analytical run is the next 
lowest acceptable calibration standard of the calibration curve. This new lower limit calibration 
standard will retain its original acceptance criteria (i.e., ±15%). If the highest calibration standard is 
rejected, the ULOQ for this analytical run is the next acceptable highest calibration standard of the 
calibration curve. The revised calibration range should cover at least 3 QC concentration levels (low, 
medium and high). Study samples outside of the revised range should be reanalysed. If replicate 
calibration standards are used and only one of the LLOQ or ULOQ standards fails, the calibration range 
is unchanged.  
At least 2/3 of the total QCs and at least 50% at each concentration level should be within ±15% of 
the nominal values. If these criteria are not fulfilled the analytical run should be rejected. A new 
analytical batch should be prepared for all study samples within the failed analytical run for subsequent 
analysis. In the cases where the failure is due to an assignable technical cause, samples may be 
reinjected. 
Analytical runs containing samples that are diluted and reanalysed should include dilution QCs to verify 
the accuracy and precision of the dilution method during study sample analysis. The concentration of 
the dilution QCs should exceed that of the study samples being diluted (or of the ULOQ) and they 
should be diluted using the same dilution factor. If multiple dilution factors are used in one analytical 
run, then dilution QCs need only be diluted by the highest and lowest dilution factors. The within-run 
acceptance criteria of the dilution QC(s) will only affect the acceptance of the diluted study samples 
and not the outcome of the analytical run.  
When several analytes are assayed simultaneously, there should be one calibration curve for each 
analyte studied. If an analytical run is acceptable for one analyte but has to be rejected for another 
analyte, the data for the accepted analyte should be used. The determination of the rejected analyte 
requires re-extraction and analysis only for the analyte that is reanalysed. Only data for this 
reanalysed analyte needs to be reported. 
The back-calculated concentrations of the calibration standards and QCs of passed and accepted runs 
should be reported. The overall (between-run) accuracy and precision of the QCs of all accepted runs 
should be calculated at each concentration level and reported in the analytical report (refer to Section 
8 Documentation and Table 1). If the overall mean accuracy and/or precision fails the 15% criterion, 
an investigation to determine the cause of the deviation should be conducted. In the case of 
comparative BA/BE studies, it may result in the rejection of the data. 
3.3.3.  Calibration range 
If a narrow range of analyte concentrations of the study samples is known or anticipated before the 
start of study sample analysis, it is recommended to either narrow the calibration curve range, adapt 
the concentrations of the QCs, or add new QCs at different concentration levels as appropriate, to 
adequately reflect the concentrations of the study samples. 
At the intended therapeutic dose(s), if an unanticipated clustering of study samples at one end of the 
calibration curve is encountered after the start of sample analysis, the analysis should be stopped and 
either the standard calibration range narrowed (i.e., partial validation), existing QC concentrations 
revised, or QCs at additional concentrations added to the original curve within the observed range 
before continuing with study sample analysis. It is not necessary to reanalyse samples analysed before 
optimising the calibration curve range or QC concentrations. 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 16/45 
 
The same applies if a large number of the analyte concentrations of the study samples are above the 
ULOQ. The calibration curve range should be changed, if possible, and QC(s) added or their 
concentrations modified. If it is not possible to change the calibration curve range or the number of 
samples with a concentration above the ULOQ is not large, samples should be diluted according to the 
validated dilution method. 
At least 2 QC levels should fall within the range of concentrations measured in study samples. If the 
calibration curve range is changed, the bioanalytical method should be revalidated (partial validation) 
to verify the response function and to ensure accuracy and precision. 
3.3.4.  Reanalysis of study samples 
Possible reasons for reanalysis of study samples, the number of replicates and the decision criteria to 
select the value to be reported should be predefined in the protocol, study plan or SOP, before the 
actual start of the analysis of the study samples. For study samples in which multiple analytes are 
being analysed, a valid result for one analyte should not be rejected if the other analyte fails the 
acceptance criteria. 
The number of samples (and percentage of total number of samples) that have been reanalysed should 
be reported and discussed in the Bioanalytical Report. For comparative BA/BE studies, a separate table 
should report values from rejected runs. 
Some examples of reasons for study sample reanalysis are: 
• 
Rejection of an analytical run because the run failed the acceptance criteria with regard to accuracy 
of the calibration standards and/or the precision and accuracy of the QCs 
• 
IS response significantly different from the response for the calibration standards and QCs (as pre-
defined in an SOP) 
• 
The concentration obtained is above the ULOQ 
• 
The concentration observed is below the revised LLOQ in runs where the lowest calibration 
standard has been rejected from a calibration curve, resulting in a higher LLOQ compared with 
other runs 
• 
Improper sample injection or malfunction of equipment 
• 
The diluted study sample is below the LLOQ 
• 
Identification of quantifiable analyte levels in pre-dose samples, control or placebo samples 
• 
Poor chromatography (as pre-defined in an SOP) 
For comparative BA/BE studies, reanalysis of study samples for a PK reason (e.g., a sample 
concentration does not fit with the expected profile) is not acceptable, as it may bias the study result. 
Any reanalysed samples should be identified in the Bioanalytical Report and the initial value, the 
reason for reanalysis, the values obtained in the reanalyses, the final accepted value and a justification 
for the acceptance should be provided. Further, a summary table of the total number of samples that 
have been reanalysed for each reason should be provided. In cases where the first analysis yields a 
non-reportable result, a single reanalysis is considered sufficient (e.g., concentration above the ULOQ 
or equipment malfunction). In cases where the value needs to be confirmed (e.g., pre-dose sample 
with measurable concentrations) replicate determinations are required if sample volume allows.  
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 17/45 
 
The safety of trial subjects should take precedence over any other aspect of the trial. Consequently, 
there may be other circumstances when it is necessary to reanalyse specific study samples for the 
purpose of a safety investigation.  
 
3.3.5.  Reinjection of study samples 
Reinjection of processed samples can be made in the case of equipment failure if reinjection 
reproducibility has been demonstrated during validation or provided in the Bioanalytical Report where 
it was conducted. Reinjection of a full analytical run or of individual calibration standards or QCs simply 
because the calibration standards or QCs failed, without any identified analytical cause, is not 
acceptable. 
3.3.6.  Integration of chromatograms 
Chromatogram integration and reintegration should be described in a study plan, protocol or SOP. Any 
deviation from the procedures described a priori should be discussed in the Bioanalytical Report. The 
list of chromatograms that required reintegration, including any manual integrations, and the reasons 
for reintegration should be included in the Bioanalytical Report. Original and reintegrated 
chromatograms and initial and repeat integration results should be kept for future reference and 
submitted in the Bioanalytical Report for comparative BA/BE studies. 
4.  Ligand binding assays 
4.1.  Key reagents  
4.1.1.  Reference standard 
The reference standard should be well characterised and documented (e.g., CoA and origin). A 
biological drug has a highly complex structure and its reactivity with binding reagents for bioanalysis 
may be influenced by a change in the manufacturing process of the drug substance. It is recommended 
that the manufacturing batch of the reference standard used for the preparation of calibration 
standards and QCs is derived from the same batch of drug substance as that used for dosing in the 
nonclinical and clinical studies whenever possible. If the reference standard batch used for bioanalysis 
is changed, bioanalytical evaluation should be carried out with QCs from the original material and the 
new material prior to use to ensure that the performance characteristics of the method are within the 
acceptance criteria. 
4.1.2.  Critical reagents 
Critical reagents, including binding reagents (e.g., binding proteins, aptamers, antibodies or 
conjugated antibodies) and those containing enzymatic moieties, have direct impact on the results of 
the assay and, therefore, their quality should be assured. Critical reagents bind the analyte and, upon 
interaction, lead to an instrument signal corresponding to the analyte concentration. The critical 
reagents should be identified and defined in the assay method.  
Reliable procurement of critical reagents, whether manufactured in-house or purchased commercially, 
should be considered early in method development. The data sheet for the critical reagent should 
include at a minimum identity, source, batch/lot number, purity (if applicable), concentration (if 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 18/45 
 
applicable) and stability/retest date/storage conditions (refer to Table 1). Additional characteristics 
may be warranted.  
A critical reagent lifecycle management procedure is necessary to ensure consistency between the 
original and new batches of critical reagents. Reagent performance should be evaluated using the 
bioanalytical method. Minor changes to critical reagents would not be expected to influence the method 
performance, whereas major changes may significantly impact the performance. If the change is minor 
(e.g., the source of one reagent is changed), a single comparative accuracy and precision assessment 
is sufficient for characterisation. If the change is major, then additional validation experiments are 
necessary. Ideally, assessment of changes will compare the method with the new reagents to the 
method with the old reagents directly. Major changes include, but are not limited to, change in 
production method of antibodies, additional blood collection from animals for polyclonal antibodies and 
new clones or new supplier for monoclonal antibody production.  
Retest dates and validation parameters should be documented in order to support the extension or 
replacement of the critical reagent. Stability testing of the reagents should be based upon the 
performance in the bioanalytical method and upon general guidance for reagent storage conditions. It 
can be extended beyond the expiry date from the supplier. The performance parameters should be 
documented in order to support the extension or replacement of the critical reagent. 
4.2.  Validation 
Most often microtitre plates are used for LBAs and study samples can be analysed using an assay 
format of 1 or more well(s) per sample. The assay format should be specified in the protocol, study 
plan or SOP. If method development and method validation are performed using 1 or more well(s) per 
sample, then study sample analysis should also be performed using 1 or more well(s) per sample, 
respectively. If multiple wells per sample are used, the reportable sample concentration should be 
determined either by calculating the mean of the responses from the replicate wells or by averaging 
the concentrations calculated from each response. Data evaluation should be performed on reportable 
concentrations. 
4.2.1.  Specificity 
Specificity is related to the concept of cross-reactivity in LBA. It is important that the binding reagent 
specifically binds to the target analyte but does not cross-react with coexisting structurally related 
molecules (e.g., endogenous compounds, isoforms or structurally related concomitant medication). 
Specificity is evaluated by spiking blank matrix samples with related molecules at the maximal 
concentration(s) of the structurally related molecule anticipated in study samples.  
The accuracy of the target analyte at the LLOQ and at the ULOQ should be investigated in the presence 
of related molecules at the maximal concentration(s) anticipated in study samples. The response of 
blank samples spiked with related molecules should be below the LLOQ. The accuracy of the target 
analyte in presence of related molecules should be within ±25% of the nominal values.  
In the event of non-specificity, the impact on the method should be evaluated by spiking increasing 
concentrations of interfering molecules in blank matrix and measuring the accuracy of the target 
analyte at the LLOQ and ULOQ. It is essential to determine the minimum concentration of the related 
molecule where interference occurs. Appropriate mitigation during sample analysis should be 
employed, e.g., it may be necessary to adjust the LLOQ/ULOQ accordingly or consider a new method. 
During method development and early method validation, these “related molecules” are frequently not 
available. Additional evaluation of specificity may be conducted after the original validation is 
completed. 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 19/45 
 
4.2.2.  Selectivity 
Selectivity is the ability of the method to detect and differentiate the analyte of interest in the presence 
of non-specific matrix components. The matrix can contain non-specific matrix component such as 
degrading enzymes, heterophilic antibodies or rheumatoid factor which may interfere with the analyte 
of interest. 
Selectivity should be evaluated at the low end of an assay where problems occur in most cases, but it 
is recommended that selectivity is also evaluated at higher analyte concentrations. Therefore, 
selectivity is evaluated using blank samples obtained from at least 10 individual sources and by spiking 
the individual blank matrices at the LLOQ and at the high QC level. Use of fewer sources may be 
acceptable in the case of rare matrices. The response of the blank samples should be below the LLOQ 
in at least 80% of the individual sources.  
The accuracy should be within ±25% at the LLOQ and within ±20% at the high QC level of the nominal 
concentration in at least 80% of the individual sources evaluated.  
Selectivity should be evaluated in lipaemic samples and haemolysed samples (refer to Section 3.2.1). 
For lipaemic and haemolysed samples, tests can be evaluated once using a single source of matrix. 
Selectivity should be assessed in samples from relevant patient populations (e.g., renally or hepatically 
impaired patients, inflammatory or immuno-oncology patients if applicable). In the case of relevant 
patient populations, there should be at least five individual patients. 
4.2.3.  Calibration curve and range 
The calibration curve demonstrates the relationship between the nominal analyte concentration and 
the response of the analytical platform to the analyte. Calibration standards, prepared by spiking 
matrix with a known quantity of analyte, span the calibration range and comprise the calibration curve. 
Calibration standards should be prepared in the same biological matrix as the study samples. The 
calibration range is defined by the LLOQ, which is the lowest calibration standard, and the ULOQ, which 
is the highest calibration standard. There should be one calibration curve for each analyte studied 
during method validation and for each analytical run. If needed, the use of surrogate matrix should be 
scientifically justified. 
A calibration curve should be generated with at least 6 concentration levels of calibration standards, 
including LLOQ and ULOQ standards, plus a blank sample. The blank sample should not be included in 
the calculation of calibration curve parameters. Anchor point samples at concentrations below the 
LLOQ and above the ULOQ of the calibration curve may also be used to improve curve fitting. The 
relationship between response and concentration for a calibration curve is most often fitted by a 4- or 
5-parameter logistic model if there are data points near the lower and upper asymptotes. Oher models 
should be suitably justified. 
A minimum of 6 independent runs should be evaluated over several days considering the factors that 
may contribute to between-run variability. 
The accuracy and precision of back-calculated concentrations of each calibration standard should be 
within ±25% of the nominal concentration at the LLOQ and ULOQ, and within ±20% at all other levels. 
At least 75% of the calibration standards excluding anchor points, and a minimum of 6 concentration 
levels of calibration standards, including the LLOQ and ULOQ, should meet the above criteria. The 
anchor points do not require acceptance criteria since they are beyond the quantifiable range of the 
curve. 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 20/45 
 
The calibration curve should preferably be prepared using freshly spiked calibration standards. If 
freshly spiked calibration standards are not used, the frozen calibration standards can be used within 
their defined period of stability. 
 
4.2.4.  Accuracy and precision 
4.2.4.1.  Preparation of quality control samples 
The QCs are intended to mimic study samples and should be prepared by spiking matrix with a known 
quantity of analyte, stored under the conditions anticipated for study samples and analysed to assess 
the validity of the analytical method. 
The dilution series for the preparation of the QCs should be completely independent from the dilution 
series for the preparation of calibration standard samples. They may be prepared from the same stock 
solution (or working stock) provided the accurate preparation and stability have been verified. The QCs 
should be prepared at a minimum of 5 concentration levels within the calibration curve range: The 
analyte should be spiked at the LLOQ, within three times of the LLOQ (low QC), around the geometric 
mean of the calibration curve range (medium QC), and at least at 75% of the ULOQ (high QC) and at 
the ULOQ. 
For non-accuracy and precision validation runs, low, medium and high QCs may be analysed in 
duplicate. These QCs, along with the calibration standards, will provide the basis for the acceptance or 
rejection of the run. 
4.2.4.2.  Evaluation of accuracy and precision 
Accuracy and precision should be determined by analysing the QCs within each run (within-run) and in 
different runs (between-run). Accuracy and precision should be evaluated using the same runs and 
data. 
Accuracy and precision should be determined by analysing at least 3 replicates per run at each QC 
concentration level (LLOQ, low, medium, high, ULOQ) in at least 6 runs over 2 or more days. Reported 
method validation data and the determination of accuracy and precision should include all results 
obtained, except those cases where errors are obvious and documented. Within-run accuracy and 
precision data should be reported for each run. If the within-run accuracy or precision criteria are not 
met in all runs, an overall estimate of within-run accuracy and precision for each QC level should be 
calculated. Between-run (intermediate) precision and accuracy should be calculated by combining the 
data from all runs. 
 
The overall within-run and between-run accuracy at each concentration level should be within ±20% of 
the nominal values, except for the LLOQ and ULOQ, which should be within ±25% of the nominal 
value. Within-run and between-run precision of the QC concentrations determined at each level should 
not exceed 20%, except at the LLOQ and ULOQ, where it should not exceed 25%.  
For non-accuracy and precision validation runs, at least 2/3 of the total QCs and at least 50% at each 
concentration level should be within ±20% of the nominal values. 
Furthermore, the total error (i.e., sum of absolute values of the errors in accuracy (%) and precision 
(%)) should be evaluated. The total error should not exceed 30% (40% at LLOQ and ULOQ). 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 21/45 
 
4.2.5.  Carry-over 
Carry-over is generally not an issue for LBA analyses. However, if the analytical platform is prone to 
carry-over, the potential of carry-over should be investigated by placing blank samples after the 
calibration standard at the ULOQ. The response of blank samples should be below the LLOQ. 
4.2.6.  Dilution linearity and hook effect 
Due to the narrow assay range in many LBAs, study samples may require dilution in order to achieve 
analyte concentrations within the range of the assay. Dilution linearity should be assessed to confirm: 
(i) that measured concentrations are not affected by dilution within the calibration range and (ii) that 
sample concentrations above the ULOQ of a calibration curve are not impacted by hook effect (i.e., a 
signal suppression caused by high concentrations of the analyte), whereby yielding an erroneous 
result. 
The same matrix as that of the study sample should be used for preparation of the QCs for dilution. 
Dilution linearity should be demonstrated by generating a dilution QC, i.e., spiking the matrix with an 
analyte concentration above the ULOQ, analysed undiluted (for hook effect) and diluting this sample 
(to at least 3 different dilution factors) with blank matrix to a concentration within the calibration 
range. For each dilution factor tested, at least 3 independently prepared dilution series should be 
performed using the number of replicates that will be used in sample analysis. The absence or 
presence of response reduction (hook effect) is checked in the dilution QCs and, if observed and unable 
to be eliminated with reasonable measures, steps should be taken to mitigate this effect during the 
analysis of study samples. 
The calculated mean concentration for each dilution should be within ±20% of the nominal 
concentration after correction for dilution and the precision should not exceed 20%.  
The dilution factor(s) applied during study sample analysis should be within the range of dilution 
factors evaluated during validation. 
4.2.7.  Stability 
Stability evaluations should be carried out to ensure that every step taken during sample preparation, 
processing and analysis as well as the storage conditions used do not affect the concentration of the 
analyte.  
The storage and analytical conditions applied to the stability tests, such as the sample storage times 
and temperatures, sample matrix, anticoagulant, and container materials should reflect those used for 
the study samples. Reference to data published in the literature is not considered sufficient. Validation 
of storage periods should be performed on QCs that have been stored for a time that is equal to or 
longer than the study sample storage periods. 
Stability of the analyte in the studied matrix should be evaluated using low and high concentration 
QCs. Aliquots of the low and high QCs are analysed at time zero and after the applied storage 
conditions that are to be evaluated. One bulk QC should be prepared at each concentration level. For 
each concentration tested, the bulk sample should be divided into a minimum of three aliquots that will 
be stored, stressed and analysed. 
The QCs are analysed against a calibration curve, obtained from freshly spiked calibration standards in 
a run with its corresponding freshly spiked QCs or QCs for which stability has been proven. While the 
use of freshly prepared calibration standards and QCs is the preferred approach, it is recognised that in 
some cases, for macromolecules, it may be necessary to freeze them overnight. In such cases, valid 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 22/45 
 
justification should be provided and freeze-thaw stability demonstrated. QCs should be kept frozen for 
at least 12 hours between the thawing cycles. The mean concentration at each QC level should be 
within ±20% of the nominal concentration.  
Since sample dilution may be required for many LBA methods due to a narrow calibration range, the 
concentrations of the study samples may be consistently higher than the ULOQ of the calibration 
curve. If this is the case, the concentration of the QCs should be adjusted, considering the applied 
sample dilution, to represent the actual sample concentration range. 
For fixed dose combination products and specifically labelled drug regimens, the freeze-thaw, bench-
top and long-term stability tests of an analyte in matrix should be conducted with the matrix spiked 
with all of the dosed compounds, on a case-by-case basis.  
As mentioned in Section 3.2.8, the investigation of stability should cover bench top (short-term) 
stability at room temperature or sample preparation temperature and freeze-thaw stability. In 
addition, long-term stability should be studied.  
For chemical drugs, the stability at one temperature (e.g., -20°C) can be extrapolated to lower 
temperatures (e.g., -70/-80°C). 
For biological drugs, a bracketing approach can be applied, e.g., in the case that the stability has been 
demonstrated at -70/-80°C and at -20°C, then it is not necessary to investigate the stability at 
temperatures in between those two points at which study samples will be stored. 
4.3.  Study sample analysis 
The analysis of study samples can be carried out after validation has been completed, however, it is 
understood that some parameters may be completed at a later stage (e.g., long-term stability). By the 
time the data are submitted to a regulatory authority, the bioanalytical method validation should have 
been completed. The study samples, QCs and calibration standards should be processed in accordance 
with the validated analytical method. Refer to Table 1 for expectations regarding documentation. 
4.3.1.  Analytical run 
An analytical run consists of a blank sample, calibration standards at a minimum of 6 concentration 
levels, at least 3 levels of QCs (low, medium and high) applied as two sets (or at least 5% of the 
number of study samples, whichever is higher) and the study samples to be analysed. The blank 
sample should not be included in the calculation of calibration curve parameters. The QCs should be 
placed in the run in such a way that the accuracy and precision of the whole run is ensured taking into 
account that study samples should always be bracketed by QCs.  
Most often microtitre plates are used for LBAs. An analytical run may comprise of one or more plate(s). 
Typically, each plate contains an individual set of calibration standards and QCs. If each plate contains 
its own calibration standards and QCs then each plate should be assessed on its own. However, for 
some platforms the sample capacity may be limited. In this case, sets of calibration standards may be 
placed on the first and the last plate, but QCs should be placed on every single plate. QCs should be 
placed at least at the beginning (before) and at the end (after) of the study samples of each plate. The 
QCs on each plate and each calibration curve should fulfil the acceptance criteria for an analytical run 
(refer to Section 4.3.2). For the calculation of concentrations, the calibration standards should be 
combined to conduct one regression analysis. If the combined calibration curve does not pass the 
acceptance criteria the whole run fails. 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 23/45 
 
4.3.2.  Acceptance criteria for an analytical run 
Criteria for the acceptance or rejection of an analytical run should be defined in the protocol, in the 
study plan or in an SOP. In the case that a run contains multiple batches, acceptance criteria should be 
applied to the whole run and to the individual batches. It is possible for the run to meet acceptance 
criteria, even if a batch within that run is rejected for failing to meet the batch acceptance criteria. 
Calibration standards in a failed batch cannot be used to support the acceptance of other batches 
within the analytical run. 
The back-calculated concentrations of the calibration standards should be within ±20% of the nominal 
value at each concentration level, except for the LLOQ and the ULOQ, for which it should be within 
±25%. At least 75% of the calibration standards, with a minimum of 6 concentration levels, should 
fulfil this criterion. This requirement does not apply to anchor calibration standards. If more than 6 
calibration standards are used and one of the calibration standards does not meet these criteria, this 
calibration standard should be rejected and the calibration curve without this calibration standard 
should be re-evaluated and a new regression analysis performed. 
If the rejected calibration standard is the LLOQ, the new lower limit for this analytical run is the next 
lowest acceptable calibration standard of the calibration curve. If the highest calibration standard is 
rejected, the new upper limit for this analytical run is the next acceptable highest calibration standard 
of the calibration curve. The new lower and upper limit calibration standard will retain their original 
acceptance criteria (i.e., 20%). The revised calibration range should cover all QCs (low, medium and 
high). The study samples outside of the revised assay range should be reanalysed. 
Each run should contain at least 3 levels of QCs (low, medium and high). During study sample 
analysis, the calibration standards and QCs should mimic the analysis of the study sample with regard 
to the number of wells used per study sample. At least 2/3 of the QCs and 50% at each concentration 
level should be within 20% of the nominal value at each concentration level. Exceptions to these 
criteria should be justified and predefined in the SOP or protocol. 
The overall mean accuracy and precision of the QCs of all accepted runs should be calculated at each 
concentration level and reported in the analytical report. In the case that the overall mean accuracy 
and/or precision exceeds 20%, additional investigations should be conducted to determine the 
cause(s) of this deviation. In the case of comparative BA/BE studies, it may result in the rejection of 
the data. 
4.3.3.  Calibration range 
At least 2 QC sample levels should fall within the range of concentrations measured in study samples. 
At the intended therapeutic dose(s), if an unanticipated clustering of study samples at one end of the 
calibration curve is encountered after the start of sample analysis, the analysis should be stopped and 
either the standard calibration range narrowed (i.e., partial validation), existing QC concentrations 
revised, or QCs at additional concentrations added to the original curve within the observed range 
before continuing with study sample analysis. It is not necessary to reanalyse samples analysed before 
optimising the calibration curve range or QC concentrations. 
4.3.4.  Reanalysis of study samples 
Possible reasons for reanalysis of study samples, the number of reanalyses and the decision criteria to 
select the value to be reported should be predefined in the protocol, study plan or SOP, before the 
actual start of the analysis of the study samples. 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 24/45 
 
The number of samples (and percentage of total number of samples) that have been reanalysed should 
be reported and discussed in the Bioanalytical Report. For comparative BA/BE studies, a separate table 
should report values from rejected runs. 
Some examples of reasons for study sample reanalysis are: 
• 
Rejection of an analytical run because the run failed the acceptance criteria with regard to accuracy 
of the calibration standards and/or the precision and accuracy of the QCs, 
• 
The concentration obtained is above the ULOQ 
• 
The concentration obtained is below the LLOQ in runs where the lowest calibration standard has 
been rejected from a calibration curve, resulting in a higher LLOQ compared with other runs 
• 
Malfunction of equipment 
• 
The diluted sample is below the LLOQ 
• 
Identification of quantifiable analyte levels in pre-dose samples, control or placebo samples. 
• 
When samples are analysed in more than one well and non-reportable values are obtained due to 
one replicate failing the pre-defined acceptance criteria (e.g., excessive variability between wells, 
one replicate being above the ULOQ or below the LLOQ). 
For comparative BA/BE studies, reanalysis of study samples for a PK reason (e.g., a sample 
concentration does not fit with the expected profile) is not acceptable, as it may bias the study result. 
The reanalysed samples should be identified in the Bioanalytical Report and the initial value, the 
reason for reanalysis, the values obtained in the reanalyses, the final accepted value and a justification 
for the acceptance should be provided. Further, a summary table of the total number of samples that 
have been reanalysed due to each reason should be provided. In cases where the first analysis yields a 
non-reportable result, a single reanalysis is considered sufficient (e.g., concentration above the ULOQ 
or excessive variability between wells). The analysis of the samples should be based on the same 
number of wells per study sample as in the initial analysis. In cases where the value needs to be 
confirmed, (e.g., pre-dose sample with measurable concentrations) multiple determinations are 
required where sample volume allows.  
The safety of trial subjects should take precedence over any other aspect of the trial. Consequently, 
there may be other circumstances when it is necessary to reanalyse specific study samples for the 
purpose of an investigation. 
5.  Incurred sample reanalysis (ISR) 
The performance of study samples may differ from that of the calibration standards and QCs used 
during method validation, which are prepared by spiking blank matrix. Differences in protein binding, 
back-conversion of known and unknown metabolites, sample inhomogeneity, concomitant medications 
or biological components unique to the study samples may affect measured concentrations of the 
analyte in study samples. ISR is intended to verify the reliability of the reported sample analyte 
concentrations.  
ISR should be performed at least in the following situations: 
• 
For nonclinical studies within the scope of this guideline, ISR should, in general, be performed at 
least once per species.  
• 
All pivotal comparative BA/BE studies 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 25/45 
 
• 
First clinical trial in subjects 
• 
Pivotal early patient trial(s), once per patient population 
• 
First or pivotal trial in patients with impaired hepatic and/or renal function 
ISR is conducted by repeating the analysis of a subset of samples from a given study in separate (i.e., 
different to the original) runs on different days using the same bioanalytical method. 
The extent of ISR depends upon the analyte and the study samples and should be based upon an in-
depth understanding of the analytical method and analyte. However, as a minimum, if the total 
number of study samples is less than or equal to 1000, then 10% of the samples should be 
reanalysed; if the total number of samples is greater than 1000, then 10% of the first 1000 samples 
(100) plus 5% of the number of samples that exceed 1000 samples should be assessed. Objective 
criteria for choosing the subset of study samples for ISR should be predefined in the protocol, study 
plan or an SOP. While the subjects/animals should be picked as randomly as possible from the dosed 
study population, adequate coverage of the concentration profile is important. Therefore, it is 
recommended that the samples for ISR be chosen around the maximum concentration (Cmax) and 
some in the elimination phase. Additionally, the samples chosen should be representative of the whole 
study. 
Samples should not be pooled, as pooling may limit anomalous findings. ISR samples and QCs should 
be processed and analysed in the same manner as in the original analysis. ISR should be performed 
within the stability window of the analyte, but not on the same day as the original analysis.  
The percent difference between the initial concentration and the concentration measured during the 
repeat analysis should be calculated in relation to their mean value using the following equation:  
% difference = repeat value − initial value
mean value
  ×  100 
For chromatographic methods, the percent difference should be within ±20% for at least 2/3 of the 
repeats. For LBAs, the percent difference should be within ± 30% for at least 2/3 of the repeats. 
If the overall ISR results fail the acceptance criteria, an investigation should be conducted and the 
causes remediated. There should be an SOP that directs how investigations are triggered and 
conducted. If an investigation does not identify the cause of the failure, the potential impact of an ISR 
failure on study validity should also be provided in the Bioanalytical Report. If ISR meets the 
acceptance criteria yet shows large or systemic differences between results for multiple samples, this 
may indicate analytical issues and it is advisable to investigate this further. 
Examples of trends that are of concern may include: 
• 
All ISR samples from one subject fail 
• 
All ISR samples from one run fail 
All aspects of ISR evaluations should be documented to allow reconstruction of the study and any 
investigations. Individual samples that are quite different from the original value (e.g., > 50%, 
“flyers”) should not trigger reanalysis of the original sample and do not need to be investigated. ISR 
sample data should not replace the original study sample data. 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 26/45 
 
6.  Partial and cross validation 
6.1.  Partial validation 
Partial validations evaluate modifications to already fully validated bioanalytical methods. Partial 
validation can range from as little as one within-run accuracy and precision determination, to a nearly 
full validation. If stability is established at one facility it does not necessarily need to be repeated at 
another facility.  
For chromatographic methods, typical bioanalytical method modifications or changes that fall into this 
category include, but are not limited to, the following situations: 
• 
Analytical site change using same method (i.e., bioanalytical method transfers between 
laboratories) 
• 
A change in analytical method (e.g., change in detection systems, platform) 
• 
A change in sample processing procedures  
• 
A change in sample volume (e.g., the smaller volume of paediatric samples)  
• 
Changes to the calibration concentration range 
• 
A change in anticoagulant (but not changes in the counter-ion) in biological fluids (e.g., heparin to 
ethylenediaminetetraacetic acid (EDTA)) 
• 
Change from one matrix within a species to another (e.g., switching from human plasma to serum 
or cerebrospinal fluid) or changes to the species within the matrix (e.g., switching from rat plasma 
to mouse plasma)  
• 
A change in storage conditions 
For LBAs, typical bioanalytical method modifications or changes that fall into this category include, but 
are not limited to, the following situations: 
• 
Changes in LBA critical reagents (e.g., lot-to-lot changes) 
• 
Changes in MRD 
• 
A change in storage conditions 
• 
Changes to the calibration concentration range 
• 
A change in analytical method (e.g., change in detection systems, platform) 
• 
Analytical site change using same method (i.e., bioanalytical method transfers between 
laboratories) 
• 
A change in sample preparation  
• 
A change in anticoagulant (but not changes in the counter-ion) in biological fluids (e.g., heparin to 
ethylenediaminetetraacetic acid (EDTA)) 
The parameters of the partial validations should meet the full validation criteria. If these criteria are 
not satisfied, additional investigation and validation is warranted. 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 27/45 
 
6.2.  Cross validation 
Cross validation is required to demonstrate how the reported data are related when multiple 
bioanalytical methods and/or multiple bioanalytical laboratories are involved. 
Cross validation is required under the following situations: 
• 
Data are obtained from different fully validated methods within a study. 
• 
Data are obtained within a study from different laboratories with the same bioanalytical method.  
• 
Data are obtained from different fully validated methods across studies that are going to be 
combined or compared to support special dosing regimens, or regulatory decisions regarding 
safety, efficacy and labelling. 
If data are obtained from different fully validated methods, and these data are not to be combined 
across studies, cross validation is not generally required.  
Cross validation should be performed in advance of study samples being analysed, if possible. 
Cross validation should be assessed by measuring the same set of QCs (low, medium and high) at 
least in triplicate and study samples (if available) that span the study sample concentration range (n≥
30) with both methods, or in both laboratories.  
Bias can be assessed by Bland-Altman plots or Deming regression. Other methods appropriate for 
assessing agreement between two methods (e.g., concordance correlation coefficient) may be used 
too. Alternatively, the concentration vs. time curves for study samples could be plotted for samples 
analysed by each method to assess bias.  
The use of multiple bioanalytical methods for the measurement of the same analyte in the conduct of 
one comparative BA/BE study is strongly discouraged. 
7.  Additional considerations 
7.1.  Methods for analytes that are also endogenous molecules 
For analytes that are also endogenous molecules (e.g., replacement therapies), the accuracy of the 
measurement of the analytes poses a challenge when the method cannot distinguish between the 
therapeutic agent and the endogenous molecule. Furthermore, the endogenous levels of the analyte 
may vary because of age, gender, race, diurnal variations, illness or as side effect of drug treatment. 
This section describes some of the approaches that may be used to assess concentrations of analytes 
that are also endogenous molecules. As a reminder, biomarkers are outside of the scope of this 
guideline. 
If available, biological matrix to prepare calibration standards and QCs should be the same as the 
study samples (i.e., authentic biological matrix) and it should be free of matrix effect and interference, 
as described in Sections 3 and 4. The endogenous concentration in the biological matrix chosen should 
be low enough to obtain an adequate signal-to-noise ratio (e.g., <20% of the LLOQ). 
In those cases where matrices without interference are not available, the following approaches can be 
used to calculate the concentration of the analyte in the study samples: 1) the surrogate matrix 
approach, 2) the surrogate analyte approach, 3) the background subtraction approach and 4) the 
standard addition approach. 
1. Surrogate matrix approach: 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 28/45 
 
The matrix for the calibration standards is substituted by a surrogate matrix. Surrogate matrices can 
vary widely in complexity from simple buffers or artificial matrices that try to mimic the authentic one, 
to stripped matrices or matrices from other species.  
2. Surrogate analyte approach: 
Stable-isotope labelled analytes are used as surrogate standards in mass spectrometric methods to 
construct the calibration curve for the quantification of endogenous analytes. In this approach, it is 
assumed that the physicochemical properties of the authentic and surrogate analytes are the same 
with the exception of molecular weight. However, isotope standards may differ in retention time and 
MS sensitivity, therefore, before application of this approach, the ratio of the MS responses (i.e., the 
response factor) of the labelled to unlabelled analyte should be close to unity and remain constant over 
the entire calibration range. If the response factor does not comply with these requirements, it should 
be incorporated into the regression equation of the calibration curve. 
3. Background subtraction approach: 
The concentration of the endogenous analyte observed in a pooled/representative matrix is subtracted 
from the concentration observed in the spiked standards, subsequently the net differences are used to 
construct the calibration curve. 
When the background concentrations are lowered by dilution of the blank matrices before spiking with 
the standards (e.g., if a lower LLOQ is required) the composition of the matrices in the study samples 
and the calibration standards is different, which may cause different recoveries and matrix effects. 
These differences should be considered when validating the method.  
4. Standard addition approach: 
The standard addition approach is only applicable for analytical platforms with linear responses. 
Typically, the standard addition method is used to determine the concentration of the endogenous 
analyte in the authentic matrix to be used for preparation of standards and QCs. However, this 
approach can be employed for determination of study samples as well. In this approach, every study 
sample is divided into aliquots of equal volume. All aliquots, but one, are separately spiked with known 
and varying amounts of the analyte standards to construct a calibration curve for either the authentic 
blank matrix or every study sample (e.g., with 3 to 5 points). The endogenous blank concentration or 
the study sample concentration is then determined as the negative x-intercept of the standard 
calibration curve prepared in that particular study sample. 
Validation of an analytical method for an analyte that is also an endogenous compound will require the 
following considerations, in addition to the validation shown in the Sections 3 and 4. 
7.1.1.  Quality control samples for methods for analytes that are also 
endogenous molecules 
The endogenous concentrations of the analyte in the biological matrix should be evaluated prior to QC 
preparation. The matrices with the lowest possible level of the interfering endogenous analyte should 
be used. The concentrations of the QCs should account for the endogenous concentrations in the 
authentic matrix and be representative of the expected study sample concentrations. 
The QCs should resemble study samples and should be prepared in the same matrix. In principle, all 
QC concentrations used for validation should be aliquots of the authentic biological matrix unspiked 
(endogenous QC with a concentration between LLOQ and low QC, if possible) and spiked with known 
amounts of the authentic analyte (low QC, middle QC and high QC). In spiked samples (e.g., LLOQ, 
low QC), the added amount should be enough to provide concentrations that are three-fold higher or 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 29/45 
 
statistically different from the endogenous concentration. If options such as multiple lots, alternative 
vendors and matrices of special populations that might contain a lower concentration of the analyte 
continue to yield matrices in which the endogenous levels are so high that it is not possible to prepare 
the low QC in the authentic matrix, diluted (surrogate) matrix may be used. 
7.1.2.  Selectivity, recovery and matrix effects for methods for analytes that 
are also endogenous molecules 
The assessment of selectivity is complicated by the absence of interference-free matrix. For 
chromatography, peak purity should be investigated as part of method validation by analysing matrices 
obtained from several donors (at least six normal blanks, one haemolysed blank and one lipemic 
blank) using a discriminative detection system (e.g., tandem mass spectrometry (MS/MS)). Other 
approaches, if justified by scientific principles, may also be considered. 
For the Standard Addition and Background Subtraction Approaches, as the same biological matrix and 
analyte are used for study samples and calibration standards, the same recovery and matrix effect 
occurs in the study samples and the calibration standards. However, if the endogenous component 
were not completely identical (e.g., recombinant proteins) the potential difference in recovery should 
be assessed in a parallelism test. For the Surrogate Matrix and Surrogate Analyte Approaches, the 
matrix effect and the extraction recovery may differ between calibration standards and study samples. 
Matrix effect should be evaluated to ensure that it does not impact accuracy and precision, mainly at 
LLOQ level. 
• 
If the Surrogate Matrix Approach is used, the impact of the different matrix effect and recovery in 
both the surrogate and authentic matrix should be assessed. This should be investigated in an 
experiment using QCs spiked with analyte in the matrix, the endogenous matrix only and spiked 
analyte in the surrogate matrix alone against the surrogate calibration curve. 
• 
If the Surrogate Analyte Approach is used in mass spectrometry chromatographic methods, the 
impact of the different matrix effect and recovery between surrogate and authentic endogenous 
analytes should be evaluated. This should be investigated in an experiment using QCs spiked with 
analyte in the matrix, the endogenous matrix only and spiked surrogate analyte in the matrix 
against the surrogate calibration curve. 
• 
In certain cases, dilution of the QCs with surrogate matrix may be necessary (e.g., for background 
subtraction methods) where the endogenous level is high and the LLOQ needs to be reduced. In 
these cases, the recovery and matrix effect experiments should be repeated with authentic 
biological matrices with endogenous concentrations between LLOQ and low QC, if available. 
Refer to Sections 3 and 4 for the acceptance criteria for chromatography and LBAs, respectively. 
Since the composition of the biological matrix might affect method performance, it is necessary to 
investigate matrices from at least six (chromatographic methods)/ten (LBA) different donors, except in 
the Standard Addition Approach, where each sample is analysed with its own calibration curve. 
7.1.3.  Parallelism for methods for analytes that are also endogenous 
molecules 
Parallelism assures that observed changes in response per given changes in analyte concentrations are 
equivalent for the surrogate and the authentic biological matrix across the range of the method. 
Parallelism should be evaluated in the Surrogate Matrix and Surrogate Analyte Approaches, taking into 
account that parallelism is assessed differently in LBA and chromatographic methods. 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 30/45 
 
7.1.4.  Accuracy and precision for methods for analytes that are also 
endogenous molecules 
Accuracy and precision should meet criteria specified in Sections 3 and 4 for chromatography and 
LBAs, respectively. 
In case of using a surrogate matrix or analyte approaches, the assessment of accuracy and precision 
should be performed by analysing the QCs against the surrogate calibration curve.  
The concentration of the endogenous molecule in the blank matrix may be determined and subtracted 
from the total concentrations observed in the spiked samples. Accuracy is recommended to be 
calculated using the following formula when QCs are spiked with the authentic analyte in the matrix 
containing endogenous levels of the analyte: 
������������������������������������������������������������������������������������������������ (%) = 100 × (Measured concentration of spiked sample − endogenous concentration )
Spiked concentration
 
Only the precision can be determined from the analysis of each unspiked/endogenous QC. 
7.1.5.  Stability for methods for analytes that are also endogenous 
molecules 
To mimic study samples as much as possible, stability experiments should be investigated with the 
authentic analyte in the authentic biological matrix and with unspiked/endogenous QCs (blank matrix 
with endogenous molecule) as well as spiked low QC and high QCs as defined in Section 7.1.1. 
However, if a surrogate matrix is used for calibration standards, stability should also be demonstrated 
for the analyte in the surrogate matrix, as this could differ from stability in the authentic biological 
matrix. 
7.2.  Parallelism 
Parallelism is defined as a parallel relationship between the calibration curve and serially diluted study 
samples to detect any influence of dilution on analyte measurement. Although lack of parallelism is a 
rare occurrence for bioanalytical methods for PK evaluation, parallelism of LBA should be evaluated on 
a case-by-case basis, e.g., where interference caused by a matrix component (e.g., presence of 
endogenous binding protein) is suspected during study sample analysis. Parallelism investigations, or 
the justification for its absence, should be included in the Bioanalytical Report. Some methods may 
demonstrate parallelism for one patient population, but lack it for another population. Generally, these 
experiments should be conducted during the analysis of the study samples due to the unavailability of 
study samples during method development or validation. A study sample with a high concentration 
(preferably close to Cmax) should be diluted to at least three concentrations with blank matrix. The 
consistency of the back calculated concentrations between samples in a dilution series should not 
exceed 30% CV. However, when applying the 30% criterion, data should be carefully monitored as 
results that pass this criterion may still reveal trends of non-parallelism. In the case that the sample 
does not dilute linearly (i.e., in a non-parallel manner), a procedure for reporting a result should be 
defined a priori. 
7.3.  Recovery 
For methods that employ sample extraction, the recovery (extraction efficiency) should be evaluated. 
Recovery is reported as a percentage of the known amount of an analyte carried through the sample 
extraction and processing steps of the method. Recovery is determined by comparing the analyte 
response in a biological sample that is spiked with the analyte and processed, with the response in a 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 31/45 
 
biological blank sample that is processed and then spiked with the analyte. Recovery of the analyte 
does not need to be 100%, but the extent of recovery of an analyte and of the IS (if used) should be 
consistent. Recovery experiments are recommended to be performed by comparing the analytical 
results for extracted samples at multiple concentrations, typically three concentrations (low, medium 
and high). 
7.4.  Minimum required dilution 
MRD is a dilution factor employed in samples that are diluted with buffer solution to reduce the 
background signal or matrix interference on the analysis using LBA. The MRD should be identical for all 
samples including calibration standards and the QCs and it should be determined during method 
development. If MRD is changed after establishment of the method, partial validation is necessary. 
MRD should be defined in the Validation Report of the analytical method. 
7.5.  Commercial and diagnostic kits   
Commercial or diagnostic kits (referred to as kits) are sometimes co-developed with new chemical or 
biological drugs for point-of-care patient diagnosis. The recommendations in this section of the 
guideline do not apply to the development of kits that are intended for point-of-care patient diagnosis 
(e.g., companion or complimentary diagnostic kits). Refer to the appropriate guideline documents 
regarding regulatory expectations for the development of these kits.  
If an applicant repurposes a kit (instead of developing a new method) or utilises “research use only” 
kits to measure chemical or biological drug concentrations during the development of a novel drug, the 
applicant should assess the kit validation to ensure that it conforms to the drug development standards 
described in this guideline. 
• 
Validation considerations for kit assays include, but are not limited to, the following: 
• 
If the reference standard in the kit differs from that of the study samples, testing should evaluate 
differences in assay performance of the kit reagents. The specificity, accuracy, precision and 
stability of the kit assay should be demonstrated under actual conditions of use in the facility 
conducting the sample analysis. Modifications from kit processing instructions should be completely 
validated.   
• 
Kits that use sparse calibration standards (e.g., one- or two-point calibration curves) should 
include in-house validation experiments to establish the calibration curve with a sufficient number 
of standards across the calibration range. 
• 
Actual QC concentrations should be known. Concentrations of QCs expressed as ranges are not 
sufficient for quantitative applications. In such cases QCs with known concentrations should be 
prepared and used, independent of the kit-supplied QCs. 
• 
Calibration standards and QCs should be prepared in the same matrix as the study samples. Kits 
with calibration standards and QCs prepared in a matrix different from the study samples should 
be justified and appropriate experiments should be performed.  
• 
If multiple kit assay lots are used within a study, lot-to-lot variability and comparability should be 
addressed for any critical reagents included in the kits.   
• 
If a kit using multiple assay plates is employed, sufficient replicate QCs should be used on each 
plate to monitor the accuracy of the assay. Acceptance criteria should be established for the 
individual plates and for the overall analytical run.  
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 32/45 
 
7.6.  New or alternative technologies 
When a new or alternative technology is used as the sole bioanalytical technology from the onset of 
drug development, cross validation with an existing technology is not required.  
The use of two different bioanalytical technologies for the development of a drug may generate data 
for the same product that could be difficult to interpret. This outcome can occur when one platform 
generates drug concentrations that differ from those obtained with another platform. Therefore, when 
a new or alternative analytical platform is replacing a previous platform used in the development of a 
drug it is important that the potential differences are well understood. The data generated from the 
previous platform/technology should be cross validated to that of the new or alternative 
platform/technology. Seeking feedback from the regulatory authorities is encouraged early in drug 
development. The use of two methods or technologies within a comparative BA/BE study is strongly 
discouraged. 
The use of new technology in regulated bioanalysis should be supported by acceptance criteria 
established a priori based on method development and verified in validation. 
7.6.1.  Dried matrix methods 
Dried matrix methods (DMM) is a sampling methodology that offers benefits such as collection of 
reduced blood sample volumes as a microsampling technique for drug analysis and ease of collection, 
storage and transportation. In addition to the typical methodological validation for LC-MS or LBA, use 
of DMM necessitates further validation of this sampling approach before using DMM in studies that 
support a regulatory application, such as:  
• 
Haematocrit (especially for spotting of whole blood into cards) 
• 
Sample homogeneity (especially for sub-punch of the sample on the card/device) 
• 
Extraction of the sample from the dried matrix 
• 
DMM sample collection for ISR 
• 
Care should be taken to ensure sufficient sample volumes or numbers of replicates are retained 
for ISR  
• 
Should be assessed by multiple punches of the sample or samples should be taken in duplicate  
When DMM is used for clinical or nonclinical studies in addition to typical liquid approaches (e.g., liquid 
plasma samples) in the same studies, these two methods should be cross validated as described (refer 
to Section 6.2). For nonclinical TK studies, refer to Section 4.1 of ICH S3A Q&A. Feedback from the 
appropriate regulatory authorities is encouraged in early drug development. 
8.  Documentation 
General and specific SOPs and good record keeping are essential to a properly validated analytical 
method. The data generated for bioanalytical method validation should be documented and available 
for data audit and inspection. Table 1 describes the recommended documentation for submission to the 
regulatory authorities and documentation that should be available at the analytical site at times of 
inspection. This documentation may be stored at the analytical site or at another secure location. In 
this case the documentation should be readily available when requested. 
All relevant documentation necessary for reconstructing the study as it was conducted and reported 
should be maintained in a secure environment. Relevant documentation includes, but is not limited to, 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 33/45 
 
source data, protocols and reports, records supporting procedural, operational, and environmental 
concerns and correspondence records between all involved parties.   
Regardless of the documentation format (i.e., paper or electronic), records should be 
contemporaneous with the event and subsequent alterations should not obscure the original data. The 
basis for changing or reprocessing data should be documented with sufficient detail, and the original 
record should be maintained. 
8.1.  Summary information 
Summary information should include the following items in Section 2.6.4/2.7.1 of the Common 
Technical Document (CTD; or electronic CTD, eCTD)) or reports: 
• 
A summary of methods used for each study should be included. Each summary should provide the 
method title, method identification code, the assay type, the Bioanalytical Report code, effective 
date of the method, and the associated Validation Report codes. 
• 
A summary table of all the relevant Validation Reports should be provided for each analyte, 
including Partial Validation and Cross Validation Reports. The table should include the method 
identification code, the type of method, the reason for the new method or additional validation 
(e.g., to lower the limit of quantification). Changes made to the method should be clearly 
identified. 
• 
A summary table cross-referencing multiple identification codes should be provided when a method 
has different codes for the method, the Validation Reports and the Bioanalytical Reports. 
• 
Discussion of method changes (e.g., evolution of methods, reason(s) for revisions, unique aspects) 
• 
For comparative BA/BE studies, a list of regulatory site inspections including dates and outcomes 
for each analytical site if conducted over the last three years, and one year post study completion. 
8.2.  Documentation for validation and bioanalytical reports 
Table 1 describes the recommended documentation for the Validation and Bioanalytical Reports 
 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 34/45 
 
Table 1: Documentation and Reporting  
Items 
Documentation at the analytical site 
Validation report* 
Bioanalytical report* 
Chromatographic 
system suitability 
Dates, times, and samples used for 
suitability testing 
Not applicable 
Not applicable 
Synopsis 
overview of method 
evolution 
History/evolution of methods (e.g., to 
explain revisions, unique aspects with 
supportive data, if available) 
Not applicable   
Not applicable 
Reference standards 
 
CoA or equivalent alternative to ensure 
quality (including purity), 
stability/expiration/retest date(s), batch 
number, and manufacturer or source 
Records of receipt, use, and storage 
conditions. 
If expired, recertified CoA, or retest of 
quality and identity with retest dates 
A copy of the CoA or equivalent 
alternative including batch/lot 
number, source, quality (including 
purity), storage conditions, and 
expiration/retest date, or table with 
this information. 
If expired, quality and stability at the 
time of use and retest dates and 
retested values. 
A copy of the CoA or equivalent alternative 
including batch /lot number, source, quality 
(including purity), storage conditions, and 
expiration/retest date or a table with this 
information.  
If expired, quality and stability at the time 
of use and retest dates and retested values.  
Internal standard 
IS quality or demonstration of suitability  
Records of receipt, use, and storage 
conditions 
Name of reagent or standard 
Origin 
Name of reagent or standard 
Origin 
 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 35/45 
 
Table 1 continued: Documentation and Reporting 
Items 
Documentation at the analytical site 
Validation report* 
Bioanalytical report* 
Critical reagents 
Name of reagent 
Batch/Lot number 
Source/Origin 
Concentration, if applicable 
Retest date (expiry date) 
Storage conditions        
Name of reagent 
Batch/Lot number 
Source/Origin 
Concentration, if applicable 
Retest date (expiry date) 
Storage conditions    
Name of reagent 
Batch/Lot number 
Source/Origin 
Concentration, if applicable 
Retest date (expiry date) 
Storage conditions    
Stock/Working 
solutions 
Record of preparation, and use of 
stock/working solutions 
Storage location and condition 
 
Notation that solutions were used 
within stability period 
Stock/working solution stability  
Storage conditions 
Notation that solutions were used within 
stability period 
Stock/working solution stability † 
Storage conditions†  
Blank matrix 
Records of matrix descriptions, lot 
numbers, receipt dates, storage 
conditions, and source/supplier 
Description, lot number, receipt dates  
Description, lot number, receipt dates†† 
Calibration 
standards and QCs 
Records and date of preparation 
Record of storage temperature (e.g., log 
of in/out dates, analyst, temperatures, 
and freezer(s)) 
Description of preparation including 
matrix 
Batch number, preparation dates and 
stability period 
Storage conditions (temperatures, 
dates, duration, etc.)  
Description of preparation† 
Preparation dates and stability period  
Storage conditions†   
 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 36/45 
 
Table 1 continued: Documentation and Reporting 
Items 
Documentation at the analytical site 
Validation report* 
Bioanalytical report* 
Standard Operating 
Procedures (SOPs; 
Procedures) 
Procedures for all aspects of analysis, 
such as: 
Method/procedure (validation/analytical) 
Acceptance criteria (e.g., run, calibration 
curve, QCs)  
Instrumentation 
Reanalysis 
ISR 
Record of changes to SOP (change, date, 
reason, etc.) 
A detailed description of the method 
procedures 
 
A list of procedures/analytical protocols 
used for the method  
 
Sample tracking 
Study sample receipt, and condition on 
receipt 
Records that indicate how samples were 
transported and received. Sample 
inventory and reasons for missing 
samples 
Location of storage (e.g., freezer unit) 
Tracking logs of QCs, calibration 
standards, and study samples 
Freezer logs for QCs, calibration 
standards, and study samples entry and 
exit 
Not applicable 
For all studies 
Dates of receipt of shipments number of 
samples  
Sample condition on receipt  
Analytical site storage condition and 
location 
Storage: total duration from sample 
collection to analysis 
List of any deviations from planned storage 
conditions, and potential impact 
Additionally, for Comparative BA/BE Studies 
also include: 
The subject ID 
 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 37/45 
 
Table 1 continued: Documentation and Reporting 
Items 
Documentation at the analytical site 
Validation report* 
Bioanalytical report* 
Analysis 
Documentation and data for system 
suitability checks for chromatography 
Instrument use log, including dates of 
analysis for each run 
Sample extraction logs including 
documentation of processing of calibration 
standards, QCs, and study samples for 
each run, including dates of extraction 
Identity of QCs and calibration standard 
lots, and study samples in each run 
Documentation of instrument settings and 
maintenance 
Laboratory information management 
system (LIMS) 
Validation information, including 
documentation and data for: 
Selectivity, specificity, sensitivity, precision 
and accuracy, carry-over, dilution, 
recovery, matrix effect 
Bench-top, freeze-thaw, long-term, 
extract, and stock solution stability 
Cross/partial validations, if applicable 
For all studies: 
Table of all runs (including failed runs), and 
analysis dates 
Table of calibration standard concentration and 
response functions results (calibration curve 
parameters) of all accepted runs with accuracy 
and precision. 
Table of within- and between- run QC results and 
calibration standards (from accuracy and 
precision runs). Values outside the acceptance 
criteria should be clearly marked.  
Include total error for LBA methods 
Data on selectivity, specificity, dilution linearity 
and sensitivity (LLOQ), carry-over, recovery. 
Bench-top, freeze-thaw, long-term, extract, and 
stock solution stability 
Partial/cross-validation, if applicable  
Append separate report for additional validation, 
if any 
Additionally, for comparative BA/BE studies 
also include: 
Instrument ID for each run in comparative BA/BE 
studies † 
100% of run summary table of accepted and 
failed runs 
For all studies: 
Table of all runs, status (accepted and 
failed), reason for failure, and analysis 
dates. 
Table of calibration standard 
concentration and response function 
results (calibration curve parameters) of 
all accepted runs with accuracy and 
precision.  
Table of QCs results of all accepted runs 
with overall (between-run) accuracy and 
precision results of the QCs and between-
run accuracy and precision results from 
accepted runs.  
Table of reinjected runs with results from 
reinjected runs and reason(s) for 
reinjection 
QCs graphs trend analysis encouraged 
Study concentration results table. 
Additionally, for comparative BA/BE 
studies also include: 
Instrument ID for each run in comparative 
BA/BE studies† 
IS response plots for each analytical run, 
including failed runs  
100% of run summary table of accepted 
and failed runs   
 
 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 38/45 
 
 
Table 1 continued: Documentation and Reporting 
 
Items 
Documentation at the analytical site 
Validation report* 
Bioanalytical report* 
Chromatograms 
and reintegration 
Electronic audit trail:  
100% e-chromatograms of original and 
reintegration from accepted and fail runs 
Reason for reintegration 
Mode of reintegration 100% of run summary 
tables of accepted and failed runs, including 
calibration curve, regression, weighting 
function, analyte and IS response and retention 
time, response ratio, integration type 
 
For all studies: 
Representative chromatograms 
(original and reintegration) 
Reason for reintegration 
Chromatograms may be 
submitted as a supplement 
Additionally, for comparative 
BA/BE studies also include 
100% chromatograms of original 
and reintegration from accepted 
and fail runs.  
100% of run summary table of 
accepted and failed runs 
 
For all studies: 
Chromatograms may be submitted as a 
supplement 
For studies other than comparative BA/BE, 
randomly selected chromatograms from 5% 
of samples submitted in application dossiers 
Reason for reintegration 
Identification and discussion of 
chromatograms with manual reintegration 
SOP for reintegration, as applicable 
Additionally, for comparative BA/BE 
studies also include 
100% of chromatograms.  
Original and reintegrated chromatograms 
and initial and repeat integration results 
100% of run summary table of accepted 
and failed runs 
 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 39/45 
 
Table 1 continued: Documentation and Reporting 
 
Items 
Documentation at the analytical site 
Validation report* 
Bioanalytical report* 
Deviations from procedures 
Contemporaneous documentation of 
deviations/unexpected events 
Investigation of unexpected events 
Impact assessment 
Description of deviations 
Impact on study results 
Description and supporting 
data of significant 
investigations 
Description of deviations 
Impact on study results 
Description and supporting data of 
significant investigations 
Reanalysis/Repeat analysis 
Procedures for conducting reanalysis/repeat 
analysis (define reasons for reanalysis, etc.) 
Retain 100% of repeat/reanalysed data 
Contemporaneous records of reason for repeats 
Not applicable 
For all studies: 
Table of sample IDs, reason for 
repeat analysis, original and repeat 
analysis values, reason for reported 
values, run IDs  
Additionally, for comparative 
BA/BE studies also include 
For comparative BA/BE studies, 
values from rejected runs should be 
included in a separate table. 
ISR 
Procedure for ISR 
ISR data: Run IDs, run summary sheets, 
chromatograms or other electronic instrument 
data files 
Document ISR failure investigations, if any 
Not applicable 
ISR data table (original and 
reanalysis values and run IDs, 
percent difference, percent passed) 
ISR failure investigations, if any†† 
 
 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 40/45 
 
Table 1 continued: Documentation and Reporting 
Items 
Documentation at the analytical site 
Validation report* 
Bioanalytical report* 
Communication 
Between involved parties (Applicant, contract 
research organizations, and consultants) related 
to study/method 
Not applicable 
Not applicable 
Audits and inspections 
Evidence of audit and inspections  
Not applicable/Refer to 
Section 8.1 for summary 
information to include in the 
eCTD 
Not applicable/Refer to Section 8.1 
for summary information to include 
in the eCTD 
*The applicant is expected to maintain data at the analytical site to support summary data submitted in validation and bioanalytical reports. validation and 
bioanalytical reports should be submitted in the application. 
† May append or link from validation report. 
††Submit either in validation report or in bioanalytical report
 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 41/45 
 
9.  Glossary 
Accuracy:  
The degree of closeness of the measured value to the nominal or known true value under prescribed 
conditions (or as measured by a particular method). In this document accuracy is expressed as percent 
of the nominal value. 
Accuracy (%) = (Measured Value/Nominal Value) × 100 
Analysis:  
A series of analytical procedures from sample processing/dilution to measurement on an analytical 
instrument. 
Analyte:  
A specific chemical moiety being measured, including an intact drug, a biomolecule or its derivative or a 
metabolite in a biological matrix. 
Analytical Run (also referred to as “Run”):  
A complete set of analytical and study samples with appropriate number of calibration standards and 
QCs for their validation. Several runs may be completed in one day or one run may take several days to 
complete.  
Anchor Calibration Standards/Anchor Points:  
Spiked samples set at concentrations below the LLOQ or above the ULOQ of the calibration curve and 
analysed to improve curve fitting in LBAs. 
Batch (for Bioanalysis): 
A batch is comprised of QCs and study samples, and possibly blanks, zero samples and calibration 
standards, which are handled during a fixed period of time and by the same group of analysts with the 
same reagents under homogenous conditions. 
Batch (for Reference Standards and Reagents): 
A specific quantity of material produced in a process or series of processes so that it is expected to be 
homogeneous within specified limits. Also referred to as “Lot”.  
Bias: 
The tendency of a measurement process to over- or under-estimate the value of a population parameter. 
Bioanalytical Method:  
Analytical method used in the quantitative determination of analytes in biological matrices. 
Biological Drugs: 
Drugs that are made by living organisms or cells (e.g., therapeutic proteins). 
Biological Matrix:  
A biological material including, but not limited to, blood, serum, plasma and urine. 
Binding Reagent:  
A reagent that binds to the analyte in LBA-based bioanalytical methods. 
Blank Sample:  
A sample of a biological matrix to which no analyte, no IS and no additional-alternative matrix or buffer 
has been added.  
 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 42/45 
 
Calibration Curve:   
The relationship between the instrument response (e.g., peak area, height or signal) and the 
concentration (amount) of analyte in the calibration standards within a given range. Also referred to as 
Standard Curve.  
Calibration Range:  
The interval between the upper and lower concentration (amounts) of analyte in the calibration standards 
(including these concentrations)  
Calibration Standard:  
A matrix to which a known amount of analyte has been added or spiked. Calibration standards are used 
to construct calibration curves. 
Carry-over:  
The appearance of an analyte signal in a sample from a preceding sample.  
Chemical Drugs: 
Chemically synthesised drugs.  
Critical Reagent:  
Critical reagents for LBAs include binding reagents (e.g., antibodies, binding proteins, peptides) and 
those containing enzymatic moieties that have a direct impact on the results of the assay. 
Cross Validation:  
Assessment of potential bias between two bioanalytical methods or the same bioanalytical method used 
in different laboratories in order to determine whether reported data are comparable. 
Dilution Integrity:  
Assessment of the sample dilution procedure to confirm that the procedure does not impact the measured 
concentration of the analyte. 
Dilution Linearity:  
A parameter demonstrating that the method can appropriately analyse samples at a concentration 
exceeding the ULOQ of the calibration curve without influence of prozone (hook) effect and that the 
measured concentrations are not affected by dilution within the calibration range in LBAs. 
Dilution Factor: 
The magnitude by which a sample is diluted. 
Full Validation:   
Establishment of all validation parameters that ensure the integrity of the method when applied to 
sample analysis. 
Hook Effect:  
Suppression of response due to very high concentrations of a particular analyte. A hook effect may occur 
in LBAs that use a liquid-phase reaction step for incubating the binding reagents with the analyte. Also 
referred to as prozone effect. 
Incurred Sample:  
A sample obtained from study subjects or animals.  
Incurred Sample Reanalysis (ISR):   
Reanalysis of a portion of the incurred samples in a separate analytical run on a different day to 
determine whether the original analytical results are reproducible.  
 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 43/45 
 
Interfering Substance:  
A substance that is present in the matrix that may affect the quantification of an analyte. 
Internal Standard (IS):  
A structurally similar analogue or stable isotope labelled compound added to calibration standards, QCs 
and study samples at a known and constant concentration to facilitate quantification of the target 
analyte. 
Ligand Binding Assay (LBA):  
A method to analyse an analyte of interest using reagents that specifically bind to the analyte. The 
analyte is detected using reagents labelled with e.g., an enzyme, radioisotope, fluorophore or 
chromophore. Reactions are carried out in microtitre plates, test tubes, disks, etc. 
Lower Limit of Quantification (LLOQ):  
The lowest amount of an analyte in a sample that can be quantitatively determined using a method with 
predefined precision and accuracy. 
Matrix Effect:  
The direct or indirect alteration or interference in response due to the presence of unintended analytes 
or other interfering substances in the sample. 
Minimum Required Dilution (MRD):  
The initial dilution factor by which biological samples are diluted with buffer solution for the analysis by 
LBAs. The MRD may not necessarily be the ultimate dilution but should be identical for all samples 
including calibration standards and QCs. However, samples may require further dilution. 
Nominal Concentration:  
Theoretical or expected concentration. 
Parallelism:   
Parallelism demonstrates that the serially diluted incurred sample response curve is parallel to the 
calibration curve. Parallelism is a performance characteristic that can detect potential matrix effects.  
Partial Validation:  
Validation based on evaluation of selected validation parameters. Applicable to methods that were 
changed after full validation. 
Precision:   
The closeness of agreement (i.e., degree of scatter) among a series of measurements. Precision is 
expressed as the coefficient of variation (CV) or the relative standard deviation (RSD) expressed as a 
percentage. 
%CV = (Standard Deviation/Mean) x 100 
Processed Sample:  
The final sample that has been subjected to various manipulations (e.g., extraction, dilution, 
concentration). 
Quality Control Sample (QC):  
A biological matrix spiked with a known quantity of analyte that is used to monitor the performance of a 
bioanalytical method and assess the integrity and validity of the results of the unknown samples analysed 
in an individual batch or run.  
Reanalysis: 
An additional evaluation of a previously assayed sample. Also referred to as Repeat Analysis. 
 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 44/45 
 
Recovery:  
The extraction efficiency of an analytical process, reported as a percentage of the known amount of an 
analyte carried through the sample extraction and processing steps of the method.  
Reference Standard: 
A well-characterised substance of known purity and identity used to prepare calibration and quality 
control samples.  
Reintegration:  
Change of the original integration of a chromatographic peak 
Replicate:  
One of several determinations or measurements of a sample, calibration standards or QC. 
Reproducibility:  
The extent to which consistent results are obtained when an experiment is repeated.  
Response Function:   
A mathematical expression which adequately describes the relationship between instrument response 
(e.g., peak area or height ratio or signal) and the concentration (amount) of analyte in the calibration 
standards. Response function is defined within a given range. See also Calibration Curve. 
Run Summary Table: 
Tabular output of all data from individual samples, QCs and calibration standards within the analytical 
run (e.g., for chromatography retention times, analyte and IS responses, concentrations, and dilution 
factors if any; for ligand binding assays analyte responses concentrations, dilution factors). 
Selectivity:  
Ability of an analytical method to differentiate and measure the analyte in the presence of interfering 
substances in the biological matrix (non-specific interference).  
Sensitivity:  
The lowest analyte concentration that can be measured with acceptable accuracy and precision (i.e., 
LLOQ). 
Specificity:  
Ability of an analytical method to detect and differentiate the analyte from other substances, including 
its related substances (e.g., substances that are structurally similar to the analyte, metabolites, isomers, 
impurities or concomitant medications). 
Stability: 
Measure of the intactness of an analyte (lack of degradation) in a given matrix under specific storage 
and use conditions relative to the starting material for given time intervals. 
Standard Curve:  
The relationship between the instrument response (e.g., peak area, height or signal) and the 
concentration (amount) of analyte in the calibration standards within a given range. Also referred to as 
calibration Curve.  
Standard Operating Procedure (SOP):  
Detailed written instructions to achieve uniformity of the performance of a specific function and/or 
process(es). 
Stock Solution: 
 
 
ICH guideline M10 on bioanalytical method validation  
 
EMA/CHMP/ICH/172948/2019  
Page 45/45 
 
An analyte in a solvent or mixture of solvents at a known concentration, which is used to prepare 
calibration standards or QCs. 
Study samples:  
Samples from animals or subjects enrolled in nonclinical or clinical studies. 
Surrogate Matrix:  
An alternative to a study matrix of limited availability (e.g., tissue, cerebrospinal fluid, bile) or where the 
study matrix contains an interfering endogenous counterpart. 
System Suitability:  
Determination of instrument performance (e.g., signal-to-noise ratio, peak shape, retention time) by 
analysis of a prepared, spiked sample conducted prior to the analytical run and is not a part of the 
sample analysis. 
Total Error:   
The sum of the absolute value of the errors in accuracy (%) and precision (%). Total error is reported 
as percent (%) error. 
Upper Limit of Quantification (ULOQ):  
The highest amount of an analyte in a sample that can be quantitatively determined with pre-defined 
precision and accuracy. 
Validation:  
Demonstration that a bioanalytical method is suitable for its intended purpose. 
Working Solution:  
A non-matrix solution prepared by diluting the stock solution in an appropriate solvent. It is mainly added 
to matrix to prepare calibration standards and QCs. 
Zero Sample:  
A blank sample spiked with an IS. 
 
  
 